Paper
18 August 2018 Dual sensing-light-sheet OCT for microfluidic PTV
Author Affiliations +
Abstract
A novel dual beam Optical Coherence Tomography (OCT) system is presented for microfluidics applications. The dual sensing-light-sheet approach presented here achieves high velocity flow measurement because the rapid re-imaging by the second sensing-light-sheet allows for particle tracking between each image-frame of the pair. We have demonstrated measurements of microchannel flow using 10 μm particles in a 700 μm deep channel with peak velocities of approximately 0.28 metres-per-second (m/s), approximately two orders of magnitude higher than previously reported OCT systems operating at a few millimeters-per-second.
© (2018) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
J. M. Hallam, E. Rigas, H. D. Ford, T. O. H. Charrett, and R. P. Tatam "Dual sensing-light-sheet OCT for microfluidic PTV", Proc. SPIE 10749, Interferometry XIX, 107490M (18 August 2018); https://doi.org/10.1117/12.2319582