1.INTRODUCTIONPrecision-aligned, robust, ultra-stable optical assemblies are required in an increasing number of space-based applications in areas including fundamental physics, metrology and geodesy. Often these optical systems are constructed from low expansion materials such as Corning’s Ultra Low Expansion Glass (ULE) or Schott’s Extremely Low Expansion Glass Ceramic (Zerodur), and with fused silica optics. One technique for bonding any combination of these materials is hydroxide catalysis bonding. Hydroxide catalysis bonding is essentially a glueless technique that uses an extremely small volume of hydroxide solution to catalyse bonding between the materials1 producing extremely strong, stable structures. Structures built using this technique have already been flown successfully in NASA’s Gravity Probe-B2, and in the European Space Agency’s LISA Pathfinder mission3,4,5 and are planned for the European Space Agency’s planned LISA mission6 to observe low frequency gravitational waves. 2.ALIGNMENT AND BONDINGOur system of aligning and bonding is described in detail in a separate papers7,8, but it is useful to summarise briefly the alignment requirements and technologies used. Figure 1 shows a component to be bonded and the axes used to describe its location. 2.1Component preparation and manipulationThe surfaces to be bonded must be conformal. Typically this mean that both are polished flat to λ/10 (λ =633nm) over the bond area. The bond is thin and uniform (of order 100nm) so the final alignment of the optical surface in γ (about Y axis) is determined by the manufacture of the component and baseplate. Typically for a high-quality component a perpendicularity between optical surface and bonding surface of 5 μradians (1 arc second) can be achieved. Cleaning of the surfaces prior to bonding is also critical to successful bonding. The component is mounted from a 6 degree of freedom hexapod actuator (Physik Instrument H824). This allows control of the position and angle of the component to better than 0.1μm and 0.3μradians, which are well within our alignment goals. 2.2MetrologyIn order for the component to bond in the desired position, they should be brought together parallel, i.e. aligned in angles β and γ so that there is minimal slippage as the bonding process takes place. The overall automation of the process also requires continuous readout of the component-to-baseplate separation. These measurements are made by the bonding surface interferometer. A frequency modulated laser beam is directed through the component along the Z axis and an interference signal results from the overlap of the reflections from the base of the component and the baseplate. The interference pattern is spatially analysed to derive angle information while the observed depth of modulation allows calibration of the separation readout. An angular accuracy of around 1 μradian and displacement accuracy – once calibrated by using known steps of the Hexapod actuator – is better than 30μm, both well within our requirements. In our tests to demonstrate our automated system we define a target position, X, and angle, α, of the optical surface. To monitor the alignment we reflect a positionally stable, amplitude modulated, laser beam from the optical surface and measure the position of the reflected beam on two quadrant diodes located at different distances from the reflection point9. 3.PROTOTYPE SYSTEMOur prototype system, developed to refine the bonding procedure and to produce a series of test bonds, is shown in Figure 2. 3.1ProcessThe process for a single bond is almost completely automated and follows the following sequence:
The overall process can now be reliably completed by a single operator in a timescale of around 30 minutes; previously bonds of this precision took a team of several operators around 1 day to complete. 4.ALIGNMENT RESULTSA series of 9 test bonds were completed using the described procedure using two subtly different bonding fluids. The overall alignment results were extremely good as shown in Table 1. In all cases the reflecting optical surface is within 4μm of its nominal position and the angular errors are below 10μradians. Full details of the bonding timeline and the fluids used are given in7. Table 1.Alignment errors for the 9 test bonds. Bonds 1 to 4 were made with a pure sodium hydroxide solution and bonds 5 to 9 were with the sodium hydroxide solution with added sodium silicate. There are no significant differences in final alignment achieved with the two solutions. The uncertainty for the measurements is also shown.
5.PRODUCTION FACILITY DESIGNThe work described above has demonstrated a system and procedure to achieve the precision alignment and bonding of a single component in a test setup. The next step is to design and build a full production facility for complex optical assemblies. 5.1Target designOur current design is driven by the need to build optical benches doer the planned LISA mission, but is relevant for any complicated optical system. As an indicator of the scale and complexity of the assembly scenario we are targeting, the preliminary design of the LISA optical bench is shown in Figure 3. The design of the facility is based on a conservative LISA optical bench of diameter 540mm that can have components placed on it in any position and orientation on either of the two flat surfaces. In addition there must be space around the optical bench to mount beam measuring devices9 and the whole volume must be accessible for measurement by a coordinate measuring machine (CMM). Our draft facility design is shown in Figure 4 in which a compact bonding surface interferometer and hexapod are mounted on a boom from a wide range XY translation stage. The interface between the boom and hexapod is a rotation stage. The whole assembly is mounted within the measurement volume of a coordinate measuring machine. The bonding surface interferometer and the hexapod component positioner already demonstrated are being re-engineered into a more compact form suitable for this build facility. An initial prototype of the redesigned system is shown in Figure 5 which shows the interferometer and component holder mounted under a smaller hexapod. We are currently testing a prototype interferometer using a larger hexapod and initial performance is good. 6.CONCLUSIONSWe have demonstrated an automatic alignment system that allows us to build optical systems with optics precision-aligned to within 4μm and 10μrad of target and bonded within these limits using the hydroxide catalysis process. The alignment technique uses interferometric sensing of the separation and relative alignment of the surfaces to be bonded, and uses these signals, along with the position and angle of a beam reflected from the optical surface, to control a 6 degree of freedom component manipulator. The overall process is reliable and reduces the time taken for a precision bond from of order 1 day, to less than 30 minutes. We are further developing the system with the aim of incorporating it into a facility for the building of complex, precision aligned, optical systems such as those required for LISA. 7.ACKNOWLEDGEMENTSThis work was funded by the UK Space Agency and by the European Space Agency. We acknowledge support by the University of Glasgow, the UK Astronomy Technology Centre, and the Scottish Universities Physics Alliance. REFERENCESvan Veggel, A.A. and Killow, C.J.,
“Hydroxide catalysis bonding for astronomical instruments,”
Advanced Optical Technologies,
(3),
(2013). https://doi.org/10.1515/aot-2014-0022 Google Scholar
Everitt, C.W.F. DeBra, D.B. Parkinson, B.W. Turneaure, J.P. Conklin, J.W. Heifetz, M.I. Keiser, G.M. Silbergleit, A.S. Holmes, T. Kolodziejczak, J. Al-Meshari, M. Mester, J.C. Muhlfelder, B. Solomonik, V. Stahl, K. Worden, P. Bencze, W. Buchman, S. Clarke, B. Al-Jadaan, A. Al-Jibreen, H. Li, J. Lipa, J.A. Lockhart, J.M. Al-Suwaidan, B. Taber, M. and Wang, S.,
“Gravity Probe B: Final Results of a Space Experiment to Test General Relativity,”
PRL,
(106), 221101
(2011). https://doi.org/10.1103/PhysRevLett.106.221101 Google Scholar
Armano, M. et al,
“Sub-femto-g free fall for space-based gravitational wave observatories: LISA Pathfinder results,”
Phys. Rev. Lett.,
(116), 231101
(2016). https://doi.org/10.1103/PhysRevLett.116.231101 Google Scholar
Armano, M. et al,
“Beyond the Required LISA Free-Fall Performance: New LISA Pathfinder Results down to 20uHz,”
Phys. Rev. Lett.,
(120), 061101
(2018). https://doi.org/10.1103/PhysRevLett.120.061101 Google Scholar
Robertson, D.I Fitzsimons, E.D. Killow, C.J. Perreur-Lloyd, M. Ward, H. Bryant, J. Cruise, A.M. Dixon, G. Hoyland, D. Smith, D. and Bogenstahl, J.,
“Construction and testing of the optical bench for LISA Pathfinder,”
Classical and Quantum Gravity,
(30), 085006
(2013). https://doi.org/10.1088/0264-9381/30/8/085006 Google Scholar
K. Danzmann, K. et al.,
“LISA: Unveiling a Hidden Universe,”
3
(2011). Google Scholar
Robertson, D.I. Fitzsimons, E.D. Killow, C.J. Fitzsimons, E.D. Perreur-Lloyd, M. and Ward, H.,
“Automated precision alignment of optical components for hydroxide catalysis bonding,”
(2018). https://doi.org/10.1364/OE.26.028323 Google Scholar
Killow, C.J. Fitzsimons, E.D. Perreur-Lloyd, M. Robertson, D. I. and Ward, H.,
“Mechanisation of precision placement and catalysis bonding of optical components,”
in International Conference on Space Optics, Proc. of SPIE,
(2016). Google Scholar
Fitzsimons, E.D. Bogenstahl, J. Hough, J. Killow, C.J. Perreur-Lloyd, M. Robertson, D.I. and Ward, H.,
“Precision absolute positional measurement of laser beams,”
Applied Optics,
(52), 2527
–2530
(2013). https://doi.org/10.1364/AO.52.002527 Google Scholar
|