1 May 1992 Comparison of detectability in pulsed versus continuous fluoroscopy: a simulation study
Richard Aufrichtig, Cecil W. Thomas, Grover W. Gilmore, David L. Wilson
Author Affiliations +
Abstract
X-ray fluoroscopy is a significant source of x-ray dose to patients and hospital staff. One technique proposed for reducing dose is pulsed fluoroscopy at reduced frame rates, typically 15 frames/sec for cardiac angiography. Because the human visual system acts as a temporal low-pass filter, simply reducing the frame rate may not allow a dose reduction. In fact, one can argue that for equivalent visualization, the dose per frame should be doubled when the frame rate is halved. We address the question of proper dose by comparing detectability in simulated pulsed and continuous fluoroscopy displays on a unique device called the video tachistoscope. The visual task for the subjects is to detect stationary, computer-generated, low- contrast disks on a flat background in the presence of noise. The disks form a contrast-detail phantom and are displayed in continuous and pulsed mode on either side of a single video monitor. We find similar detectability and visibility in noise when the dose for pulsed is reduced by 25% of that for continuous.
© (1992) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Richard Aufrichtig, Cecil W. Thomas, Grover W. Gilmore, and David L. Wilson "Comparison of detectability in pulsed versus continuous fluoroscopy: a simulation study", Proc. SPIE 1653, Medical Imaging VI: Image Capture, Formatting, and Display, (1 May 1992); https://doi.org/10.1117/12.59515
Lens.org Logo
CITATIONS
Cited by 5 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Fluoroscopy

Visibility

Visualization

Video

X-rays

Angiography

Visual system

RELATED CONTENT

Reduction of Display Artifacts by Random Sampling
Proceedings of SPIE (January 09 1984)
Rotatable Spoke Fluoroscopic Test Object
Proceedings of SPIE (August 28 1984)
Quality metric design: a closer look
Proceedings of SPIE (June 02 2000)
CdZnTe detector arrays for low-dose x-ray fluoroscopy
Proceedings of SPIE (October 06 1999)

Back to Top