Abstract
A rugged, easy to implement, line-of-sight absorption instrument, which utilizes a low- pressure water vapor microwave discharge cell as the light source, has been developed to make simultaneous measurements of the OH concentration and temperature at 10 spatial positions. The design, theory, and capability of the instrument are discussed. Results of the measurements obtained on a methane/air flat flame burner are compared with those obtained using a single-frequency, tunable dye laser system.
© (1994) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Behrooz Shirinzadeh and Ray W. Gregory "Resonance lamp absorption technique for simultaneous determination of the OH concentration and temperature at 10 spatial positions in combustion environments", Proc. SPIE 2122, Laser Applications in Combustion and Combustion Diagnostics II, (18 March 1994); https://doi.org/10.1117/12.171298