1 April 1994 Perception in x-ray fluoroscopy
David L. Wilson, Ping Xue, Richard Aufrichtig
Author Affiliations +
Abstract
We find that pulsed fluoroscopy of stationary objects gives average dose savings of 22 percent, 38 percent, and 49 percent at 15, 10, and 7.5 acq/sec, respectively. Dose savings depend on object size, with less savings for smaller objects. A model that includes the spatio-temporal response of the human visual system (HVS) describes these results with one free parameter. Pulsed-30 is always better than continuous for detecting moving objects. Likewise, pulsed-15 saves dose as compared to pulsed-30, but the savings decrease for high velocities and small objects.
© (1994) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
David L. Wilson, Ping Xue, and Richard Aufrichtig "Perception in x-ray fluoroscopy", Proc. SPIE 2166, Medical Imaging 1994: Image Perception, (1 April 1994); https://doi.org/10.1117/12.171745