Paper
8 September 1995 Research on calibration technique for a high-precision optical angle gauge
Author Affiliations +
Abstract
Optical angle gauge is one of the measuring instruments standardized by the National Metrological System of Chian for verification of plane angle. In this paper a new method for verifying the optical angle gauge is described and a related photoelectric goniometer developed by ourself is introduced. As a datum instrument for measuring deviation angle of optical wedge, its maximum error is less than 0.1 inch. In addition, some examples of its application for calibrating indication error of various angle measuring devices are presented, and its distinctive advantages, especially in calibrating 2D autocollimators, are demonstrated.
© (1995) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Jiaming Lin, Genrui Cao, and Datu Su "Research on calibration technique for a high-precision optical angle gauge", Proc. SPIE 2536, Optical Manufacturing and Testing, (8 September 1995); https://doi.org/10.1117/12.218459