26 May 1997 Posterior fixation keratoprostheses and mechanical biocompatibility: determination of critical intraocular pressure causing aqueous humor leak and/or keratoprosthesis extrusion
Hassan Tahi, Bernard Duchesne M.D., Jean-Marie A. Parel, Izuru Nose, David B. Denham, Franck L. Villain, Emmanuel Lacombe
Author Affiliations +
Proceedings Volume 2971, Ophthalmic Technologies VII; (1997) https://doi.org/10.1117/12.275109
Event: BiOS '97, Part of Photonics West, 1997, San Jose, CA, United States
Abstract
The effect of increased intraocular pressure (IOP) in human cadaver eyes implanted with posterior fixation keratoprosthesis was evaluated. Experiments were carried out with six fresh pairs of human cadaver eyes. One eye of each pair was implanted with a PCL-5 keratoprosthesis (8.60 mm diameter with an optic of 5.60 nm diameter) and the contralateral eye was used as a control. The keratoprosthesis was inserted through a 6 mm diameter opening trephined in the cornea. The resistance of the implanted eye to pressure on `aqueous humor' leak and/or keratoprosthesis extrusion was tested by infusing water at a constant flow of 60 mmHg/second into the anterior chamber. IOP variations were recorded with a transducer connected to a computer. IOP could be increased up to 1520 to 2324 mmHg before aqueous humor leaks occurred. Leaks were always located at the keratoprosthesis-cornea interface. No prosthesis extrusion was observed. Implanted eyes that did not leak aqueous and control eyes tore at the sclera. All posterior fixation keratoprostheses implanted eyes resisted more than 100 times the normal physiological intraocular pressure and on this standpoint is safe. Additional experiments were needed to assess the influence of suture fixation and wound healing in an animal model.
© (1997) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Hassan Tahi, Bernard Duchesne M.D., Jean-Marie A. Parel, Izuru Nose, David B. Denham, Franck L. Villain, and Emmanuel Lacombe "Posterior fixation keratoprostheses and mechanical biocompatibility: determination of critical intraocular pressure causing aqueous humor leak and/or keratoprosthesis extrusion", Proc. SPIE 2971, Ophthalmic Technologies VII, (26 May 1997); https://doi.org/10.1117/12.275109
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Eye

Cornea

Haptic technology

Tissues

Wound healing

Information operations

Optical components

RELATED CONTENT

Novel posterior fixation keratoprosthesis
Proceedings of SPIE (August 14 1992)
Specialized modal tomography of human eye aberrations
Proceedings of SPIE (August 01 2007)
Design improvement in keratoprosthesis
Proceedings of SPIE (June 23 1994)

Back to Top