28 June 2001 Megavoltage radiation dose response of digital computed radiography
Zhiheng Wang, Qingrong J. Wu, Claudio H. Sibata
Author Affiliations +
Abstract
The response of a digital computed radiography system to the megavoltage therapeutic radiation beams was invested. A narrow slit of radiation beam was used to test the line spread function of the system. The effects of various facts such as cassette, beam energy, radiation dose, scanning orientation and timing on the line spread function were investigated. The calibration curves were established to calibrate the image intensity to the megavoltage radiation dose. The calibration curves were applied to measure the beam profiles of the radiation fields with various wedges.
© (2001) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Zhiheng Wang, Qingrong J. Wu, and Claudio H. Sibata "Megavoltage radiation dose response of digital computed radiography", Proc. SPIE 4320, Medical Imaging 2001: Physics of Medical Imaging, (28 June 2001); https://doi.org/10.1117/12.430958