Paper
28 December 2007 Calibration procedure of extreme ultraviolet detectors
Author Affiliations +
Proceedings Volume 6937, Photonics Applications in Astronomy, Communications, Industry, and High-Energy Physics Experiments 2007; 69370F (2007) https://doi.org/10.1117/12.784595
Event: Photonics Applications in Astronomy, Communications, Industry, and High-Energy Physics Experiments 2007, 2007, Wilga, Poland
Abstract
The paper presents a procedure and laboratory setup for calibration of extreme ultraviolet detectors. The procedure determinates a responsivity of a photodiode applied in the technology of EUV radiation. The main elements of the described system are a gas-puff laser plasma source, metrology chamber with an optical beam splitter and a model instrument. The applied measurement method is based on a comparison of the detected signals from the model instrument and the tested detector. The reduction of an error resulting from instabilities of the source radiation is ensured using a special construction of the optical beam splitter. Because of the cost and the metrology features, the presented setup is future alternative for small factories and laboratories producing EUV detectors. In the paper, analysis and preliminary research of the setup are described.
© (2007) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
J. Mikołajczyk, Z. Bielecki, W. Kołosowski, and E. Sędek "Calibration procedure of extreme ultraviolet detectors", Proc. SPIE 6937, Photonics Applications in Astronomy, Communications, Industry, and High-Energy Physics Experiments 2007, 69370F (28 December 2007); https://doi.org/10.1117/12.784595
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Sensors

Calibration

Extreme ultraviolet

Photodiodes

Instrument modeling

Metrology

Signal detection

RELATED CONTENT

Reading a CD-ROM without a photodiode
Proceedings of SPIE (April 28 2016)
Absolute extreme-ultraviolet metrology
Proceedings of SPIE (June 18 2001)
Testing system for extreme ultraviolet detectors
Proceedings of SPIE (October 07 2005)
Description of the HIRDLS radiometric model (HIRAM)
Proceedings of SPIE (November 18 1998)

Back to Top