11 August 2009 VIIRS polarization sensitivity testing and analysis
Author Affiliations +
Abstract
A requirement for the Visible/Infrared Imager Radiometer Suite (VIIRS) is that its polarization sensitivity be 3% or less for all VISNIR bands (412-865 nm). A test using a rotating polarizer sheet was performed on the sensor to validate this requirement, and though the test results show that the requirement is met, they also show a large variation in this polarization sensitivity (as much as 2%) across the field of view (FOV) in track. Though this result is unexpected, it may be the result of natural variations in the diattenuation and retardance of the VIIRS optics as a function of field angle. To test this theory, a raytracing model of the system was constructed using measured ellipsometric data from the VIIRS optics, and the polarization sensitivity of the model was computed. Using the nominal ellipsometric data, good correlation between the predicted and measured polarization sensitivity was not achieved. However, by applying small variations to the ellipsometric data as a function of position on the optics, it was possible to achieve good correlation. This paper gives the details of the sensor polarization sensitivity measurements, ellipsometric measurements, and raytracing analysis.
© (2009) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Eric C. Fest "VIIRS polarization sensitivity testing and analysis", Proc. SPIE 7461, Polarization Science and Remote Sensing IV, 746102 (11 August 2009); https://doi.org/10.1117/12.828154
Lens.org Logo
CITATIONS
Cited by 3 scholarly publications and 1 patent.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Polarization

Sensors

Mirrors

Data modeling

Optical components

Coating

Ray tracing

RELATED CONTENT


Back to Top