PERSONAL Sign in with your SPIE account to access your personal subscriptions or to use specific features such as save to my library, sign up for alerts, save searches, etc.
The objective of this study is to validate a method for the measurement of tissue oxygen saturation level via a
thermal imaging bundle in-vitro and in-vivo. The method consists of a thermal imaging system and an algorithm
which estimates the compound concentration according to the temperature rise of the tissue. A temperature rise
is obtained by illuminating the tissue in the NIR range and is measured using a thermal camera and a coherent
thermal imaging bundle for non-invasive transendoscopic use. The system was validated using agar phantoms of
varying concentrations of Methylene Blue and ICG as well as blood samples. The algorithm estimated the
Methylene Blue relative amount and the results were compared to the real relative amount. The calculated RMS
of the error was 5.12%, a satisfying value for this stage. In the blood samples, for oxygenation levels higher than
50% the RMS of the error was 5.79%. Once the system was verified a portable system was built for clinical use,
this system was also evaluated on agar phantoms and the RMS of the error was 10.64%. As a result of the
encouraging experiments in-vivo, animal trials were performed. The oxygenation levels of mice were decreased
and were estimated respectively using our system. The system determined a small decrease in the tissue oxygen
saturation of the mice. These results verify the algorithm's and bundle's suitability for the use in a non-invasive
system. They provide motivation for performing more complex in-vitro experiments and moving on to clinical
trials.
Yonat Milstein,Dror M. Allon,James Harrington,Carlos Bledt, andIsrael Gannot
"Photothermal imaging bundle system for estimating tissue oxygen saturation", Proc. SPIE 8218, Optical Fibers and Sensors for Medical Diagnostics and Treatment Applications XII, 82180D (1 February 2012); https://doi.org/10.1117/12.917008
ACCESS THE FULL ARTICLE
INSTITUTIONAL Select your institution to access the SPIE Digital Library.
PERSONAL Sign in with your SPIE account to access your personal subscriptions or to use specific features such as save to my library, sign up for alerts, save searches, etc.
The alert did not successfully save. Please try again later.
Yonat Milstein, Dror M. Allon, James Harrington, Carlos Bledt, Israel Gannot, "Photothermal imaging bundle system for estimating tissue oxygen saturation," Proc. SPIE 8218, Optical Fibers and Sensors for Medical Diagnostics and Treatment Applications XII, 82180D (1 February 2012); https://doi.org/10.1117/12.917008