Using data from our previously described otoscope1 that uses 1310 nm phase-sensitive spectral domain optical coherence tomography (PS-SDOCT), we demonstrate a software technique for improving the signal-to-noise (SNR). This method is a software post-processing algorithm applicable to generic PS-SDOCT data describing phase versus time at a specific depth position. By sub-sampling the time trace and shifting the phase of the subsamples to maximize their correlation, the subsamples can be coherently averaged, which increases the SNR.
© (2013) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Tracy C. Petrie, Sripriya Ramamoorthy, Steven L. Jacques, and Alfred L. Nuttall "Increasing PS-SDOCT SNR using correlated coherent averaging", Proc. SPIE 8565, Photonic Therapeutics and Diagnostics IX, 85651O (8 March 2013); https://doi.org/10.1117/12.2008963