Paper
22 July 2015 High definition in vivo retinal volumetric video rate OCT at 0.6 Giga-voxels per second
Jan Philip Kolb, Thomas Klein, Wolfgang Wieser, Wolfgang Draxinger, Robert Huber
Author Affiliations +
Abstract
We present full volumetric high speed OCT imaging of the retina with multiple settings varying in volume size and volume rate. The volume size ranges from 255x255 A-scans to 160x40 A-scans with 450 samples per depth scan with volume rates varying between 20.8 V/s for the largest volumes to 195.2 V/s for the smallest. The system is based on a 1060nm Fourier domain mode locked (FDML) laser with 1.6MHz line rate. Scanning along the fast axis is performed with a 2.7 kHz or 4.3 kHz resonant scanner operated in bidirectional scanning mode, while a standard galvo scanner is used for the slow axis. The performance is analyzed with respect to various potential applications, like intraoperative OCT.
© (2015) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Jan Philip Kolb, Thomas Klein, Wolfgang Wieser, Wolfgang Draxinger, and Robert Huber "High definition in vivo retinal volumetric video rate OCT at 0.6 Giga-voxels per second", Proc. SPIE 9541, Optical Coherence Imaging Techniques and Imaging in Scattering Media, 95410Z (22 July 2015); https://doi.org/10.1117/12.2183768
Lens.org Logo
CITATIONS
Cited by 2 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Optical coherence tomography

Video

Scanners

Computing systems

Eye

In vivo imaging

Mode locking