1 April 2016 High frame rate synthetic aperture vector flow imaging for transthoracic echocardiography
Carlos A. Villagómez-Hoyos, Matthias B. Stuart, Thor Bechsgaard, Michael Bachmann Nielsen, Jørgen Arendt Jensen
Author Affiliations +
Abstract
This work presents the first in vivo results of 2-D high frame rate vector velocity imaging for transthoracic cardiac imaging. Measurements are made on a healthy volunteer using the SARUS experimental ultrasound scanner connected to an intercostal phased-array probe. Two parasternal long-axis view (PLAX) are obtained, one centred at the aortic valve and another centred at the left ventricle. The acquisition sequence was composed of 3 diverging waves for high frame rate synthetic aperture flow imaging. For verification a phantom measurement is performed on a transverse straight 5 mm diameter vessel at a depth of 100 mm in a tissue-mimicking phantom. A flow pump produced a 2 ml/s constant flow with a peak velocity of 0.2 m/s. The average estimated flow angle in the ROI was 86.22° ± 6.66° with a true flow angle of 90°. A relative velocity bias of −39% with a standard deviation of 13% was found. In-vivo acquisitions show complex flow patterns in the heart. In the aortic valve view, blood is seen exiting the left ventricle cavity through the aortic valve into the aorta during the systolic phase of the cardiac cycle. In the left ventricle view, blood flow is seen entering the left ventricle cavity through the mitral valve and splitting in two ways when approximating the left ventricle wall. The work presents 2-D velocity estimates on the heart from a non-invasive transthoracic scan. The ability of the method detecting flow regardless of the beam angle could potentially reveal a more complete view of the flow patterns presented on the heart.
© (2016) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Carlos A. Villagómez-Hoyos, Matthias B. Stuart, Thor Bechsgaard, Michael Bachmann Nielsen, and Jørgen Arendt Jensen "High frame rate synthetic aperture vector flow imaging for transthoracic echocardiography", Proc. SPIE 9790, Medical Imaging 2016: Ultrasonic Imaging and Tomography, 979004 (1 April 2016); https://doi.org/10.1117/12.2216707
Lens.org Logo
CITATIONS
Cited by 6 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Blood

Tissues

Transducers

Heart

In vivo imaging

Point spread functions

Ultrasonography

Back to Top