14 November 2024 Impact of skin tone on target size detectability in photoacoustic breast imaging
Rhea D. Rasquinha, Mardava R. Gubbi, Muyinatu A. Lediju Bell
Author Affiliations +
Abstract

Significance

Photoacoustic imaging has the potential to improve non-invasive breast cancer diagnosis. However, illumination through the skin introduces a skin tone bias, as greater melanin content increases optical absorption and can create acoustic clutter, reducing the visibility of various target sizes.

Aim

We investigate the impact of skin tone bias as a function of target sizes in three photoacoustic image reconstruction methods: fast Fourier transform (FFT)-based reconstruction, delay-and-sum (DAS) beamforming, and short-lag spatial coherence (SLSC) beamforming.

Methods

The three imaging methods were applied to channel data from multidomain simulations with 757, 800, and 1064 nm wavelengths, 11 target sizes (ranging 0.5 to 3 mm), 18 skin tones [ranging individual typology angles (ITAs), of –54 deg to 60 deg], and a previously validated realistic 3D breast model. The signal-to-noise ratio (SNR) and generalized contrast-to-noise ratio (gCNR) were measured to assess visibility.

Results

With 757 and 800 nm wavelengths, targets underlying dark skin tones (ITA range: 54 deg to 33 deg) with sizes 3 mm were poorly visible with 2.38 to 4.21 mean SNR and 0.46 to 0.74 mean gCNR, with smaller targets generally being more visible with lighter skin tones. A 1064-nm transmit wavelength improved visualization with DAS and SLSC beamforming, relative to both FFT reconstruction with a 1064 nm wavelength and DAS or SLSC beamforming with 757 and 800 nm wavelengths. When combined with SLSC beamforming, the 1064-nm wavelength offered the greatest improvements, enabling visualization of simulated target sizes ranging from 0.5 to 3 mm underlying very light (ITA = 60 deg) to dark (ITA=54 deg) skin tones, with mean SNR 10.01 and mean gCNR 1. Visualization of simulated vessel structures derived from in vivo photoacoustic images was consistent with simulation-based target size expectations.

Conclusions

Results are promising for advancing next-generation photoacoustic imaging systems for breast cancer diagnosis across the range of skin tones represented in healthcare systems throughout the world.

Statement of Discovery

This work provides the first known in silico demonstrations to support the benefits of 1064 nm transmit wavelength combined with short-lag spatial coherence (SLSC) beamforming when visualizing target sizes ranging 0.5–3 mm underlying very light (ITA = 60 deg) to dark (ITA = –54 deg) skin tones.

1.

Introduction

Photoacoustic imaging is an emerging modality for non-invasive screening and detection of breast cancer.1 Photoacoustic imaging is based on the photoacoustic effect, by which optical energy is converted to acoustic energy.2 Biological tissues absorb energy from a laser light source which causes thermal expansion and the emission of acoustic waves3 that can be detected by conventional clinical ultrasound transducers and used to reconstruct an image.4 X-ray mammography is the current standard for routine breast cancer screening, but the sensitivity of this imaging modality decreases in patients with greater breast density.5,6 Alternative screening methods include magnetic resonance imaging and ultrasound imaging, but these are limited by high cost and a high false positive rate, respectively.1,7,8 As the scattering of acoustic waves in tissue is significantly less than that of optical waves,9 photoacoustic imaging can be used to generate high-resolution images of hemoglobin content, which is linked to the presence of cancer.

Hemoglobin has high optical absorption in the near-infrared range relative to water and other tissues,1 which enables photoacoustic imaging systems to use this wavelength range to detect vascular structures. Tumor angiogenesis is a known indication of malignancy,10 and other reported indicators include hypoxia11 and elevated hemoglobin levels.12 Several photoacoustic imaging systems have been developed to differentiate between the vasculature of benign and malignant breast lesions.3 Abeyakoon et al.13 developed a feature set based on combined photoacoustic-ultrasound images that improved diagnostic accuracy of malignant lesions. However, these diagnostic vessels vary in size.1317

To non-invasively image these vascular structures, tissue illumination through the skin is required,18 yet it is limited by the high optical absorption coefficient of skin in the visible and near-infrared range due to the presence of melanin,19,20 which decreases optical penetration depths, particularly for vessels located deeper in the tissue. This high optical absorption can also result in acoustic wave emission from the skin into the tissue below, which are reflected by typical ultrasound scatterers and create artifacts and acoustic clutter, reducing image quality.21,22 As skin optical absorption coefficients increase with greater melanin content,20 these two limitations significantly compromise photoacoustic image quality for individuals with darker skin tones, creating a skin tone bias in photoacoustic imaging systems.2224 In addition, Fernandes et al.25 confirmed skin as the primary source of clutter in photoacoustic images of a 3D breast tissue model. However, the relationship between skin tone and target size detectability in photoacoustic imaging is currently unknown.

Initially developed to predict the risk of sunburn and skin cancer,26 the Fitzpatrick skin type scale has been widely used in dermatology to classify skin tone,27 although this scale is limited by inaccuracy and inconsistent use.28 The individual typology angle (ITA) provides a more objective evaluation that can be calculated based on the Commission Internationale de l’Eclairage L*a*b* color system.29 ITA is also reported to be highly correlated with melanin content in the skin30,31 and is related to the skin optical absorption coefficient and wavelength through the equation proposed by Junior et al.32

Short-lag spatial coherence (SLSC) beamforming is an image reconstruction method shown to reduce clutter and improve photoacoustic image quality in noisy environments with low optical fluence.33,34 The SLSC beamformer computes and sums the spatial coherence between signals received at different transducer elements across the short spatial lag region to construct an image.35 Clutter sources have low spatial coherence, which enables the mitigation of clutter artifacts in SLSC images.3638 Fernandes et al.34 demonstrated a skin tone bias in photoacoustic imaging of the radial artery in human volunteers, which was successfully mitigated by SLSC beamforming, allowing for clear radial artery visualization in skin tones ranging from light to dark. We expect that SLSC beamforming will similarly improve target and vessel visualization in photoacoustic images used to determine the presence of breast cancer when considering a range of vessel sizes in both lighter and darker skin tones.

In this paper, the impact of skin tone on target size detectability in photoacoustic breast imaging is investigated with multidomain simulations implemented using a previously validated realistic 3D breast model.39 Representative target sizes, skin tones, and wavelengths of interest were simulated to quantify target visibility for a range of imaging scenarios with two amplitude-based beamforming methods for comparison with SLSC beamforming. We additionally simulated vessels derived from previously published features of interest (i.e., splayed vessel and claw features13) and then compared results with existing in vivo data.

2.

Methods

Eleven simulation volumes were constructed based on a realistic 3D breast model immersed in water, with breast tissue classified as extremely dense.39 The model volumes were 45×15×20  mm in the lateral, axial, and elevation directions, respectively, with a voxel size of 100  μm in each dimension. In each volume, a spherical photoacoustic target with diameters ranging from 0.5 to 3 mm was generated, in 0.25 increments. This range was chosen based on measurements of blood vessels and other diagnostic features from in vivo patient photoacoustic images,1317 and the optical and acoustic properties of the targets were set to those of blood, which is an important component of tumor detection.1013 The targets were centered in the lateral and elevation dimensions of the model, and the center of each target was placed at a depth of 8.6 mm from the skin surface (10 mm from the top surface of the volume in the axial direction), as shown in Fig. 1.

Fig. 1

Cross-sectional view of the simulated volumes within the imaging plane, demonstrating various tissue components and ranges of simulated photoacoustic target sizes.

BIOS_2_1_012502_f001.png

Optical simulations were conducted using MCXLAB40 to compute the fluence of light in the volumes. A collimated Gaussian beam with a 4-mm waist radius (i.e., 50.3  mm2 illumination area) emitting 108 photons was centered in the lateral and elevation dimensions and placed at the top of the volume. Although the simulated fluence is reported as normalized values, any incident energy divided by the illumination area is expected to provide information about the fluence that can be expected in a real system. To assess the impact of skin tone on target size detectability, 18 ITA values were simulated (ranging from 60 deg to 54deg, including three values per skin tone category, as defined in Refs. 29, 41, and 42 and summarized in Table 1). These ITA values correspond to previously reported melanin content-to-basal length ratios ranging from 0  mm2/mm at an ITA of 60 deg to 2.1  mm2/mm at an ITA of 60deg,43 melanosomes/μm2 ranging 20  μm2 at an ITA of 36.1deg to 100  μm2 at an ITA of 57.4 deg,43 and melanin volume fractions (v/v) ranging from 3.6% at an ITA of 56 deg to 40% at an ITA of 47deg.23 We additionally acknowledge that although ITA quantifies perceived skin tone, this standalone metric does not include information about associated molecular or physiological origins (e.g., whether the skin tone perception is due to increases in melanin, blood flow, or veins near the skin surface),44 and a recent review article recommends the establishment of an effective melanosome volume fraction parameter based on an assumed epidermal thickness, which can otherwise vary when simply reporting the volume fraction.45 For each simulated target size and skin tone, three wavelengths were simulated (i.e., 757, 800, and 1064 nm). These wavelengths were chosen because they are used in existing commercial photoacoustic imaging systems.3,13,15,46

Table 1

Skin tone categories, corresponding ITA values,29,41,42 and specific ITA values simulated.

Skin tone categoryITA (deg)Simulated ITA (deg)
Very Light>5560, 58, 56
Light>41 and 5550, 47, 43
Intermediate>28 and 4138, 35, 30
Tan>10 and 2826, 20, 12
Brown30 and 105, −10, −20
Dark<30−33, −45, −54

The optical absorption coefficients of water,47 blood,47 fat,39 and fibroglandular tissue39 were set based on reported values, and the optical absorption coefficients of skin were calculated using the equation proposed by Junior et al.32 The optical scattering coefficients were similarly set based on reported values,20 and standardized anisotropy factor and index of refraction values of 0.9 and 1.37, respectively, were used. The optical properties of each tissue component20,32,39,47 used in the simulations are summarized in Table 2.

Table 2

Optical properties of tissue components.20,32,39,47

ComponentOptical absorption coefficient (mm−1)Optical scattering coefficient (mm−1)
Water0.0029 (757 nm) 0.0022 (800 nm) 0.0144 (1064 nm)1
SkinCalculated for each wavelength and ITA value using:32μa=45.4·100.004·λ·100.014·ITA20.6104 (757 nm) 19.0949 (800 nm) 13.6070 (1064 nm)
Fat0.0058.5436 (757 nm) 8.3016 (800 nm) 7.1575 (1064 nm)
Fibroglandular0.00413.0649 (757 nm) 11.8491 (800 nm) 7.1567 (1064 nm)
Target (i.e., blood)0.3042 (757 nm) 0.4370 (800 nm) 0.5484 (1064 nm)8.5436 (757 nm) 8.3016 (800 nm) 7.1575 (1064 nm)

Acoustic simulations were conducted using k-Wave.48 The initial pressure was computed as Φ(r)·μa(r)·Γ, where Φ(r) is the light fluence map derived from the output of the optical simulation, μa(r) is the optical absorption coefficient map, and Γ is the Grüneisen parameter, which was set to 1. The initial pressure was multiplied by a randomly generated binary absorbers map in which 50% of the pixels were absorbers. To record the acoustic pressure, a linear array ultrasound transducer defined with 128 elements, 0.3 mm pitch, and a 7-MHz center frequency was centered at the top of the volume to simulate an Ultrasonix L14-5/38 linear array transducer.49,50 The sampling frequency was 40 MHz, and the simulated acoustic properties of each tissue component25,39 are summarized in Table 3.

Table 3

Acoustic properties of tissue components.25,39

ComponentSpeed of sound (m/s)Density (kg/m3)Attenuation (dB/MHz·cm)
Water150010000.0025
Skin165011500.35
Fat14709370.60
Fibroglandular151510400.75
Target (i.e., blood)158410400.20

As some level of background noise is present in most experimental channel data, noise corresponding to a channel signal-to-noise ratio (SNR) of 20 dB was added to the simulated channel data. Photoacoustic images were then reconstructed using three beamforming methods: (i) amplitude-based beamforming using the fast Fourier transform (FFT) reconstruction technique available in k-Wave,48 (ii) standard delay-and-sum (DAS) beamforming, and (iii) SLSC beamforming.35 A short-lag value M=10 was used for the SLSC beamformed images, and the axial correlation kernel length was 0.270 mm (i.e., seven samples, which are equivalent to one acoustic wavelength). Envelope detection was applied to the outputs of the DAS and SLSC beamforming algorithms. Each photoacoustic image was normalized by its brightest pixel. Prior to image display, the normalized image data from the FFT and DAS beamforming algorithms were log compressed. The log-compressed (FFT, DAS) or linear (SLSC) image data were then displayed with the same dynamic range per image formation method to facilitate qualitative comparisons.

The axial and lateral resolution of the FFT and DAS images, determined with full-width at half maximum (FWHM) measurements of a point target surrounded by water, located at a 10-mm depth, is 0.84 and 0.73 mm for the DAS images and 0.79 and 0.41 mm for the FFT images, respectively. These axial and lateral FWHM values were calculated by averaging the columns and rows, respectively, immediately adjacent to the target center line (i.e., an average of three rows or columns total). The corresponding SLSC lateral FWHM determined with PhocoSpace33,51,52 and target sizes of 0.5 and 3 mm are 2.11 and 3.16 mm, respectively, when using the same parameters described above (e.g., a Gaussian light profile width of 8 mm, which corresponds to 4-mm waist radius; an M value of 10; an initial pressure distribution SNR of 20 dB; an imaging field depth of 15 mm; a target center depth of 10 mm with background and target optical absorption coefficients of 0.144 and 5.484  cm1, respectively; an absorber density of 5×107  absorbers/m2, which corresponds to 50% of pixels being absorbers; and a transducer with 128 elements, 0.3 mm pitch, 4.9 MHz bandwidth, and 7 MHz center frequency). However, it is known that multiple factors can affect SLSC lateral resolution (e.g., light profile size, target size, noise, M value, and correlation kernel size33,5154). The SLSC axial resolution is expected to be greater than the axial correlation kernel length54 (i.e., >0.270  mm).

To quantify target detectability, the SNR and generalized contrast-to-noise ratio (gCNR)5557 were calculated from the normalized image data. The SNR was calculated using the equation

Eq. (1)

SNR=μiσo,
where μi is the mean signal amplitude (i.e., after envelope detection and/or normalization, prior to any log compression) within the circular region of interest (ROI) inside the target and σo is the standard deviation of signal amplitudes within the annular ROI outside the target. Each ROI was determined once per target size, using FFT images with no added noise, created with the 757 nm wavelength and 60 deg ITA, as these images were the closest to the ground truth. The circular target ROI was the same size as the target size and concentric with the diameter of the target. The annular background ROI was created with an inner radius 2·r, where r is the radius of the target ROI, and an outer radius such that the area of the background and the target ROIs were equal. The same ROIs were then replicated for each wavelength, ITA value, and beamforming method investigated, and were applied to the following gCNR equation (which is a discretization of the equation used to calculate gCNR with probability density functions56,57):

Eq. (2)

gCNR=1k=0N1min{hi(xk),ho(xk)},
where N is the number of bins centered at {x0,x1,,xN1}; hi and ho are the histograms computed for the target and background ROIs, respectively; and k is the bin index. A total of N=32 bins were used to create the histograms. These image analyses were performed with MATLAB software (Mathworks, Natick, MA, United States).

Although standardized spherical targets are useful in analyzing target detectability limits, clinical photoacoustic targets are more complex in shape and structure. To translate results to more realistic models, two additional simulation volumes were generated with incorporated blood vessel structures based on patient photoacoustic images.13 A splayed vessel and claw, which are features of benign and malignant lesions, respectively, were generated in Fusion 360 (Autodesk, San Francisco, CA, United States) and incorporated into the 3D breast model39 using MATLAB by creating a triangulation from the exported .stl files. A slice of each volume is shown in Fig. 2. The optical and acoustic properties of the vessel structures were set to those of blood, and the multidomain simulations described above were conducted. Photoacoustic images were generated using the same three beamformers described above (i.e., FFT, DAS, and SLSC).

Fig. 2

Cross-sectional view of the simulated volumes with splayed vessel and claw structures within the imaging plane.

BIOS_2_1_012502_f002.png

3.

Results

3.1.

Photoacoustic Images of Small Targets

Figure 3 shows representative examples of photoacoustic images generated for simulations using 757 nm wavelength and FFT, DAS, and SLSC beamforming. With FFT beamforming [Fig. 3(a)], targets smaller than 1.75 mm are generally difficult to visualize regardless of skin tone. In addition, for very light (ITA=60  deg) and light (ITA=47  deg) skin tones, target visibility generally increases as the target size increases. However, the simulated target sizes cannot be visualized for intermediate (ITA=35  deg) to dark (ITA=54  deg) skin tones. With DAS beamforming [Fig. 3(b)], targets smaller than 1.75 mm are visible for very light and light skin tones but are difficult to visualize with darker skin tones (ITA35  deg) due to the presence of increasing levels of acoustic clutter. With SLSC beamforming [Fig. 3(c)], smaller targets (<1.75  mm) are visible for very light to intermediate skin tones. Although targets 2.5  mm can be visualized for the tan (ITA=20  deg) skin tone, target visualization is generally difficult for tan to dark skin tones.

Fig. 3

Photoacoustic images created with 757 nm wavelength, multiple skin tones, and (a) FFT, (b) DAS, and (c) SLSC beamforming. The white scale bar represents 5 mm. The skin signal does not appear uniformly across the top of these images because of the size of the light profile (i.e., 4 mm radius, as noted in Sec. 2).

BIOS_2_1_012502_f003.png

Figure 4 shows representative photoacoustic images using 800 nm wavelength and the same three beamformers noted above (i.e., FFT, DAS, and SLSC). With FFT beamforming [Fig. 4(a)], target sizes as small as 0.5 mm are visible with very light and light skin tones, target visualization is more difficult with the intermediate skin tone, and none of the simulated targets are visible with the tan to dark skin tones. With DAS beamforming [Fig. 4(b)], target sizes as small as 0.5 mm are visible for very light to intermediate skin tones, yet they are more difficult to visualize for tan to dark skin tones due to absorption at the skin surface and the associated skin photoacoustic signals and corresponding presence of acoustic clutter. With SLSC beamforming [Fig. 4(c)], the simulated targets are visible with very light to tan skin tones, yet they are difficult to visualize with brown (ITA=10  deg) and dark (ITA=54  deg) skin tones.

Fig. 4

Photoacoustic images created with 800 nm wavelength, multiple skin tones, and (a) FFT, (b) DAS, and (c) SLSC beamforming. The white scale bar represents 5 mm.

BIOS_2_1_012502_f004.png

Figure 5 shows representative examples of photoacoustic images using the 1064 nm wavelength and the same three beamformers noted above (i.e., FFT, DAS, and SLSC). With FFT beamforming [Fig. 5(a)], target visualization is generally poor. This is due to increased optical absorption in the initial water layer,25 which decreases light fluence into the tissue. With DAS beamforming [Fig. 5(b)], targets ranging from 0.5 to 3 mm can be partially visualized with the dark skin tone. Although the 2 to 2.5 mm targets were better visualized than the smaller or larger targets that were simulated with the very light to dark skin tones, larger targets (2.5 to 3 mm) are generally more visible than smaller targets (1.75  mm) for very light to brown skin tones due to acoustic clutter. With SLSC beamforming [Fig. 5(c)], target sizes ranging from 0.5 to 3 mm are visible with the entire range of simulated skin tones.

Fig. 5

Photoacoustic images created with 1064 nm wavelength, multiple skin tones, and (a) FFT, (b) DAS, and (c) SLSC beamforming. The white scale bar represents 5 mm.

BIOS_2_1_012502_f005.png

3.2.

Target Detectability Metrics

Figure 6 shows the mean ± one standard deviation of the SNR and gCNR as functions of the target size per skin tone category per beamforming method when imaging with the 757 nm wavelength. With FFT beamforming, the SNR and gCNR generally increase as the target size increases for very light to light skin tones. The SNR and gCNR are lower with intermediate to dark skin tones compared with the SNR and gCNR achieved with lighter skin tones. With DAS beamforming, although the gCNR of targets underlying intermediate to dark skin tones consistently increase as the target size increases, the SNR of targets underlying very light to light skin tones is greater than values achieved with the intermediate to dark skin tones. In this case, the differences in quantitative SNR results are more consistent with qualitative observations of target visibility (when compared with the corresponding gCNR results). With SLSC beamforming, the SNR and gCNR values generally increase as the target size increases in very light to tan skin tones, whereas the magnitude of the increase is lower for brown and dark skin tones compared with that of lighter skin tones. Therefore, very light to tan skin tones are more visible with smaller target sizes with SLSC beamforming.

Fig. 6

Mean ± one standard deviation of the SNR and gCNR of photoacoustic images created with 757 nm wavelength, per skin tone category, per beamforming method, displayed as functions of the target size.

BIOS_2_1_012502_f006.png

Figure 7 shows the mean ± one standard deviation of the SNR and gCNR as functions of the target size per skin tone category per beamforming method when imaging with an 800 nm wavelength. With FFT beamforming, the SNR and gCNR generally increase as the target size increases in very light to intermediate skin tones. However, the same rate of increase was not achievable with tan to dark skin tones, which also have lower SNR and gCNR values. With DAS beamforming, the SNR and gCNR of target sizes underlying very light to intermediate skin tones are greater than corresponding values achieved with tan to dark skin tones. With SLSC beamforming, the SNR and gCNR of target sizes underlying very light to tan skin tones are generally greater than values achieved with dark skin tones, whereas brown skin tones have inconsistent quantitative results when compared with the qualitative observations.

Fig. 7

Mean ± one standard deviation of the SNR and gCNR of photoacoustic images created with 800 nm wavelength, per skin tone category, per beamforming method, displayed as functions of the target size.

BIOS_2_1_012502_f007.png

Figure 8 shows the mean ± one standard deviation of the SNR and gCNR as functions of the target size per skin tone category for the three beamforming methods (FFT, DAS, and SLSC) when imaging with the 1064 nm wavelength. FFT beamforming produces the lowest SNR and gCNR for most target sizes, which is consistent with the qualitative observation of poor target visibility in corresponding images. With DAS and SLSC beamforming, the SNR is maximized with 2 to 2.5 mm target sizes imaged through very light to dark skin tones, with the greatest gCNR achieved with the 2.25-mm-diameter target. With SLSC beamforming, resulting images generally have the greatest gCNR of the three image formation methods for the simulated target sizes, and there is generally minimal variation among skin tone groups. These quantitative DAS and SLSC observations are consistent with qualitative observations.

Fig. 8

Mean ± one standard deviation of the SNR and gCNR of photoacoustic images created with 1064 nm wavelength, per skin tone category, per beamforming method, displayed as functions of the target size.

BIOS_2_1_012502_f008.png

3.3.

Splayed Vessel and Claw Feature Comparisons

Figure 9 shows a schematic diagram and representative photoacoustic image of a splayed vessel structure, created with a standard backprojection algorithm13 and 800 nm wavelength [Fig. 9(a)], along with FFT and SLSC photoacoustic images of the simulated splayed vessel structure imaged through very light [Fig. 9(b)] and dark [Fig. 9(c)] skin tones with 757, 800, and 1064 nm wavelengths. The green arrows point to a segment of the proximal vessel that is 0.5 mm in diameter. The corresponding target size simulations predict that this target size is difficult to visualize with FFT reconstruction, 757 and 1064 nm wavelengths, and very light skin tone (ITA=60deg) and similarly with FFT reconstruction, 1064 nm wavelength, and dark skin tone (ITA=54deg). In these three cases, the proximal vessel can be visualized with less clarity compared with the 800-nm FFT result obtained with the very light skin tone. By contrast, the target size simulations predict that 0.5-mm-diameter vessels can be visualized with (1) the very light skin tone, FFT reconstruction, and 800 nm wavelength; (2) the very light skin tone, SLSC beamforming, and 757 to 1064 nm wavelengths; and (3) the dark skin tone, SLSC beamforming, and 1064 nm wavelength. These three cases are consistent with the 0.5-mm-diameter proximal vessel visibility in Fig. 9. The distal vessel (0.5 to 0.75 mm diameter) is difficult to visualize with the FFT reconstruction, yet the blue arrows highlight better visualization of the 0.75-mm-diameter segment of this vessel with SLSC beamforming, using 757 to 1064 nm wavelengths with very light skin tone or a 1064 nm wavelength with dark skin tone (which is comparable to visualization with the very light skin tone).

Fig. 9

(a) Schematic diagram and corresponding in vivo photoacoustic image of a splayed vessel structure visualized with 800 nm wavelength (adapted from Abeyakoon et al.13). (b) FFT and SLSC photoacoustic images of a splayed vessel structure for very light (ITA=60  deg) and (c) dark (ITA=54  deg) skin tones, visualized with 757, 800, and 1064 nm wavelengths. The green arrows indicate a vessel component with a 0.5 mm diameter. The blue arrows indicate vessel components with a 0.75 mm diameter.

BIOS_2_1_012502_f009.png

Figure 10 shows a schematic diagram and representative photoacoustic image of a claw structure, created with a standard backprojection algorithm13 and 800 nm wavelength [Fig. 10(a)], along with FFT and SLSC photoacoustic images of the simulated claw structure for very light [Fig. 10(b)] and dark [Fig. 10(c)] skin tones, imaged with 757, 800, and 1064 nm wavelengths. The green and yellow arrows indicate components of the main vessel with diameters of 0.5 and 1.25 mm, respectively, which the corresponding target size FFT simulations predict are less visible with 757 to 1064 nm wavelengths and most simulated skin tones. With the very light skin tone (ITA=60  deg) and FFT reconstruction, the main vessel of the claw structure (diameter ranging from 0.5 to 1.25 mm and as small as 0.3 mm in the center) can be visualized with 757 and 800 nm wavelengths, yet is not visible with the 1064-nm wavelength. With SLSC beamforming, 757 to 1064 nm wavelengths, and the very light skin tone, the components of the claw structure highlighted by the green and yellow arrows are clearly visible. The blue arrows indicate areas where the branching vessels (diameter 0.4  mm) are partially visible. With the dark skin tone (ITA=54  deg), visualization of the claw structure is limited to SLSC beamforming and the 1064 nm wavelength, with comparable visualization to the very light skin tone.

Fig. 10

(a) Schematic diagram and corresponding in vivo photoacoustic image of a claw structure visualized with 800 nm wavelength (adapted from Abeyakoon et al.13). (b) FFT and SLSC photoacoustic images of a simulated claw structure visualized through very light (ITA = 60 deg) and (c) dark (ITA=54  deg) skin tones, using 757, 800, and 1064 nm wavelengths. The green and yellow arrows point to main vessel components with diameters of 0.5 and 1.25 mm, respectively. The blue arrows highlight a branching vessel component with a diameter 0.4  mm that is visible in SLSC images but not visible in corresponding FFT images.

BIOS_2_1_012502_f010.png

3.4.

Comparisons of Photoacoustic Images to Ground Truth Model Images

It may appear that the targets in the SLSC beamformed images are elongated in the axial direction relative to the amplitude-based images, particularly when viewing Figs. 3Fig. 45. To provide more context, Fig. 11 shows examples of model images from Fig. 1 overlaid on photoacoustic images from the minimum (i.e., 0.5 mm) and maximum (i.e., 3 mm) target sizes simulated with 800 and 1064 nm wavelengths and the very light and dark skin tones. In addition to being caused by the axial kernel length required to create SLSC images, the elongated appearance also occurs in the larger targets because the full target is visualized. In particular, the corresponding amplitude-based images (i.e., FFT and DAS images) of these larger targets primarily visualize the proximal region of the larger targets as the remainder of the target suffers from depth-dependent, amplitude-based fluence attenuation. It can also be appreciated from these images that the signal proximal to the target in most images originates from either the skin or water signal in Figs. 3Fig. 45.

Fig. 11

Photoacoustic images from the very light skin tone and dark skin tone results obtained with 800 and 1064 nm wavelengths (Figs. 4 and 5, respectively), overlaid on the corresponding ground truth images from Fig. 1.

BIOS_2_1_012502_f011.png

Finally, it may appear as if the photoacoustic images in Figs. 9 and 10 are spatially shifted relative to the simulation model. Therefore, Fig. 12 shows simulation model images from Fig. 2 overlaid on corresponding photoacoustic image examples from the 800 nm very light skin tone and the 1064 nm dark skin tone. As opposed to spatial shifts, the overlay format reveals artifacts surrounding the targets in the SLSC images compared with the FFT images, likely due to a combination of the axial correlation kernel length and the presence of coherent artifacts in the simulations. With SLSC beamforming, these types of artifacts are known to decrease with increasing levels of channel noise.54,58

Fig. 12

Photoacoustic images from the 800 nm very light skin tone and 1064 nm dark skin tone results in Figs. 9 and 10, overlaid on the corresponding ground truth images from Fig. 2.

BIOS_2_1_012502_f012.png

4.

Discussion

This paper is the first to qualitatively and quantitatively assess the impact of skin tone on target size detectability, followed by demonstrations of multiple combinations of optical wavelengths and beamforming methods that can be used to mitigate the associated skin tone bias, with three key insights. First, smaller target sizes (0.5 to 1.75 mm) are difficult to visualize in general and increasingly difficult to visualize with darker skin tones. Second, the 1064-nm wavelength relieves some of this difficulty when implementing DAS and SLSC beamforming (compared with results obtained with 757 and 800 nm wavelengths). Third, SLSC beamforming, which is a coherence-based approach, further alleviates this difficulty when compared with FFT reconstruction and DAS beamforming, which are both amplitude-based approaches. This insight is consistent with previous reports about SLSC photoacoustic imaging,25,3336 which provide foundational evidence that SLSC beamforming outperforms FFT reconstruction when imaging with 750, 810, and 870 nm wavelengths,34 and simulated breast imaging was previously demonstrated with 1064 nm and FFT imaging.25 With respect to these existing reports, the new and remarkable insight introduced herein is that the SLSC beamformer offers improved target detectability (i.e., gCNR) over FFT and DAS beamforming with a 1064 nm wavelength (Fig. 8).

In most cases, the qualitative observations of Figs. 3Fig. 45 are consistent with the quantitative results reported in Figs. 6Fig. 78, respectively. For example, the SLSC beamformer improved target detectability compared with the FFT reconstruction when photoacoustic images were acquired with 757 and 800 nm wavelengths in lighter skin tones [shown qualitatively in Figs. 3(c) and 4(c) and quantitatively in Figs. 6 and 7]. In addition, with 1064 nm and DAS and SLSC beamforming, the qualitative (Fig. 5) and quantitative (Fig. 8) variations among skin tones are minimal for each target size, likely due to the greater optical absorption in the water layer at 1064 nm, which decreases optical absorption by the skin and thereby decreases the magnitude of acoustic clutter caused by acoustic wave emission from the skin.21,22,34 Also, the gCNR values obtained with the 1064-nm wavelength are worse with FFT reconstruction compared with DAS and SLSC imaging in Fig. 8 (which corresponds to the poor visibility of targets with FFT reconstruction in [Fig. 5(a)]), and the acoustic clutter in DAS images of smaller targets [Fig. 5(b)] is reduced with SLSC beamforming [Fig. 5(c)]. Therefore, the combination of SLSC beamforming and the 1064 nm wavelength appears to be the best choice for small target visualization among varying skin tones.

There are a few inconsistencies with the general trends described above. For example, gCNR better supports the qualitative advantages achieved with SLSC beamforming and the 1064 nm wavelength, relative to the SNR, when comparing Figs. 5 and 8. However, the associated gCNR (and SNR) results in Fig. 8 do not support the reduced clutter observed with the dark skin tone DAS images in Fig. 5(b) (relative to the clutter present with lighter skin tones due to optical absorption of water at the 1064 nm wavelength). In addition, with tan to dark skin tones and DAS beamforming, gCNR increases as the target size increases when imaging with 757 nm (Fig. 6) and 800 nm (Fig. 7) wavelengths, although target visualization is qualitatively difficult in Figs. 3(b) and 4(b), respectively. Similarly, with 800 nm and SLSC beamforming [Fig. 4(c)], although the photoacoustic target is less bright when imaging targets underlying the tan skin tone (relative to lighter skin tones), the SNR obtained with the entire range of simulated target sizes and the gCNR obtained with targets 2.25  mm are greater with tan skin tones (Fig. 7).

Possible reasons for the above inconsistencies include known differences among the purpose of the SNR and gCNR image quality metrics55,56 and competing differences in the optical absorption of breast tissue39 (i.e., 0.004 to 0.005  mm1) relative to that of skin tissue32 at the 1064 nm wavelength (i.e., 0.0144  mm1 with the dark skin tone, as opposed to 0.0004 to 0.0035  mm1 with lighter skin tones, based on details in Table 2). The competing optical absorption of the water layer proximal to the skin tissue was also hypothesized to be responsible for the minimal skin absorption in the images created with a 1064 nm wavelength25 (i.e., water optical absorption of 0.0144  mm1 versus skin optical absorption ranging 0.0004 to 0.0144  mm1, which is within the range of previously reported values59,60). Additional contributing factors to the inconsistencies noted above include acoustic clutter (e.g., originating from skin, breast tissue, or water optical absorption) that increases the signal in the target ROIs relative to the background ROIs even though no visible target is present, as well as artifacts from the target that increase the overall signal and variance in the background ROI. However, none of these possible reasons for inconsistencies impact the generally optimal combination of SLSC beamforming and 1064 nm wavelength noted above.

It is promising that the benefits observed with 1064 nm and SLSC beamforming translated well from small targets to the splayed vessel (Fig. 9) and claw (Fig. 10) features. In particular, visualization of vessel components as small as 0.5 mm (Fig. 9, green arrows), 0.75 mm (Fig. 9, blue arrows), or 1.25 mm (Fig. 10, green arrows) aligned with target detectability predictions from the corresponding target sizes in Figs. 3Fig. 4Fig. 5Fig. 6Fig. 78. In some cases, vessel components with diameters <0.5  mm can be visualized (e.g., blue arrows in Fig. 10), which is promising as these target sizes were not individually simulated in Figs. 3Fig. 4Fig. 5Fig. 6Fig. 78. It is additionally promising that, with the 1064-nm wavelength and SLSC beamformer, visualization of the vessel structures underlying the dark skin tone (Figs. 9 and 10) were each comparable to visualization with the very light skin tone.

There are no details about patient skin tone reported by Abeyakoon et al.13 when describing the source of the in vivo photoacoustic images in Figs. 9 and 10. Therefore, direct comparisons with the simulated 800 nm images for very light and dark skin tones are not possible. In addition to differences between the backprojection algorithm used to generate the in vivo images and the three image reconstruction methods used in our simulations (FFT, DAS, and SLSC), there are differences in the light profile and transducer parameters used to generate the in vivo images compared with the simulations conducted in this study. There are also likely differences in the level of channel noise, the components of the vessel structures contained within the imaging plane, and post-processing filters that may have been applied to the in vivo images after the beamforming step (which our images do not implement). As the literature on in vivo photoacoustic images of blood vessels in breast tissue is relatively limited, our goal in simulating the splayed vessel and claw structures was to investigate if our findings regarding SLSC beamforming with the spherical targets translated to more anatomically relevant structures. It is promising that the 1064-nm dark skin tone SLSC photoacoustic images in Figs. 9 and 10 are comparable to the very light skin tone SLSC images created with the 800-nm wavelength (as well as with 757 and 1064 nm wavelengths) and comparable to the in vivo photoacoustic images in Figs. 9 and 10, which further supports the use of 1064 nm and SLSC beamforming in a wide range of skin tones. Overall, this comparison to clinical data highlights the importance of including skin tone when reporting results and demonstrates the potential of SLSC beamforming for improving the visualization of blood vessels relevant to breast cancer diagnosis in patients with darker skin tones.

In comparison with previously published in vivo forearm results, wherein SLSC beamforming enabled visualization of in vivo radial arteries in dark skin tones (ITA range from 53.74deg to 33.57deg) with 750, 810, and 870 nm wavelengths,34 our simulated results show that targets are not visible with 757 and 800 nm wavelengths and the dark skin tone (ITA range from 54deg to 33deg). Given the similar optical properties between 750 and 757 nm wavelengths and between 800 and 810 nm wavelengths,20,32,39,47 this discrepancy with the previous report34 is likely due to differences in tissue types (e.g., primarily skin and muscle tissue in the forearm as opposed to primarily fatty and fibroglandular tissue in the breast model) and associated differences in optical and acoustic properties.

Although our simulations included multiple tissue components that are known to have varying Grüneisen parameters (e.g., 0.81, 0.16, and 0.12 for subcutaneous fat, blood, and water, respectively61), we assumed a constant value across varying tissue types, which could potentially be perceived as a study limitation. However, acceptable values for the Grüneisen parameter of other components of our simulations (e.g., skin and fibroglandular tissue) are not well documented in the literature. The Grüneisen parameter is also temperature-dependent,61 has been reported to vary within the same tissue type,62,63 and was previously assumed to be constant in photoacoustic simulations.39,64,65 As this parameter scales the initial pressure distribution,66 incorporating chromophore-specific values is expected to increase the signal from fat tissue relative to components with lower Grüneisen parameters, which could potentially decrease acoustic clutter from skin yet introduce additional clutter from fat.

One limitation of our work is that three wavelengths were investigated, whereas a wider range of wavelengths can possibly be deployed. However, the three investigated wavelengths (i.e., 757, 800, and 1064 nm) correspond to wavelengths utilized in existing clinical photoacoustic imaging systems.3,13,15,46 In addition, a wider range of target sizes can potentially be investigated in the future. However, the range of target sizes simulated in this study was based on measurements of diagnostic structures from in vivo breast photoacoustic images1317 and translated well to comparisons with the in vivo photoacoustic images in Figs. 9 and 10. Target depth is another variable that can be investigated, although previous work22,25,67,68 demonstrated that skin tone influences target detectability at multiple target depths with amplitude-based image reconstruction methods. In particular, deeper targets were difficult to visualize in vivo with amplitude-based image reconstruction methods,22,25 this difficulty is further compounded with darker skin tones,22,25 and this compounded difficulty persists as a function of wavelength in controlled phantom67 and simulation68 experiments. The range of target depths previously investigated includes the target depth of 10 mm from the transducer surface, which was the fixed depth investigated in our study. The coherence-based SLSC beamformer is also known to visualize deeper targets than otherwise possible with amplitude-based beamformers in phantoms36,38,53 and in vivo.37 Considering this previous work, we focused our attention on other important parameter combinations—i.e., target size, wavelength, and image reconstruction methods. The results in this paper, when combined with previous findings, add new insights into the detrimental effects of (and potential remedies to) skin tone impacting multiple imaging parameters of interest.

Future testing with phantoms can incorporate synthetic melanosomes that will closely match the absorption coefficient and reduced scattering coefficient of skin.69 With a real photoacoustic imaging system, more noise is expected when using low-energy diodes (e.g., light emitting diodes,22 pulsed laser diodes,38 and continuous wave laser diodes70), rather than higher-energy lasers. Although MCXLAB does not support the simulation of specific laser sources or laser energies,40 diode lasers are typically commercially available at wavelengths greater than 750 nm,71 which aligns with the simulated wavelengths. However, diode sources require extensive signal averaging to compensate for the lower illumination energy, which was previously shown to be resolvable with SLSC beamforming38 (assuming the magnitudes of acoustic waves generated after tissue expansion are greater than the noise-equivalent pressure of the transducer72). In addition, as the performance of SLSC beamforming is influenced by the presence of channel noise,33,54,58 future work will investigate the impact of different levels of channel noise on the detectability of targets underlying various skin tones and the relationship to ground truth target sizes.

5.

Conclusion

The work presented in this paper provides simulation-based evidence that there is a skin tone bias that limits the detectability of small targets (i.e., 0.5 to 3 mm in diameter) underlying darker skin tones when using traditional amplitude-based photoacoustic image reconstruction methods (e.g., FFT and DAS). With SLSC beamforming and transmitted wavelengths of 757 and 800 nm, target visibility remained sufficient with very light to light skin tones (ITA range from 43 deg to 60 deg) and improved with intermediate to tan skin tones (ITA range from 12 deg to 38 deg), relative to amplitude-based FFT and DAS imaging. Adding to these benefits, SLSC beamforming further mitigated the skin tone bias present with brown and dark skin tones (ITA range from –54 deg to 5 deg) when using the 1064 nm wavelength, as demonstrated qualitatively with improved photoacoustic target visibility and quantitatively with increased SNR and gCNR values relative to FFT and DAS images. These findings were further supported when comparing FFT and SLSC images of two realistic vessel structures, as the SLSC images created with a dark skin tone and with a 1064-nm wavelength were comparable to those of a very light skin tone. Differences among skin tone groups were also minimal with 1064 nm per target size per image reconstruction method. This work highlights the promise of the 1064 nm wavelength and SLSC beamforming to address challenges with detecting small targets across a wide range of skin tones. Upon implementation, these parameter choices are expected to improve future designs of next-generation photoacoustic imaging systems.

Disclosures

The authors declare no competing interests.

Code and Data Availability

Tabulated data to recreate figures that support the major findings of this study are publicly available at https://gitlab.com/pulselab/skin_tone_simulations.73 These data were derived from the following software, data, and code available in the public domain:

  • Breast phantom database39

  • k-Wave MATLAB toolbox48

Additional data and code related to the paper will be publicly available whenever possible utilizing the above PULSE Lab GitLab repository.73

Acknowledgments

This work is supported by the Chan Zuckerberg Initiative DAF (an advised fund of the Silicon Valley Community Foundation) (Grant No. 2022-309513), the National Institutes of Health (NIH) (Grant No. R01 EB032960), and the National Science Foundation (NSF) Alan T. Waterman Award (Grant No. IIS-2431810). This work was also completed in partnership with the NSF Computational Sensing and Medical Robotics Research Experience for Undergraduates program (Grant No. EEC-1852155).

References

1. 

N. Nyayapathi and J. Xia, “Photoacoustic imaging of breast cancer: a mini review of system design and image features,” J. Biomed. Opt., 24 (12), 121911 https://doi.org/10.1117/1.JBO.24.12.121911 JBOPFO 1083-3668 (2019). Google Scholar

2. 

J. Xia, J. Yao and L. V. Wang, “Photoacoustic tomography: principles and advances,” Electromagn. Waves, 147 1 –22 https://doi.org/10.2528/PIER14032303 (2014). Google Scholar

3. 

S. Manohar and M. Dantuma, “Current and future trends in photoacoustic breast imaging,” Photoacoustics, 16 100134 https://doi.org/10.1016/j.pacs.2019.04.004 (2019). Google Scholar

4. 

Y. Wang et al., “Review of methods to improve the performance of linear array-based photoacoustic tomography,” J. Innov. Opt. Health Sci., 13 (02), 2030003 https://doi.org/10.1142/S1793545820300037 (2020). Google Scholar

5. 

R. W. Pinsky and M. A. Helvie, “Mammographic breast density: effect on imaging and breast cancer risk,” J. Natl. Compr. Cancer Netw., 8 (10), 1157 –1165 https://doi.org/10.6004/jnccn.2010.0085 (2010). Google Scholar

6. 

P. E. Freer, “Mammographic breast density: impact on breast cancer risk and implications for screening,” Radiographics, 35 (2), 302 –315 https://doi.org/10.1148/rg.352140106 (2015). Google Scholar

7. 

A. Wiacek et al., “Coherence-based beamforming increases the diagnostic certainty of distinguishing fluid from solid masses in breast ultrasound exams,” Ultrasound Med. Biol., 46 (6), 1380 –1394 https://doi.org/10.1016/j.ultrasmedbio.2020.01.016 USMBA3 0301-5629 (2020). Google Scholar

8. 

I. W. Pan, K. C. Oeffinger and Y. T. Shih, “Cost-sharing and out-of-pocket cost for women who received MRI for breast cancer screening,” J. Natl. Cancer Inst., 114 (2), 254 –262 https://doi.org/10.1093/jnci/djab150 JNCIEQ (2022). Google Scholar

9. 

L. V. Wang and H.-I. Wu, Biomedical optics: principles and imaging, John Wiley & Sons( (2007). Google Scholar

10. 

D. R. Bielenberg and B. R. Zetter, “The contribution of angiogenesis to the process of metastasis,” Cancer J., 21 (4), 267 –273 https://doi.org/10.1097/PPO.0000000000000138 (2015). Google Scholar

11. 

Y. Zhang et al., “Hypoxia in breast cancer—scientific translation to therapeutic and diagnostic clinical applications,” Front. Oncol., 11 652266 https://doi.org/10.3389/fonc.2021.652266 FRTOA7 0071-9676 (2021). Google Scholar

12. 

Q. Zhu et al., “Benign versus malignant breast masses: optical differentiation with us-guided optical imaging reconstruction,” Radiology, 237 (1), 57 –66 https://doi.org/10.1148/radiol.2371041236 RADLAX 0033-8419 (2005). Google Scholar

13. 

O. Abeyakoon et al., “An optoacoustic imaging feature set to characterise blood vessels surrounding benign and malignant breast lesions,” Photoacoustics, 27 100383 https://doi.org/10.1016/j.pacs.2022.100383 (2022). Google Scholar

14. 

G. Xi et al., “Label-free imaging of blood vessels in human normal breast and breast tumor tissue using multiphoton microscopy,” Scanning, 2019 5192875 https://doi.org/10.1155/2019/5192875 SCNNDF 0161-0457 (2019). Google Scholar

15. 

A. A. Oraevsky et al., “Clinical optoacoustic imaging combined with ultrasound for coregistered functional and anatomical mapping of breast tumors,” Photoacoustics, 12 30 –45 https://doi.org/10.1016/j.pacs.2018.08.003 (2018). Google Scholar

16. 

E. I. Neuschler et al., “A pivotal study of optoacoustic imaging to diagnose benign and malignant breast masses: a new evaluation tool for radiologists,” Radiology, 287 (2), 398 –412 https://doi.org/10.1148/radiol.2017172228 RADLAX 0033-8419 (2017). Google Scholar

17. 

I. Yamaga et al., “Vascular branching point counts using photoacoustic imaging in the superficial layer of the breast: a potential biomarker for breast cancer,” Photoacoustics, 11 6 –13 https://doi.org/10.1016/j.pacs.2018.06.002 (2018). Google Scholar

18. 

J. H. Uliana et al., “Multiangle long-axis lateral illumination photoacoustic imaging using linear array transducer,” Sensors, 20 (14), 4052 https://doi.org/10.3390/s20144052 SNSRES 0746-9462 (2020). Google Scholar

19. 

A. N. Bashkatov et al., “Optical properties of human skin, subcutaneous and mucous tissues in the wavelength range from 400 to 2000 nm,” J. Phys. D: Appl. Phys., 38 (15), 2543 https://doi.org/10.1088/0022-3727/38/15/004 JPAPBE 0022-3727 (2005). Google Scholar

20. 

S. L. Jacques, “Optical properties of biological tissues: a review,” Phys. Med. Biol., 58 (11), R37 –R61 https://doi.org/10.1088/0031-9155/58/11/R37 PHMBA7 0031-9155 (2013). Google Scholar

21. 

S. Presser et al., “Study of clutter origin in in-vivo epi-optoacoustic imaging of human forearms,” J. Opt., 18 (9), 094003 https://doi.org/10.1088/2040-8978/18/9/094003 (2016). Google Scholar

22. 

Y. Mantri and J. V. Jokerst, “Impact of skin tone on photoacoustic oximetry and tools to minimize bias,” Biomed. Opt. Express, 13 (2), 875 –887 https://doi.org/10.1364/BOE.450224 BOEICL 2156-7085 (2022). Google Scholar

23. 

T. R. Else et al., “Effects of skin tone on photoacoustic imaging and oximetry,” J. Biomed. Opt., 29 (S1), S11506 https://doi.org/10.1117/1.JBO.29.S1.S11506 JBOPFO 1083-3668 (2024). Google Scholar

24. 

G. S. P. Fernandes et al., “Impact of skin pigmentation on photoacoustic imaging using linear array transducer: a pilot in vivo study,” in IEEE Int. Ultrason. Symp. (IUS), 1 –4 (2022). https://doi.org/10.1109/IUS54386.2022.9958826 Google Scholar

25. 

G. S. P. Fernandes, T. Z. Pavan and M. A. L. Bell, “In silico demonstrations of the impact of wavelength and skin tone on photoacoustic breast imaging,” in IEEE Int. Ultrason. Symp. (IUS), 1 –4 (2023). https://doi.org/10.1109/IUS51837.2023.10307088 Google Scholar

26. 

T. B. Fitzpatrick, “The validity and practicality of sun-reactive skin types I through VI,” Arch. Dermatol., 124 (6), 869 –871 https://doi.org/10.1001/archderm.1988.01670060015008 (1988). Google Scholar

27. 

W. Coleman, K. Mariwalla and P. Grimes, “Updating the Fitzpatrick classification: the skin color and ethnicity scale,” Dermatol. Surg., 49 (8), 725 –731 https://doi.org/10.1097/DSS.0000000000003860 (2023). Google Scholar

28. 

M. Osto et al., “Individual typology angle and Fitzpatrick skin phototypes are not equivalent in photodermatology,” Photochem. Photobiol., 98 (1), 127 –129 https://doi.org/10.1111/php.13562 PHCBAP 0031-8655 (2022). Google Scholar

29. 

B. C. K. Ly et al., “Research techniques made simple: cutaneous colorimetry: a reliable technique for objective skin color measurement,” J. Invest. Dermatol., 140 (1), 3 –12.e1 https://doi.org/10.1016/j.jid.2019.11.003 JIDEAE 0022-202X (2020). Google Scholar

30. 

A. M. Pena et al., “In vivo melanin 3D quantification and z-epidermal distribution by multiphoton FLIM, phasor and pseudo-FLIM analyses,” Sci. Rep., 12 (1), 1642 https://doi.org/10.1038/s41598-021-03114-0 SRCEC3 2045-2322 (2022). Google Scholar

31. 

S. Del Bino et al., “Chemical analysis of constitutive pigmentation of human epidermis reveals constant eumelanin to pheomelanin ratio,” Pigment Cell Melanoma Res., 28 (6), 707 –717 https://doi.org/10.1111/pcmr.12410 (2015). Google Scholar

32. 

C. Junior et al., “Absorption and reduced scattering coefficient estimation in pigmented human skin tissue by experimental colorimetric fitting,” J. Opt. Soc. Am. A, Opt., Image. Sci., Vis., 40 (9), 1680 –1685 https://doi.org/10.1364/JOSAA.489892 (2023). Google Scholar

33. 

M. T. Graham and M. A. L. Bell, “Photoacoustic spatial coherence theory and applications to coherence-based image contrast and resolution,” IEEE Trans. Ultrason. Ferroelectr. Freq. Control., 67 (10), 2069 –2084 https://doi.org/10.1109/TUFFC.2020.2999343 ITUCER 0885-3010 (2020). Google Scholar

34. 

G. S. P. Fernandes et al., “Mitigating skin tone bias in linear array in vivo photoacoustic imaging with short-lag spatial coherence beamforming,” Photoacoustics, 33 100555 https://doi.org/10.1016/j.pacs.2023.100555 (2023). Google Scholar

35. 

M. A. Lediju et al., “Short-lag spatial coherence of backscattered echoes: imaging characteristics,” IEEE Trans. Ultrason. Ferroelectr. Freq. Control., 58 1377 –1388 https://doi.org/10.1109/TUFFC.2011.1957 ITUCER 0885-3010 (2011). Google Scholar

36. 

M. A. Lediju Bell et al., “Short-lag spatial coherence beamforming of photoacoustic images for enhanced visualization of prostate brachytherapy seeds,” Biomed. Opt. Express, 4 (10), 1964 –1977 https://doi.org/10.1364/BOE.4.001964 BOEICL 2156-7085 (2013). Google Scholar

37. 

M. A. Lediju Bell et al., “In vivo visualization of prostate brachytherapy seeds with photoacoustic imaging,” J. Biomed. Opt., 19 (12), 126011 https://doi.org/10.1117/1.JBO.19.12.126011 JBOPFO 1083-3668 (2014). Google Scholar

38. 

M. A. Lediju Bell et al., “Improved contrast in laser-diode-based photoacoustic images with short-lag spatial coherence beamforming,” in IEEE Int. Ultrason. Symp., 37 –40 (2014). https://doi.org/10.1109/ULTSYM.2014.0010 Google Scholar

39. 

Y. Lou et al., “Generation of anatomically realistic numerical phantoms for photoacoustic and ultrasonic breast imaging,” J. Biomed Opt., 22 (4), 041015 https://doi.org/10.1117/1.JBO.22.4.041015 (2017). Google Scholar

40. 

Q. Fang and D. A. Boas, “Monte carlo simulation of photon migration in 3D turbid media accelerated by graphics processing units,” Opt. Express, 17 (22), 20178 –20190 https://doi.org/10.1364/OE.17.020178 OPEXFF 1094-4087 (2009). Google Scholar

41. 

A. Chardon, I. Cretois and C. Hourseau, “Skin colour typology and suntanning pathways,” Int. J. Cosmet. Sci., 13 (4), 191 –208 https://doi.org/10.1111/j.1467-2494.1991.tb00561.x IJCMDW 0142-5463 (1991). Google Scholar

42. 

S. Del Bino et al., “Relationship between skin response to ultraviolet exposure and skin color type,” Pigment Cell Res., 19 (6), 606 –614 https://doi.org/10.1111/j.1600-0749.2006.00338.x PCREEA 0893-5785 (2006). Google Scholar

43. 

I. Hurbain et al., “Melanosome distribution in keratinocytes in different skin types: melanosome clusters are not degradative organelles,” J. Invest. Dermatol., 138 (3), 647 –656 https://doi.org/10.1016/j.jid.2017.09.039 JIDEAE 0022-202X (2018). Google Scholar

44. 

G. N. Stamatas et al., “Non-invasive measurements of skin pigmentation in situ,” Pigment Cell Res., 17 (6), 618 –626 https://doi.org/10.1111/j.1600-0749.2004.00204.x PCREEA 0893-5785 (2004). Google Scholar

45. 

S. Vasudevan et al., “Melanometry for objective evaluation of skin pigmentation in pulse oximetry studies,” Commun. Med., 4 (1), 138 https://doi.org/10.1038/s43856-024-00550-7 (2024). Google Scholar

46. 

I. Steinberg et al., “Photoacoustic clinical imaging,” Photoacoustics, 14 77 –98 https://doi.org/10.1016/j.pacs.2019.05.001 (2019). Google Scholar

47. 

S. Prahl, “Assorted spectra,” https://omlc.org/spectra/index.html (). Google Scholar

48. 

B. E. Treeby and B. T. Cox, “k-wave: MATLAB toolbox for the simulation and reconstruction of photoacoustic wave fields,” J. Biomed. Opt., 15 (2), 021314 https://doi.org/10.1117/1.3360308 JBOPFO 1083-3668 (2010). Google Scholar

49. 

W. Shu et al., “Broadening the detection view of 2D photoacoustic tomography using two linear array transducers,” Opt. Express, 24 (12), 12755 –12768 https://doi.org/10.1364/OE.24.012755 OPEXFF 1094-4087 (2016). Google Scholar

50. 

G. R. Kim et al., “Photoacoustic imaging of breast microcalcifications: a preliminary study with 8-gauge core-biopsied breast specimens,” PloS One, 9 (8), e105878 https://doi.org/10.1371/journal.pone.0105878 POLNCL 1932-6203 (2014). Google Scholar

51. 

M. T. Graham and M. A. L. Bell, “PhocoSpace: an open-source simulation package to implement photoacoustic spatial coherence theory,” in IEEE Int. Ultrason. Symp. (IUS), 1 –5 (2022). https://doi.org/10.1109/IUS54386.2022.9958309 Google Scholar

52. 

M. T. Graham and M. A. L. Bell, “PhocoSpace,” https://gitlab.com/pulselab/phocospace (2022). Google Scholar

53. 

M. A. Lediju Bell et al., “Transurethral light delivery for prostate photoacoustic imaging,” J. Biomed. Opt., 20 (3), 036002 https://doi.org/10.1117/1.JBO.20.3.036002 JBOPFO 1083-3668 (2015). Google Scholar

54. 

M. A. Lediju Bell, J. J. Dahl and G. E. Trahey, “Resolution and brightness characteristics of short-lag spatial coherence (SLSC) images,” IEEE Trans. Ultrason. Ferroelectr. Freq. Control, 62 (7), 1265 –1276 https://doi.org/10.1109/TUFFC.2014.006909 ITUCER 0885-3010 (2015). Google Scholar

55. 

A. Rodriguez-Molares et al., “The generalized contrast-to-noise ratio: a formal definition for lesion detectability,” IEEE Trans. Ultrason. Ferroelectr. Freq. Control, 67 (4), 745 –759 https://doi.org/10.1109/TUFFC.2019.2956855 ITUCER 0885-3010 (2020). Google Scholar

56. 

K. M. Kempski et al., “Application of the generalized contrast-to-noise ratio to assess photoacoustic image quality,” Biomed. Opt. Express, 11 (7), 3684 –3698 https://doi.org/10.1364/BOE.391026 BOEICL 2156-7085 (2020). Google Scholar

57. 

M. R. Gubbi, E. A. Gonzalez and M. A. L. Bell, “Theoretical framework to predict generalized contrast-to-noise ratios of photoacoustic images with applications to computer vision,” IEEE Trans. Ultrason. Ferroelectr. Freq. Control, 69 (6), 2098 –2114 https://doi.org/10.1109/TUFFC.2022.3169082 ITUCER 0885-3010 (2022). Google Scholar

58. 

B. Stephanian et al., “Additive noise models for photoacoustic spatial coherence theory,” Biomed. Opt. Express, 9 (11), 5566 –5582 https://doi.org/10.1364/BOE.9.005566 BOEICL 2156-7085 (2018). Google Scholar

59. 

T. Kono and J. Yamada, “In vivo measurement of optical properties of human skin for 450–800 nm and 950–1600 nm wavelengths,” Int. J. Thermophys., 40 1 –14 https://doi.org/10.1007/s10765-019-2515-3 IJTHDY 0195-928X (2019). Google Scholar

60. 

E. Salomatina et al., “Optical properties of normal and cancerous human skin in the visible and near-infrared spectral range,” J. Biomed. Opt., 11 (6), 064026 https://doi.org/10.1117/1.2398928 JBOPFO 1083-3668 (2006). Google Scholar

61. 

D. K. Yao et al., “Photoacoustic measurement of the grüneisen parameter of tissue,” J. Biomed. Opt., 19 (1), 017007 https://doi.org/10.1117/1.JBO.19.1.017007 JBOPFO 1083-3668 (2014). Google Scholar

62. 

B. T. Cox, J. G. Laufer and P. C. Beard, “The challenges for quantitative photoacoustic imaging,” Proc. SPIE, 7177 131 –139 https://doi.org/10.1117/12.806788 PSISDG 0277-786X (2009). Google Scholar

63. 

E. V. Savateeva et al., “Optical properties of blood at various levels of oxygenation studied by time-resolved detection of laser-induced pressure profiles,” Proc. SPIE, 4618 63 –75 https://doi.org/10.1117/12.469849 PSISDG 0277-786X (2002). Google Scholar

64. 

S. Agrawal et al., “Modeling combined ultrasound and photoacoustic imaging: simulations aiding device development and artificial intelligence,” Photoacoustics, 24 100304 https://doi.org/10.1016/j.pacs.2021.100304 (2021). Google Scholar

65. 

S. Park et al., “Stochastic three-dimensional numerical phantoms to enable computational studies in quantitative optoacoustic computed tomography of breast cancer,” J. Biomed. Opt., 28 (6), 066002 https://doi.org/10.1117/1.JBO.28.6.066002 JBOPFO 1083-3668 (2023). Google Scholar

66. 

C. Bench, A. Hauptmann and B. Cox, “Toward accurate quantitative photoacoustic imaging: learning vascular blood oxygen saturation in three dimensions,” J. Biomed. Opt., 25 (8), 085003 https://doi.org/10.1117/1.JBO.25.8.085003 JBOPFO 1083-3668 (2020). Google Scholar

67. 

W. C. Vogt, K. A. Wear and T. J. Pfefer, “Phantoms for evaluating the impact of skin pigmentation on photoacoustic imaging and oximetry performance,” Biomed. Opt. Express, 14 (11), 5735 –5748 https://doi.org/10.1364/BOE.501950 BOEICL 2156-7085 (2023). Google Scholar

68. 

S. Park et al., “Numerical investigation of impact of skin phototype on three-dimensional optoacoustic tomography of the breast,” Proc. SPIE, PC12379 PC123790E https://doi.org/10.1117/12.2650740 PSISDG 0277-786X (2023). Google Scholar

69. 

W. Yim et al., “3D-bioprinted phantom with human skin phototypes for biomedical optics,” Adv. Mater., 35 (3), 2206385 https://doi.org/10.1002/adma.202206385 ADVMEW 0935-9648 (2023). Google Scholar

70. 

X. Li et al., “High-speed high-resolution laser diode-based photoacoustic microscopy for in vivo microvasculature imaging,” Visual Comput. Ind. Biomed. Art, 4 1 –6 https://doi.org/10.1186/s42492-020-00067-5 (2021). Google Scholar

71. 

T. J. Allen and P. C. Beard, “High power visible light emitting diodes as pulsed excitation sources for biomedical photoacoustics,” Biomed. Opt. Express, 7 (4), 1260 –1270 https://doi.org/10.1364/BOE.7.001260 BOEICL 2156-7085 (2016). Google Scholar

72. 

Y. Zhu et al., “Light emitting diodes based photoacoustic imaging and potential clinical applications,” Sci. Rep., 8 (1), 1 –12 https://doi.org/10.1038/s41598-018-28131-4 SRCEC3 2045-2322 (2018). Google Scholar

73. 

M. A. L. Bell, “PULSE Lab public repositories,” https://gitlab.com/pulselab (). Google Scholar

Biography

Rhea D. Rasquinha is an undergraduate student majoring in biomedical engineering at Brown University. She completed the NSF Computational Sensing and Medical Robotics Research Experience for Undergraduates program in the Photoacoustic and Ultrasonic Systems Engineering (PULSE) Lab at Johns Hopkins University during the summer of 2024, where she conducted research on photoacoustic image reconstruction methods to mitigate skin-tone-based differences in small-target detectability with applications to breast cancer diagnosis and screening.

Mardava R. Gubbi is pursuing a PhD in electrical and computer engineering at Johns Hopkins University. He received his MTech degree in wireless communications and signal processing and his BTech degree in electrical engineering from the Indian Institute of Technology Madras, Chennai, India, in 2015. His research interests include photoacoustic imaging, robotics, and machine learning with a view to surgical applications.

Muyinatu A. Lediju Bell is the John C. Malone Associate Professor of electrical and computer engineering, biomedical engineering, and computer science at Johns Hopkins University, where she founded and directs the Photoacoustic and Ultrasonic Systems Engineering (PULSE) Lab. She received her PhD in biomedical engineering from Duke University in 2012, completed research abroad as a Whitaker International Fellow in 2010, and earned her SB degree in mechanical engineering (biomedical engineering minor) from Massachusetts Institute of Technology in 2006. Her research interests revolve around ultrasound and photoacoustic imaging, photoacoustic-guided surgery, robot-assisted imaging, and the design of innovative medical imaging technologies that will improve the standard of patient care. She has received numerous awards and honors, including the SPIE Early Career Achievement Award in 2021. She is a fellow of SPIE, Optica, and AIMBE.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 International License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Rhea D. Rasquinha, Mardava R. Gubbi, and Muyinatu A. Lediju Bell "Impact of skin tone on target size detectability in photoacoustic breast imaging," Biophotonics Discovery 2(1), 012502 (14 November 2024). https://doi.org/10.1117/1.BIOS.2.1.012502
Received: 2 August 2024; Accepted: 24 September 2024; Published: 14 November 2024
Advertisement
Advertisement
KEYWORDS
Skin

Target detection

Spatial filtering

Photoacoustic spectroscopy

Visualization

Signal to noise ratio

Tissue optics

Back to Top