1 September 2006 Molecular contrast optical coherence tomography: a pump-probe scheme using indocyanine green as a contrast agent
Author Affiliations +
Abstract
The use of indocyanine green (ICG), a U.S. Food and Drug Administration approved dye, in a pump-probe scheme for molecular contrast optical coherence tomography (MCOCT) is proposed and demonstrated for the first time. In the proposed pump-probe scheme, an optical coherence tomography (OCT) scan of the sample containing ICG is first acquired. High fluence illumination (~190 kJ/cm2) is then used to permanently photobleach the ICG molecules — resulting in a permanent alteration of the overall absorption of the ICG. A second OCT scan is next acquired. The difference of the two OCT scans is used to determine the depth resolved distribution of ICG within a sample. To characterize the extent of photobleaching in different ICG solutions, we determine the cumulative probability of photobleaching, ΦB,cum, defined as the ratio of the total photobleached ICG molecules to the total photons absorbed by the ground state molecules. An empirical study of ICG photobleaching dynamics shows that ΦB,cum decreases with fluence as well as with increasing dye concentration. The quantity ΦB,cum is useful for estimating the extent of photobleaching in an ICG sample (MCOCT contrast) for a given fluence of the pump illumination. The paper also demonstrates ICG-based MCOCT imaging in tissue phantoms as well as within stage 54 Xenopus laevis.

1.

Introduction

Optical coherence tomography1 (OCT) is an important noninvasive biomedical tool for high resolution2 imaging of biological samples to a depth of a few millimeters. Since its inception, OCT has matured into a useful diagnostic tool for ophthalmic applications. 3, 4, 5, 6 OCT’s applications for imaging other areas of the human body include endoscopic imaging of the gastrointestinal (GI) tract7 and bladder8 for tumor detection, monitoring and risk assessment of vulnerable lipid plaques in the vascular system,9 monitoring intracoronary stenting,10 and tracking structural weakness associated with tooth decay.11 While basic OCT imaging is able to render depth resolved structural images of the target, more sophisticated OCT imaging strategies can provide additional functional information such as flow (through Doppler OCT),12 tissue structural arrangement (through birefringence OCT),13 and depth resolved spectral signatures (through spectral OCT).14, 15 In recent years, a new and exciting functional OCT method, known as molecular contrast optical coherence tomography (MCOCT), 16, 17, 18, 19, 20, 21, 22, 23, 24 has been introduced that combines the major advantages of fluorescence microscopy (chemical or biological contrast agent’s detection and imaging capability) and OCT (higher spatial resolution and depth penetration).

In this paper, we explore the use of indocyanine green (ICG), which is used in medical diagnostics25 and photodynamic therapy,26 in a MCOCT scheme based on the photobleaching mechanism of the dye. This work differs in its contrast mechanism from an earlier work with ICG in a MCOCT scheme, where the contrast agent’s distribution inside the sample was determined from agent’s spectral differential absorption.21

Aqueous solutions of ICG degrade over time; the process is accelerated in the presence of light with spectral band that is coincident with the ICG optical absorption peak. This degradation is attributable to the photobleaching of the ICG. In addition to the ground and first excited states, ICG has three states of interest, namely the triplet (T) , P -isomer (P) , and bleached (B) states, in which the molecules can reside (see Fig. 1 ). An excited molecule will generally relax directly back into the ground state, but it does have a small probability of transiting into any of these states. The quantum yields ϕT , ϕP , and ϕB of the three states are very low ( 105 to 106 ).27 The pump-probe MCOCT scheme makes use of the fact that molecules that are temporarily shelved (for the case of triplet and P -isomer states) or permanently bleached (for the case of bleached state) have a different absorption spectrum from ground state molecules to elicit contrast. We note that relatively short lifetimes associated with the triplet and P -isomer states (triplet state lifetime, τT700μs and P -isomer state lifetime, τP10ms )27 will require elaborate pump-probe schemes for contrast data acquisition. In comparison, the permanency of the photobleached state is appealing in that given enough time, all the molecules in an illuminated sample of ICG solution can be transitioned into the bleached state; there is no stringent pump intensity requirements or fast signal acquisition scheme requirements in this case.

Fig. 1

Schematic of electronic state diagram of ICG. σG is the ground state absorption cross section, whereas ϕT , ϕP , and ϕB are quantum yields of triplet (T) , P -isomer (P) , and bleached (B) states.

054017_1_021605jbo1.jpg

The implementation scheme based on the use of photobleaching can be summarized as follows. A baseline OCT scan of the sample containing a contrast agent is first acquired, followed by photoillumination of the sample. The photoillumination excites the molecules and repeatedly cycles the molecules through the ground-excited state transition. The molecules have a small chance of ending up in the photobleached state during each transition; as such, given a high enough fluence, the majority of the molecules will eventually end up in the photobleached state. A second OCT scan is then acquired. The two OCT scans, which are different as the absorption spectrum of the contrast agent has changed, are thus processed to determine the distribution of the contrast agent within the sample. This paper provides an understanding of the ICG dynamics in the proposed pump-probe scheme and demonstrates the MCOCT technique in tissue phantoms as well as in an animal model.

Since the proposed pump-probe scheme uses the photobleached state, it is important to understand and quantify the photobleaching mechanism in ICG. Most notable among previous works on the photodegradation process in ICG is the work of Holzer 28 The study, however, is based on either very low (e.g., 10μM ) or very high (e.g., 1.5mM —toxic level29) concentrations of ICG. Moreover, the work was predominantly geared toward characterizing the stability of above two concentrations of ICG (in different solvents) based on the initial probability of photobleaching. In this work, we have studied moderate (e.g., 10 to 140μM ) concentrations of ICG. From this study, we can characterize the variation of the initial probability of photobleaching changes with different ICG concentration and the dependency of the probability of photobleaching on light fluence. We will like to note that it is the cumulative probability of photobleaching that is needed to estimate the time required to photobleach a given ICG sample with a specified irradiance. Therefore, the initial probabilities (such as the ones reported for 10 or 1.5mM ICG concentrations in Ref. 28) cannot be used to accurately estimate the photobleaching extent of 200 and 400μM ICG concentrations used in the demonstration of proposed pump-probe–based MCOCT scheme.

This paper is divided into six sections. In Sec. 2, we will report on our experimental characterizations of the photobleaching mechanism. In Sec. 3, we will elaborate on the computation scheme by which we will extract MCOCT information from our experimental data. In Sec. 4, we will describe the MCOCT experiment and report on our experimental findings. In Sec. 5, we will discuss the adaptation of this MCOCT scheme for clinical applications. Finally, we will conclude with a summary of our findings.

2.

Photobleaching Dynamics of ICG

As an initial step in the study of the photobleaching dynamics of ICG, we performed a simple experiment to verify the stability of the photobleached state. Glass cuvettes (from Starna Cells, Inc.) with 1 cm path length were filled with 50μM ICG (SDB8662, H. W. Sands Corp.) solutions in 1% bovine serum albumin (BSA) and water, and exposed to 790 nm illumination (incident optical power 1.2W ) for several hours. Figures 2a to 2c show absorption cross-section spectra as well as photographs of both unbleached and bleached samples of 50μM ICG in BSA and water. Notice that the unbleached samples of ICG solution in BSA and water are both green in color. However, the bleached samples of ICG solution in BSA and water possess shades of orange and gray, respectively. The bleached samples of ICG in BSA and water were also observed over several days to see if any recovery of the dye from its bleached state had occurred. For this purpose, the bleached samples were kept in the dark and two additional sets of absorption spectra were acquired—one on the third day, and the other on the fifth day [see Figs. 2d and 2e]. A quantitative comparison of the measured absorption spectra of bleached ICG samples shows no signs of recovery from the bleached state, which confirms that the photobleached state is permanent. The permanence of ICG photobleaching illustrates the usefulness of bleached state transition in a pump-probe scheme for OCT—the focus of this study. We note that the extent of change in absorption cross section Δσ=σGσB , where σG and σB are the ground state and bleached state absorption cross sections, respectively, is a useful parameter to estimate achievable contrast from a given ICG sample. From Fig. 2a, it can be seen that ICG solution in BSA shows a higher contrast change compared to that in water. This observation, along with the fact that ICG shows high affinity to bind with proteins,27 indicates that higher contrast should be expected from tissue locales where ICG is attached to proteins provided the ICG samples are completely bleached.

Fig. 2

Shows (a) absorption cross-section spectra and (b) and (c) pictures of unbleached as well as bleached samples of ICG in 1% BSA and water. ICG concentration is 50μM in all the cases. (d) and (e) show absorption spectra of bleached samples of ICG in BSA and water, respectively, acquired at different times. F: fresh; BS: bleached sample.

054017_1_021605jbo2.jpg

Figure 3 shows the setup to characterize the degradation dynamics of ICG. A fraction of the collimated beam from a Ti:sapphire laser (Spectra Physics Tsunami, Model 3941-X3BB) is redirected using a beamsplitter (BS) to form the reference beam. The major portion of the light (signal beam) passes straight through a square cuvette with 2mm diameter circular hole. Glass cover slips are used on each side to hold the ICG samples. The path length L of the specially designed circular cross-section cuvette is 1.52mm (see Fig. 3). Notice that the diameter of the incident collimated laser beam is approximately the size of the circular hole to reduce effects due to diffusion. A balanced photodetector (New Focus, Inc.) is used to measure the transmission through the ICG samples over time. Since, (a) ICG shows high affinity to bind with proteins (which would be the case if injected into a protein environment) and (b) agarose was used (to reduce the effects due to diffusion) in the proof-of-concept demonstrations, 50μM ICG solutions were prepared in deionized (DI) water, 1% BSA, and 0.8% agarose for initial study. ICG solutions in BSA and agarose solutions were also prepared using DI water. Transmission time traces were acquired for different irradiance levels ranging from 25Wcm2 to 80Wcm2 . Before each transmission measurement, the system was balanced (or calibrated) by using the corresponding solvent (water, agarose, or BSA) in an identical cuvette and adjusting the free-space attenuators A1 and A2 accordingly. Note that fresh samples of ICG were used for each transmission measurement. Figures 4a, 4b, 4c show the transmission time traces of 50μM ICG in water, 1% BSA, and 0.8% agarose, respectively, for different irradiance levels. A visual inspection of the transmission curves in Fig. 4 suggests that ICG solution in BSA shows a higher transmission change when bleached compared to ICG solution in water or agarose. This finding is consistent with the Δσ values for ICG in water and 1% BSA, derived from absorption spectra shown in Fig. 2a.

Fig. 3

Experimental setup to study ICG photodegradation dynamics. BS: beamsplitter; M: mirror; A i : i ’th attenuator; R: reference beam; S: sample beam.

054017_1_021605jbo3.jpg

Fig. 4

Transmission curves for 50μM ICG versus time in (a) DI water, (b) 1% BSA, and (c) 0.8% agarose for different illumination intensities (irradiance levels).

054017_1_021605jbo4.jpg

The absorption of ICG does not change linearly with concentration. This is due to the fact that ICG tends to aggregate in water at higher concentrations; aggregated ICG molecules have a different effective absorption cross section. Nevertheless, Beer’s law would still be valid if the absorption cross section were known for a given wavelength and ICG concentration N0 . Assuming that the absorption cross sections of the ground and bleached states do not change during the photobleaching process, the bleached state population density NB(t) can be written as

Eq. 1

NB(t)=1[σGσB]Lln[T(t)Tmin],
where L is the sample path length and T(t) is the transmittance (the ratio of transmitted light intensity to the input light intensity). Tmin=exp[σGN0L] is the minimum transmittance at t=0 , where N0 is the initial concentration of ICG. Because the irradiance decays through the sample, NB(t) , in Eq. 1 represents the spatial average of the bleached state population density. Therefore, the probability of photobleaching, defined as the ratio of total number of photobleached molecules to the total number of photons absorbed by ground state molecules, can be mathematically expressed as

Eq. 2

ϕB,cum(t)=NB(t)×V0tP0hν[1TG(τ)]dτ,
where V is the volume of the ICG sample, h is Planck’s constant, ν is the optical frequency of illumination, P0 is the incident optical power, and TG(t)=Tminexp[σGNBL] is the transmittance attributed to ground state ICG molecules. It can be seen from Eq. 2 that the probability of photobleaching is a function of time unless the terms in the numerator and denominator vary in a similar fashion in time. Moreover, this approach yields a probability that is cumulative in nature such that at a given time t , ϕB,cum(t) gives an average probability of photobleaching for time interval [0,t] .

The transmission data shown in Fig. 4 were processed to determine the cumulative probability of photobleaching, ϕB,cum , using Eqs. 1, 2. The cumulative probability of photobleaching was determined for each data set and averaged. Figure 5 shows average ϕB,cum versus fluence for 50μM ICG in water, 1% BSA, and 0.8% agarose; the corresponding initial cumulative probabilities, ϕB,cum,init , were determined as (1.2±0.17)×105 , (0.97±0.12)×105 , and (0.8±0.3)×105 , respectively. This empirical study also illustrates that ϕB,cum decreases with fluence. For instance, in the case of 50μM ICG in water, the probability of photobleaching decreased by 13 of the corresponding ϕB,cum,init value at 1kJcm2 fluence. Moreover, the rate at which the cumulative probability of photobleaching decreases decays with increasing fluence levels, indicating that for higher (e.g., 10kJcm2 ) fluence levels, ϕB,cum will tend toward a final value ϕB,cum,final . We note that the final cumulative probability of photobleaching, ϕB,cum,final , would be representative of the whole photobleaching event.

Fig. 5

Cumulative probability ϕB,cum for 50-μM ICG in DI water, 1% BSA, and 0.8% agarose. The irradiance in all cases is 33.55Wcm2 .

054017_1_021605jbo5.jpg

We also studied the photobleaching phenomenon as a function of ICG concentrations in DI water. To keep the initial transmissions alike for different concentrations of ICG, cuvettes with different path lengths were assembled. ICG concentrations ranging from 12 to 137μM were used in the study. The irradiance on each sample was 33Wcm2 . For each concentration, four sets of measurements were acquired and averaged for better accuracy. The transmission curves were processed to determine cumulative probability of photobleaching versus time. The study shows that although ϕB,cum,init decreases with increasing ICG concentration M (see Fig. 6 , where a decrease in ϕB,cum,init with increasing dye concentration can be clearly seen), the corresponding ϕB,cum,final values (at 10kJcm2 fluence level) converge to (2.7±0.6)×106 . This trend indicates that for high fluence levels, the above ϕB,cum,final value can also be used to estimate the extent of photobleaching in 200 and 400μM ICG concentrations (used in the MCOCT demonstrations). During this study, we also found the initial probability of photobleaching for 10μM ICG concentration in water to be 1×105 , which is approximately two orders of magnitude smaller than that reported in Ref. 28.

Fig. 6

Initial cumulative probability of photobleaching ϕB,cum,init versus ICG concentration.

054017_1_021605jbo6.jpg

To estimate the extent of photobleaching in a given sample of ICG for a specified irradiance I , consider a simple model that takes into account only the ground, excited, and bleached states (see Fig. 1). The relevant rate equations can therefore be written as

3.

Eq. 3a

NEt=σGIhνNG1τENE,

Eq. 3b

NBt=ϕB,cumτENE,

Eq. 3c

Iz=σGN0I+(σGσB)NBI,
where, NG , NE , and NB are the population densities in the ground, excited, and bleached states, respectively. I is the irradiance, τE is the lifetime of excited state, and hν is the energy of photons. Since NEtNEτE and the irradiance I is much less than saturation intensity of the excited state, Eqs. 3a, 3b can be combined to yield

Eq. 4

NBt=ϕB,cumσGIhν[N0NB].
Notice that irradiance I as well as population densities NG , NE , and NB are functions of depth z as well as time t . If NB(t) represents the spatial average of the bleached state population density varying in time, differential equation 3c can be solved as

Eq. 5

I(z)=I0exp([σGN0+(σBσG)NB(t)]z),
where I0 is the illumination intensity at z=0 . Notice that for small dye concentrations and path length z , one can effectively assume I(z)=I0 . Since the probability of photobleaching ϕB,cum also changes with time due to its dependency on fluence, we can use ϕB,cum,final to estimate the extent of photobleaching in a given sample of ICG. A simplified expression for photobleached population density NB(t) can therefore be determined from Eq. 4 as

Eq. 6

NB(t)=N0(1exp[ϕB,cum,finalσGI0hνt]).

Using ϕB,cum,final=2.7×106 in Eq. 6, the bleached state population density NB(t) was estimated for 200 and 400μM concentrations of ICG in water. The representative curves shown in Fig. 7 were produced for different irradiances. In each case, the irradiance on the ICG samples was assumed uniform through the sample depth z . Notice that in a highly scattering sample, the approximation I(z)=I0 will not hold and the above expression will overestimate the extent of photobleaching process. Nevertheless, the simple exponential model [given by Eq. 6] will help estimating a lower bound on the time required to photobleach a given sample of ICG. Notice that representative curves in Figs. 7a and 7b look similar due to the fact that the ground state absorption cross sections for 200 and 400μM ICG concentrations (in water) are comparable. It can also be estimated from Fig. 7 that for the above concentrations of ICG in water, 7kWcm2 irradiance would require approximately 1s to completely photobleach 200 and 400μM ICG samples.

Fig. 7

Estimated bleached state population density NB(t) for (a) 200μM and (b) 400μM ICG concentrations for different irradiances.

054017_1_021605jbo7.jpg

3.

Pump-Probe MCOCT Setup

Figure 8a shows our spectral domain MCOCT setup. The experiment uses a high power broadband light source, a galvoscanner-based programmable fiber-optic attenuator (PFOA), and a single 2×2 fiber-optic coupler. The PFOA is an important element in the design of this pump-probe MCOCT setup as it controls the amount of light reaching the sample during ICG photobleaching as well as OCT signal acquisition. A personal computer both controls the PFOA and acquires data from the spectrometer. Figure 8b shows the timing diagram of the pump-probe scheme in Fig. 8a. From an operational point of view, the PFOA is first set to deliver low optical power, typically less than 100μW , so that photobleaching is minimal during the first OCT scan. After the first OCT scan is acquired at time t1 , the PFOA is set to deliver high power, typically a 1000-fold increase, for a duration Δt to photobleach the ICG. The optical power transmitted by the PFOA is then reduced to the same level as for the first OCT scan to acquire the second OCT scan at time t2 .

Fig. 8

(a) Pump-probe MCOCT setup using a high power broadband source such as Ti:sapphire and a galvoscanner-based PFOA for both OCT signal acquisition as well as ICG photobleaching. (b) Timing diagram for the proposed pump-probe scheme. PC; polarization controller; SMF: single mode fiber; PFOA: programmable fiber-optic attenuator; FC: fiber collimator.

054017_1_021605jbo8.jpg

At the interferometer, the light collected from the reference and sample arms is spectrally dispersed using a diffraction grating and detected using N different detectors. N can vary from 100 to 1000s depending upon the spectrometer employed. For the spectral domain MCOCT setup shown in Fig. 8a, the spectrally resolved interferometric OCT signal can be mathematically expressed as

Eq. 7

Pinterferernce(k,z)=2S(k)η(k)[1η(k)]RRRs(λ,z)exp[0z[μa(λ,zs)+μs(λ,zs)]dzs]cos[2k(zzr)],
where k is the optical wave number, S(k) is the source spectral profile, η(k) is optical power split factor of the 2×2 coupler ( 12 in an ideal case), RR is the reflectivity of the reference mirror, Rs(λ,z) is the sample reflectivity at depth z , and (zzr) is the path length difference between the reference arm mirror and the plane of interest inside the sample. μa(λ,z) and μs(λ,z) are the absorption and scattering coefficients, respectively, of the sample at depth z . Moreover, the period of the spectral oscillation of the measured signal in k space is proportional to (zzr) indicating that an interface with a smaller (zzr) value will produce a slower sinusoidal spectral oscillation than an interface with a larger (zzr) value. Therefore, a Fourier transform of the spectral measurement will produce a line scan profile (also known as an A-scan) of the sample under study. The additional exponential term accounts for the loss of collected light due to absorption and scattering during the passage into and out of the sample.

In a pump-probe scheme such as the one discussed above using ICG as the contrast agent, the photoillumination permanently alters the absorption coefficient of the dye. In this context, the absorption coefficient can be mathematically written as

Eq. 8

μa(λ,z)=μa,intrinsic(λ,z)+μa,ICG_ground(λ,z)+Δμa,ICG(λ,z),
where μa,intrinsic(λ,z) is the localized intrinsic absorption coefficient of the sample, μa,ICG_ground(λ,z) is the localized absorption coefficient component for ICG in the ground state, and Δμa,ICG(λ,z) is the change in the absorption coefficient between the two OCT scans. The two quantities, μa,ICG_ground(λ,z) and Δμa,ICG(λ,z) can be expressed as

Eq. 9

μa,ICG_ground(λ,z)=σG(λ)N0(z),

Eq. 10

Δμa,ICG(λ,z)=μa,ICG_bleached(λ,z)μa,ICG_ground(λ,z) =[σG(λ)NG(z)+σB(λ)NB(z)]σG(λ)N0(z),
where σG(λ) and σB(λ) are the ground and bleached state absorption cross-section spectra, respectively. N0(z) is the initial ground state population density of ICG at z , whereas NG(z) and NB(z) are the ground and bleached state population densities, respectively, when the second OCT scan is acquired. As the irradiance I is much less than saturation intensity of the excited state, NG(z)NE(z) , we can safely assume that N0(z)=NG(z)+NB(z) . Therefore, Eq. 10 can be simplified as

Eq. 11

Δμa,ICG(λ,z)=[σB(λ)σG(λ)]NB(z).
If Pinterference_ground(λ,z) and Pinterference_bleached(λ,z) are the OCT signals acquired before and after the photobleaching of ICG, the change in absorption coefficient is related to the two OCT scans by

Eq. 12

0zΔμa,ICG(λ,z)dz=ln[Pinterferernce_bleached(k,z)Pinterferernce_ground(k,z)].
Solving Eqs. 11, 12 simultaneously, the distribution of ICG within the sample, represented by NB(z) , can be determined as

Eq. 13

NB(z)=ddz[1{σG(λ)σB(λ)}ln{Pinterferernce_bleached(k,z)Pinterferernce_ground(k,z)}].

4.

Experiment

A spectral domain MCOCT system shown in Fig. 8a was realized for demonstration. Light from a mode locked Ti:sapphire laser [ 12nm full width at half maximum (FWHM)] is coupled into a long 5125μm single mode fiber (SMF) for spectrum broadening. The broadband probe light ( 60nm FWHM centered at 800nm ) from the SMF is coupled into a 2×2 single mode fiber-optic coupler via a galvoscanner-based PFOA. Light from the coupler is collimated by lens L1(f1=11mm) . A 10× microscope objective lens (f2=16.5mm) from Newport Corp. is used to focus the collimated light onto the sample. The calculated 1e2 beam diameter on the sample is 8.2μm , which corresponds to a depth of focus of 135μm . Light collected from the two arms of the interferometer is detected by a spectrometer designed and built in our laboratory. The spectrometer uses a 1200linesmm diffraction grating together with 2048 pixels linear detector array and features 0.07 nm spectral resolution. The spectral oscillation data acquired by the spectrometer are first rescaled and resampled evenly in k space, before it is Fourier transformed to retrieve the sample depth profile (or A-scan). The rescaled and resampled data are also processed to adjust for any dispersion mismatch between the two arms of the interferometer using the method outlined in Ref. 30. Our laboratory MCOCT system features a measured signal-to-noise ratio (SNR) of 98dB at 1-ms integration time. The corresponding theoretical shot noise limited SNR is 108dB .

In a proof-of-concept demonstration, a 0.84 mm path length glass cuvette filled with 200μM (N0=12×1016cm3) ICG solution in 1% BSA was used as a sample [see Fig. 9a ]. The galvoscanner-based PFOA was programmed to deliver 68μW and 10mW ( 19kWcm2 irradiance based on a beam spot diameter of 8.2μm ) on the sample for OCT signal acquisition and ICG photobleaching, respectively. The linear detector array integration time for OCT signal acquisition was set to 400μs . The ICG was photobleached for duration Δt=2.5s between the two OCT scans. Figure 9b shows the two A-scans of the 0.84 mm path length cuvette filled with 200μM ICG solution in 1% BSA before and after the photobleaching. The first two interfaces of the glass cuvette show no contrast as expected. However, the third and fourth interfaces of the glass cuvette show 7dB of measured contrast, attributable to the change in absorption coefficient of the ICG solution within the cuvette. Next, the 0.84 mm path length glass cuvette was filled with a mixture of 200μM ICG and 0.25% (v/v) 0.1μm latex microspheres. A series of 10 A-scans was acquired and processed. Figure 9c shows averaged A-scans of the glass cuvette before and after the photobleaching. Again, the first two interfaces show no contrast as anticipated. The last two interfaces of the glass cuvette yield a measured contrast of 5dB . Notice that the contrast increases with ICG sample depth as it is cumulative in nature [see Eq. 12]. A ratio of the two averaged A-scans is shown in Fig. 9d to support the argument. Using Eq. 13, average bleached state population NB of ICG molecules within the cuvette was determined to be 9×1016cm3 . We note that at 19kWcm2 irradiance, it should take less than 0.5s to photobleach > 99% of the ICG molecules (see Fig. 7). However, only 75% ICG molecules were observed to have transitioned to the photobleached state with an irradiation time of 2.5s . The reason not all the ICG molecules were bleached can be attributed to the irradiance decay through the highly scattering sample [0.25% (v/v) 0.1μm latex microspheres] and possible diffusion of ICG molecules to and from the region of interest.

Fig. 9

(a) Side view of a 0.84 mm path length cuvette used as a sample. (b) A-scans of the glass cuvette filled with 200μM ICG solution in 1% BSA before and after the photobleaching. (c) A-scans of the glass cuvette was filled with a mixture of 200μM ICG and 0.25% volume concentration 0.1μm latex microspheres before and after the photobleaching. (d) Ratio of the two A-scans is shown in (c).

054017_1_021605jbo9.jpg

Tissue phantoms as well as an animal model were also used in order to evaluate ICG as a contrast agent in the pump-probe–based MCOCT scheme. In these MCOCT demonstrations, the ICG was used with agarose to reduce the effects due to the diffusion of ICG molecules. In our estimation of pump light dosage for photobleaching of ICG in agarose, it was assumed that ϕB,cum,final is approximately the same as that for ICG in water. This approximation is fairly reasonable since (a) the initial probabilities of photobleaching for ICG in water and agarose are only different by less than a factor of 2 and (b) ϕB,cum decreases with fluence in a similar fashion for both the cases (see Fig. 5). Therefore, by approximating ϕB,cum,final=2.7×106 in Eq. 6, it is estimated that 19kWcm2 irradiance would require less than 0.5s to photobleach > 99% of the ICG molecules in 400μM ICG sample in agarose.

The tissue phantoms were prepared by using a glass cuvette comprised of three cells [see Figs. 10a and 10b for side and front views, respectively]. Each of the three cells was filled with a mixture containing 0.25% (v/v) 0.1μm latex microspheres, and 0.64% (v/v) agarose. Cells (i) and (ii) also contained 400μM ICG (N0=24×1016cm3) dissolved in 0.5% BSA and DI water, respectively, whereas the third cell [marked (iii)] contained additional 0.125% (v/v) 1.437μm latex microspheres in order to bring the total attenuation of the cell contents close to that of the mixtures in (i) and (ii).

Fig. 10

(a) Side and (b) front view of a cuvette composed of 3 cells filled with a mixture containing 0.25% (v/v) 0.1μm latex microspheres and 0.64% (v/v) agarose. Cells (i) and (ii) also contained 400μM ICG dissolved in 0.5% BSA and DI water, respectively. Cell (iii) contained additional 0.125% (v/v) 1.437μm latex microspheres to bring the total attenuation of the cell contents close to that of the mixtures in (i) and (ii). (c) and (d) show B-scan images (600 A-scans each) of the glass cuvette, shown in (a), acquired with 20μm lateral scan step before and after ICG photobleaching, respectively. (e) and (g) show averaged A-scans of cells (ii) and (iii) before and after high intensity illumination, respectively, at the locations marked by dotted lines; the corresponding ratios of the averaged A-scans are shown in (f) and (h).

054017_1_021605jbo10.jpg

To acquire B-scans [two-dimensional (2D) images], the glass cuvette was mounted on a motorized linear stage in order to move the sample under the probe beam. The PFOA was programmed such that light exposure levels at the glass cuvette for OCT signal acquisition and ICG photobleaching were 60μW and 10mW , respectively. The linear detector array integration time per A-scan was set to 1ms . First, a B-scan image composed of 600 A-scans was acquired with a 20μm lateral scan step [see Fig. 10c]. During the second B-scan, the sample was also photobleached using 19kWcm2 irradiance at each A-scan position. Although the estimated time to photobleach the 400μM ICG samples was less than 0.5s , due to irradiance decay through the highly scattering phantoms (as noted earlier) we had to increase the exposure time for effective photobleaching. In summary, the phantoms were photobleached for Δt=10s duration at each A-scan location during the second B-scan, which is shown in Fig. 10d. During both B-scans, an OCT sample size of 50 repeats was acquired at each A-scan position in order to reduce speckle. A comparison of Figs. 10c and 10d shows that cells marked as (i) and (ii) containing samples of ICG in 0.5% BSA and water, respectively, show significant increase in backscatter from the latex microspheres. This enhanced backscatter is attributable to change in absorption coefficients of the ICG samples in the first two cells. Figure 10e shows averaged A-scans [at the marked location in cell (ii)] before and after the photobleaching; the corresponding ratio of the averaged A-scans is shown in Fig. 10g. Using Eq. 13, the average bleached state population density NB of ICG molecules within cell (ii) was determined as 23×1016cm3 , indicating that 95% ICG molecules were transitioned to the photobleached state. The third cell [marked (iii)] showed no change in backscatter [also see Figs. 10f and 10h for representative A-scans and their ratio at the marked location] indicating that regions without ICG do not yield any contrast in the proposed pump-probe scheme for MCOCT.

To demonstrate the ICG detection capability in an animal model, a mixture of 400μM ICG, 0.25% (v/v) 0.1μm latex microspheres, and 0.6% (v/v) of agarose was used to fill the gill arch cavities of euthanized stage 54 Xenopus laevis [see Fig. 11a ]. First, the PFOA was set to illuminate the sample with 60μW and a whole B-scan (400 A-scans; lateral resolution 7.5μm ) was acquired by moving the sample with a motorized linear stage [see Figs. 11b and 11d]. During the second B-scan, the PFOA was set to illuminate the sample with 10mW power at each A-scan position for 12 s duration. The optical power was then readjusted to 60μW to acquire the corresponding A-scan, and so forth. During both B-scans, an OCT sample size of 50 repeats was acquired at each A-scan position in order to reduce speckle. Figures 11c and 11e show the second B-scan where increased backscatter from the microspheres (indicating the localization of ICG) in gill arch cavities can be clearly seen. This increased backscatter, attributable to the change in the absorption coefficient of ICG from within the gill arch cavities of the tadpole, demonstrates the functionality of ICG as a contrast agent in the proposed pump-probe scheme. Further analysis of the xenopus images to estimate achieved contrast could not be carried out due to sample shifting between the two B-scans (acquired before and after the photobleaching).

Fig. 11

(Color online only) MCOCT imaging of a stage 54 Xenopus laevis. (a) Ventral view of the animal filled with a mixture of 400μM ICG (green), 0.25% 0.1μm latex microspheres, and 0.6% volume concentration of agarose; the red line indicates the region where OCT scans were acquired. (b) and (c) show B-scan images of the illuminated region before and after ICG photobleaching, respectively. (d) and (e) show B-scan images of gill arch cavities of another stage 54 Xenopus laevis. Increased backscatter from within the gill arch cavities of the animal can be seen.

054017_1_021605jbo11.jpg

5.

Considerations of ICG-Based Pump-Probe Scheme for Clinical Applications

This MCOCT scheme can be applied in animal model studies to tomographically visualize body cavities, blood vessel systems, and gastrointestinal tracts. The fact that ICG is approved for clinical usage points to possible usage of this MCOCT scheme for clinical applications as well. While the sensitivity achieved in our study is insufficiently high for clinical applications, the findings of our study do form a basis by which we can evaluate the improvements needed for this MCOCT scheme and other similar MCOCT schemes to be clinically applicable.

There are two major considerations that we need to account for. First, is it possible to deliver the fluence needed for photobleaching in a reasonable time frame? Second, is the method sufficiently sensitive?

The time frame for fluence delivery can be estimated by the following approach. The American National Standards Institute (ANSI) guidelines for skin exposure31 suggest that an acceptable light illumination for long time frame exposure is 300mWcm2 . For a nominal ICG concentration of 10μM , the approved ICG concentration in blood for clinical procedures,32 photobleaching of 25% of the dye will require a time duration of 350s . This time frame can be reduced by using a higher irradiance or switching to a dye with a higher quantum yield for photobleaching. This estimate implies that using ICG with this MCOCT scheme is not well-suited for time critical clinical procedures.

The sensitivity of the method can be estimated by noting that the fundamental limit on the contrast agent sensitivity is related to the SNR of the underlying OCT system as22

Eq. 14

SNRMCOCT=(εPsThν)×R(z)exp[20zμa,s(λ,z)dz]×[Δμa(z)ΔL]2,
where, ε is the detection sensitivity, Ps is the collected sample power, T is integration time, and hν is the energy of photons. R(z) is the sample reflectivity at depth z , whereas μa,s is the overall absorption and scattering extinction coefficient. Δμa represents the change in absorption coefficient of the thin absorptive layer of thickness ΔL at depth z inside the scattering sample. The fundamental limit on the contrast agent sensitivity can be determined by setting Eq. 14 equal to 1. To determine the minimal concentration of ICG detectable by our MCOCT system, we use the measured OCT SNR of 98dB at 1 ms integration time. By assuming a nominal R(z)=104 (typical scattering from tissue), we find out that the minimal detectable concentration for ΔL=100μm thick ICG layer at z=0 is 0.8μM . This concentration compares very favorably with the clinically safe ICG concentration of 10μM in blood.32

Our experiments show that this sensitivity is not reached because the above analysis assumes that the OCT signal does not show significant variation from scan to scan. While this assumption is true for immobile targets, the presence of fluid and the associated speckle fluctuations lead to significant scan-to-scan signal variation. This variation can be as large as 4 to 5dB from scan to scan in our experiments. While averaging can help reduce the variation, this variation limited our ability to discern contrast. The reduction of speckle fluctuations is crucial to enabling this and similar MCOCT schemes to reach the contrast sensitivity given in the above analysis.

6.

Conclusions

In conclusion, we have proposed a pump-probe scheme for MCOCT imaging of biological samples using ICG as a contrast agent. The photobleached state in ICG is permanent in nature that makes this transition useful in the reported pump-probe scheme. A photobleaching study of 50μM ICG solutions in DI water, BSA, and agarose shows that the cumulative probability of photobleaching decreases with fluence in all the cases. Various concentrations (ranging from 12 to 137μM ) of ICG in DI water were also studied, which illustrates that ϕB,cum,init decreases with the increasing dye concentration. We also found that at higher fluence levels, the final values of cumulative probability (for the range of ICG concentrations used in the study) tend to converge to (2.7±0.6)×106 . The final cumulative probability ϕB,cum,final is useful in estimating the extent of photobleaching in a given sample of ICG for known irradiance and exposure time. Using our laboratory spectral domain MCOCT setup, we have successfully mapped the distribution of ICG ( 400μM concentration) within tissue phantoms as well as stage 54 Xenopus laevis. The ability to detect and map depth-resolved distribution of ICG within a biological sample illustrates the potential of the proposed pump-probe scheme to provide high resolution structural information of tissue as well as the ability to differentiate between certain tissue pathologies.

As opposed to spectral triangulation MCOCT,20 which is based on spectral differential absorption of ICG, the proposed pump-probe scheme relies upon the change in absorption spectrum of ICG due to photobleaching. Since the proposed method uses the permanently photobleached state in ICG (the photobleached dye molecules do not relax back to the ground state), both the optical design as well as the associated data acquisition methodology are simple.

However, the proposed technique does rely on its ability to effectively photobleach the dye. The current pump-probe MCOCT scheme is limited in imaging speed due to the low probability of photobleaching, which requires high irradiance and relatively long exposure time to photobleach ICG. For instance, the estimated light exposure level required for complete photobleaching of 10μM ICG is 4 to 5 orders of magnitude higher than the ANSI prescribed limits. Previously, it has been shown that the probability of ICG photobleaching decreases in the presence of certain materials such as silver particles. A study of materials that would potentially increase the probability of ICG photobleaching upon conjugation would benefit the proposed ICG-based pump-probe MCOCT scheme. Using higher irradiance (subject to the availability of high power light source) and modifying the system accordingly, so as to photobleach the entire region of interest as opposed to line-by-line photobleaching (the current methodology), we can improve the overall imaging speed of the system.

Acknowledgment

This research was supported by National Institute of Health Grant No. 1RZ1EB004602-1.

References

1. 

D. Huang, E. A. Swanson, C. P. Lin, J. S. Schuman, W. G. Stinson, W. Chang, M. R. Hee, T. Flotte, K. Gregory, C. A. Puliafito, and J. G. Fujimoto, “Optical coherence tomography,” Science, 254 1178 –1181 (1991). 0036-8075 Google Scholar

2. 

B. Povazay, K. Bizheva, A. H. Unterhuber, B. Hermann, H. Sattmann, A. F. Fercher, W. Drexler, A. Apolonski, W. J. Wadsworth, J. C. Knight, P. S. J. Russell, M. Vetterlein, and E. Scherzer, “Submicrometer axial resolution optical coherence tomography,” Opt. Lett., 27 1800 (2002). 0146-9592 Google Scholar

3. 

M. R. Hee, J. A. Izatt, E. A. Swanson, D. Huang, J. S. Schuman, C. P. Lin, C. A. Puliafito, and J. G. Fujimoto, “Optical coherence tomography of the human retina,” Arch. Ophthalmol. (Chicago), 113 325 –332 (1995). 0003-9950 Google Scholar

4. 

J. J. Kaluzny, M. Wojtkowski, and A. Kowalczyk, “Imaging of the anterior segment of the eye by spectral optical coherence tomography,” Opt. Appl., 32 581 –589 (2002). 0078-5466 Google Scholar

5. 

P. Targowski, M. Wojtkowski, A. Kowalczyk, T. Bajraszewski, M. Szkulmowski, and W. Gorczynska, “Complex spectral OCT in human eye imaging in vivo,” Opt. Commun., 229 79 –84 (2004). 0030-4018 Google Scholar

6. 

A. Unterhuber, B. Povazay, B. Hermann, H. Sattmann, A. Chavez-Pirson, and W. Drexler, “In vivo retinal optical coherence tomography at 1040nm-enhanced penetration into the choroid,” Opt. Express, 13 3252 –3258 (2005). https://doi.org/10.1364/OPEX.13.003252 1094-4087 Google Scholar

7. 

A. M. Rollins, R. Ung-arunyawee, A. Chak, R. C. K. Wong, K. Kobayashi, M. V. Sivak, and J. A. Izatt, “Real-time in vivo imaging of human gastrointestinal ultrastructure by use of endoscopic optical coherence tomography with a novel efficient interferometer design,” Opt. Lett., 24 1358 –1360 (1999). 0146-9592 Google Scholar

8. 

T. Q. Xie, M. L. Zeidel, and Y. Pan, “Detection of tumorigenesis in urinary bladder with optical coherence tomography: Optical characterization of morphological changes,” Opt. Express, 10 1431 –1443 (2002). 1094-4087 Google Scholar

9. 

G. Tearney and B. Bouma, “Atherosclerotic plaque characterization by spatial and temporal speckle pattern analysis,” Opt. Lett., 27 533 –535 (2002). https://doi.org/10.1038/417533a 0146-9592 Google Scholar

10. 

B. E. Bouma, G. J. Tearney, H. Yabushita, M. Shishkov, C. R. Kauffman, D. D. Gauthier, B. D. MacNeill, S. L. Houser, H. T. Aretz, E. F. Halpern, and I. K. Jang, “Evaluation of intracoronary stenting by intravascular optical coherence tomography,” Heart, 89 317 –320 (2003). 1355-6037 Google Scholar

11. 

X. J. Wang, T. E. Milner, J. F. de Boer, Y. Zhang, D. H. Pashley, and J. S. Nelson, “Characterization of dentin and enamel by use of optical coherence tomography,” Appl. Opt., 38 2092 –2096 (1999). 0003-6935 Google Scholar

12. 

J. A. Izatt, M. D. Kulkami, S. Yazdanfar, J. K. Barton, and A. J. Welch, “In vivo bidirectional color Doppler flow imagingof picoliter blood volumes using optical coherence tomograghy,” Opt. Lett., 22 1439 –1441 (1997). 0146-9592 Google Scholar

13. 

J. F. deBoer, T. E. Milner, M. J. C. van Gemert, and J. S. Nelson, “Two-dimensional birefringence imaging in biological tissue by polarization-sensitive optical coherence tomography,” Opt. Lett., 22 934 –936 (1997). 0146-9592 Google Scholar

14. 

U. Morgner, W. Drexler, F. X. Kartner, X. D. Li, C. Pitris, E. P. Ippen, and J. G. Fujimoto, “Spectroscopic optical coherence tomography,” Opt. Lett., 25 111 –113 (2000). 0146-9592 Google Scholar

15. 

R. Leitgeb, M. Wojtkowski, A. Kowalczyk, C. K. Hitzenberger, M. Sticker, and A. F. Fercher, “Spectral measurement of absorption by spectroscopic frequency-domain optical coherence tomography,” Opt. Lett., 25 820 –822 (2000). 0146-9592 Google Scholar

16. 

K. D. Rao, M. A. Choma, S. Yazdanfar, A. M. Rollins, and J. A. Izatt, “Molecular contrast in optical coherence tomography by use of a pump-probe technique,” Opt. Lett., 28 340 –342 (2003). https://doi.org/10.1038/425340a 0146-9592 Google Scholar

17. 

T. M. Lee, A. L. Oldenburg, S. Sitafalwalla, D. L. Marks, W. Luo, F. J. J. Toublan, K. S. Suslick, and S. A. Boppart, “Engineered microsphere contrast agents for optical coherence tomography,” Opt. Lett., 28 1546 –1548 (2003). 0146-9592 Google Scholar

18. 

C. Vinegoni, J. S. Bredfeldt, D. L. Marks, and S. A. Boppart, “Nonlinear optical contrast enhancement for optical coherence tomography,” Opt. Express, 12 331 –341 (2004). https://doi.org/10.1364/OPEX.12.000331 1094-4087 Google Scholar

19. 

Y. Jiang, I. Tomov, Y. M. Wang, and Z. P. Chen, “Second-harmonic optical coherence tomography,” Opt. Lett., 29 1090 –1092 (2004). https://doi.org/10.1364/OL.29.001090 0146-9592 Google Scholar

20. 

C. Yang, L. E. Lamb, J. D. Simon, M. A. Choma, B. Applegate, and J. Izatt, “Spectral triangulation molecular contrast optical coherence tomography with indocyanine green as the contrast agent,” Opt. Lett., 29 2016 (2004). https://doi.org/10.1364/OL.29.002016 0146-9592 Google Scholar

21. 

C. H. Yang, L. E. L. McGuckin, J. D. Simon, M. A. Choma, B. E. Applegate, and J. A. Izatt, “Spectral triangulation molecular contrast optical coherence tomography with indocyanine green as the contrast agent,” Opt. Lett., 29 2016 –2018 (2004). https://doi.org/10.1364/OL.29.002016 0146-9592 Google Scholar

22. 

C. H. Yang, “Molecular contrast optical coherence tomography: A review,” Photochem. Photobiol., 81 215 –237 (2005). https://doi.org/10.1562/2004-08-06-IR-266.1 0031-8655 Google Scholar

23. 

A. L. Oldenburg, F. J. J. Toublan, K. S. Suslick, A. Wei, and S. A. Boppart, “Magnetomotive contrast for in vivo optical coherence tomography,” Opt. Express, 13 6597 –6614 (2005). https://doi.org/10.1364/OPEX.13.006597 1094-4087 Google Scholar

24. 

M. V. Sarunic, B. E. Applegate, and J. A. Izatt, “Spectral domain second-harmonic optical coherence tomography,” Opt. Lett., 30 2391 –2393 (2005). https://doi.org/10.1364/OL.30.002391 0146-9592 Google Scholar

25. 

H. Nimura, “Infrared ray electronic endoscopy combined with indocyanine green injection for detection of sentinel nodes of patients with gastric cancer,” Br. J. Surg., 91 575 –579 (2004). 0007-1323 Google Scholar

26. 

F. C. Piccolino, C. M. Eandi, L. Ventre, R. C. R. De la Longrais, and F. M. Grignolo, “Photodynamic therapy for chronic central serous chorioretinopathy,” Retina, 23 752 –763 (2003). 0275-004X Google Scholar

27. 

H. Gratz, A. Penzkofer, C. Abels, R.-M. Szeimies, M. Landthaler, and W. Baumler, “Photo-isomerisation, triplet formation, and photo-degradation dynamics of indocyanine green solutions,” J. Photochem. Photobiol., A, 128 101 –109 (1999). https://doi.org/10.1016/S1010-6030(99)00174-4 1010-6030 Google Scholar

28. 

W. Holzer, M. Mauerer, A. Penzkofer, R.-M. Szeimies, C. Abels, M. Landthaler, and W. Baumler, “Photostability and thermal stability of indocyanine green,” J. Photochem. Photobiol., B, 47 155 –164 (1998). https://doi.org/10.1016/S1011-1344(98)00216-4 1011-1344 Google Scholar

29. 

J. E. Lee, T. J. Yoon, B. S. Oum, J. S. Lee, and H. Y. Choi, “Toxicity of indocyanine green injected into the subretinal space—Subretinal toxicity of indocyanine green,” Retina, 23 675 –681 (2003). 0275-004X Google Scholar

30. 

M. Wojtkowski, V. J. Srinivasan, T. Ko, J. G. Fujimoto, A. Kowalczyk, and J. S. Duker, “Ultrahigh-resolution, high-speed, Fourier domain optical coherence tomography and methods for dispersion compensation,” Opt. Express, 12 2404 –2422 (2004). https://doi.org/10.1364/OPEX.12.002404 1094-4087 Google Scholar

31. 

American National Standards Institute, “Safe Use of Lasers,” ANSI Z 136, 1 –2000 Laser Institute of America, Orlando, FL (2000). Google Scholar

32. 

P. E. Stanga, J. I. Lim, and P. Hamilton, “Indocyanine green angiography in choriorefinal diseases: Indications and interpretation—An evidence-based update,” Ophthalmology, 110 15 –21 (2003). 0161-6420 Google Scholar
©(2006) Society of Photo-Optical Instrumentation Engineers (SPIE)
Zahid Yaqoob, Emily J. McDowell, Jigang Wu, Xin Heng, Jeff P. Fingler, and Changhuei Yang "Molecular contrast optical coherence tomography: a pump-probe scheme using indocyanine green as a contrast agent," Journal of Biomedical Optics 11(5), 054017 (1 September 2006). https://doi.org/10.1117/1.2360525
Published: 1 September 2006
Lens.org Logo
CITATIONS
Cited by 31 scholarly publications and 1 patent.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Optical coherence tomography

Absorption

Molecules

Statistical analysis

Glasses

Latex

Imaging systems

Back to Top