1 March 2008 Photoacoustic imaging of biological tissue with intensity-modulated continuous-wave laser
Author Affiliations +
Abstract
We build a photoacoustic imaging system using an intensity-modulated continuous-wave laser source, which is an inexpensive, compact, and durable 120-mW laser diode. The goal is to significantly reduce the costs and sizes of photoacoustic imaging systems. By using a bowl-shaped piezoelectric transducer, whose numerical aperture is 0.85 and resonance frequency is 2.45 MHz, we image biological tissues with a lateral resolution of 0.45 mm, an axial resolution of 1 mm, and an SNR as high as 43 dB.

1.

Introduction

Photoacoustic (PA) imaging 1, 2, 3, 4, 5, 6, 7, 8 is a hybrid technology that maps the internal distribution of optical energy deposition (specific absorption) in biological tissues by detecting laser-induced ultrasonic (so-called PA) waves. PA imaging takes advantage of the weaker ultrasonic scattering, whose magnitude per unit path length is two orders of magnitude less than that of optical scattering,9 to combine ultrasonic resolution with optical contrast. PA imaging is able to extract certain physiological parameters for functional imaging by varying the optical wavelength of the excitation laser.10

State-of-the-art PA techniques always use pulsed lasers as the PA excitation source. The major reasons for this choice include 1 high energy in each laser pulse—even if within the ANSI safety limit11—yields strong PA signals, leading to a good signal-to-noise ratio12 (SNR), 2 PA transients provide axial resolution along the ultrasonic propagation direction, and (3) difference in the time of flight of PA waves reduces signal cluttering. For example, the first dark-field confocal PA microscope (PAM) using pulsed laser illumination is capable13 of imaging biological tissue with 15μm axial resolution, 45 to 120μm lateral resolution, and 40dB SNR. Pulsed PA imaging techniques, however, suffer from laser jittering as well as acoustic and thermal noises within the wide bandwidth of the ultrasonic transducer. In addition, high-energy pulsed lasers are expensive and bulky, which limits the applications of PA imaging technologies in medicine and biology.

Continuous-wave (cw) PA excitation sources such as high-power laser diodes,14 which are inexpensive, compact, and durable, offer attractive alternatives. A cw PA detection with a narrow-band ultrasonic transducer and a narrow-band or lock-in amplifier provides high sensitivity and strong noise rejection. A cw PAM based on lock-in detection has been demonstrated to measure surface15 and subsurface16 properties of solids. Although single-frequency detection does not enable unequivocal phase-based axial resolution owing to the 2π phase ambiguity, chirping (frequency sweeping) over a bandwidth enables full axial resolution, which is comparable to the time-domain resolution acquired with the same bandwidth. However, single-frequency detection with a focused ultrasonic transducer can provide amplitude-based axial resolution as will be shown later. In addition, dual-frequency detection can be tailored to reject undesirable PA signals from a particular depth.

In this study, we investigated the possibility of cw PA imaging of biological tissue using a laser diode as a PA source. We also compared the sensitivity of the cw technique with the pulsed counterpart.10

2.

Theory

Because laser diodes produce peak powers several orders of magnitude less than Q -switched lasers, special attention should be paid to the SNR. We first consider the peak PA pressure generated by light absorption. In the case of pulsed excitation, the light intensity can be approximated by a Gaussian function: IEP(ωPτπ)exp(t2τ2) , where EP is the optical fluence, and τ is the pulse width ( ωP1τ is the cut-off frequency of the PA transient). In the case of intensity-modulated cw excitation, the light intensity is given by I=Icw[1+sin(ωct)] , where Icw is the time-averaged light intensity, and ωc is the modulation frequency. If light intensity is expressed as I=I0f(t̂) , where I0 denotes the maximum, and t̂ denotes a dimensionless time ( t̂=tτ or ωct ), the associated PA pressure is proportional17 to df(t̂)dt̂ . Consequently, regardless of the absorber geometry,18 the maximum PA pressure due to pulsed excitation is given by pP,max(2πe)12EPωP , whereas the PA pressure amplitude due to cw excitation is given by pcw,maxIcw .

In both the pulsed and cw modes, the maximum values of EP and Icw are limited by the ANSI safety standards.11 In the visible spectral region, the maximum EP is 20mJcm2 and the maximum Icw is 200mWcm2 . To achieve comparable lateral resolution between the pulsed and cw modes, we assume ωP=ωc . If ωc2π is set to 2.45MHz , we find that pP,max is approximately six orders of magnitude (120dB) stronger than pcw,max .

The difference in bandwidth between the two modes offsets the difference in signal strength because noise amplitude is proportional to the square root of the signal bandwidth. In pulsed mode, the cutoff frequency of the bandwidth is ΔfPωP2π . In cw mode, the bandwidth is limited only by the desired data acquisition rate of each pixel of the image, which can be set to the 10-Hz laser repetition rate of the table-top PAM system under comparison. This difference in bandwidth increases the ratio of the SNR of the cw mode to that of the pulsed mode by approximately three orders of magnitude (60dB) .

The difference in sensitivity between the ultrasonic transducers used in the two modes further offsets the difference in signal strength. In pulsed mode, the piezoelectric transducer, which must have a broad bandwidth, typically has an insertion loss19 of the order of 20dB . This loss is due to either the low electromechanical coupling coefficient K of piezoelectric material or the high acoustic impedance Z because no piezoelectric material can yet provide both high K and low Z . For example, polymers such as PVDF have low Z but also low K , whereas piezoceramics such as PZT have high K but also high Z . High Z ultrasonic transducers present strong acoustic impedance mismatch with soft biological tissue, which causes strong acoustic reflection at the interface. In cw mode, a low-loss resonant piezoelectric transducer can be used. When the thickness of the piezoelectric material equals one half of the acoustic wavelength, acoustic reflection at the interface in theory approaches zero. As a result, resonant transducers can be made of high K and high Z piezoceramic materials. In addition, resonant transducers are not required to use damping backing material, which is acoustically lossy. Therefore, a resonant transducer suffers from only electrical resistive loss and mechanical loss in the piezoelectric material. At a few megahertz frequency, these losses can be as little as a few decibels. In the end, this difference translates into approximately another order of magnitude (20dB) improvement in SNR for cw detection.

The difference in signal strength can also be offset by using higher intensity laser diodes that are modulated with low duty cycle tone bursts. According to the ANSI standards11 for less than 10s of exposure, the maximum permissible exposure is limited by 1100te14 in millijouls per square centimeter, where te denotes the exposure duration in seconds. For a 10-Hz pulse repetition frequency with a 1% duty cycle, the maximum permissible exposure is 20Wcm2 , at which the SNR can be improved by another order of magnitude (20dB) .

Altogether the aforementioned three factors can offset the difference in signal strength between pulsed and cw modes by approximately five orders of magnitude. Therefore, a fully implemented cw system should provide a SNR about only one order of magnitude (20dB) worse than that of a pulsed system. We so far, however, have taken advantage of only the first two factors. As a result, the SNR of the cw system is expected to be about two orders of magnitude (40 dB) worse than that of a pulse system.

3.

Instrumentation

The pulsed-laser-based PAM system is similar to the high-frequency (50-MHz) version described elsewhere.10, 13 A tunable dye laser (ND6000, Continuum, California) pumped by an Nd:YAG laser (Brilliant B, Bigsky, Montama; pulse duration, 6.5ns ; repetition rate, 10Hz ) serves as the PA excitation source. The pulsed system uses a 5-MHz spherically focused ultrasonic transducer made of composite material (113-842-320, GE Inspection Technology; diameter, 6.35mm ; focal length, 18mm ). The incident fluence on the sample surface and at the optical focus is11 <20mJcm2 (the ANSI safety limit). To acquire an image, the PAM system records the PA wave at each location of the ultrasonic transducer and performs raster-scanning in the horizontal (xy) plane with a step size of 125μm .

The cw PA imaging system uses a scanning system similar to the one in the pulsed PAM system but different optical excitation and data acquisition systems (see Fig. 1 ). An inexpensive, compact, and durable laser diode (GH0781JA2C, Sharp; power: 120mW ; wavelength: 784nm ) driven by a laser diode current controller (LDC220, Thorlabs) is the excitation source. We modulate the power of the laser diode at 2.45MHz with a 100% modulation depth by applying a 10-V sinusoidal signal from a function generator to the laser bias through a serial 50Ω resistor. Light from the laser diode is delivered via an optical fiber. The transmitted light passes to the sample through a 5-mm-diam optical collimating lens secured within the hole on the axis of the transducer. The resulting time-averaged spatial peak optical power density on the sample surface after loss in the light delivery system was measured to be near but within ANSI safety limit of 200mWcm2 . An ultrasonic transducer has a bowl-shaped piezoceramic active element (Pz27 ceramic, Ferroperm Piezoceramics A/S, Kvistgaard, Denmark) with a 38mm diameter and a 22mm radius of curvature. The through-thickness resonance frequency of the transducer is 2.45MHz . Gold electrodes are deposited on the two spherical surfaces of the active element. The electrode on the concave spherical surface is soldered to the transducer shell with low-melting-point alloy ( SnCuInCd eutectic), whereas the other electrode is connected to a resonant preamplifier via an impedance-matching line transformer. The preamplified PA signal is detected by a lock-in amplifier (SR844, Stanford Research Systems), which is used as both a phase-sensitive detector and a data acquisition system. The components within the dashed-box in Fig. 1 are translated within a container with immersion liquid (water). A window at the bottom of the water container is sealed with an optically and ultrasonically transparent polyethylene membrane (thickness, 0.044mm ). During data acquisition (DAQ), a sample is placed below the membrane and acoustically coupled with commercial ultrasound gel. The amplitude and phase of the signal are analyzed by a laptop computer, which also controls the mechanical scanner.

Fig. 1

Schematic of the cw PA imaging system.

024006_1_019802jbo1.jpg

4.

Animal Protocol

New Zealand white rabbits with body weights between 2.8 and 3.5kg were used for the experiments. Immediately after the rabbits were sacrificed by an overdose of pentobarbital, the ears and legs were removed. Wounds were sealed with surgical glue (NEXABAND, Closure Medical Corporation, North Carolina). Hairs were shaved off with a veterinary shaving device. The tissue was placed on top of a 10-mm layer of acoustic impedance-matching rubber, which absorbs transmitted ultrasound. The tissue was then imaged by the cw PA imaging system. The samples were also photographed. All experimental animal procedures were carried out in conformity with the guidelines of the National Institutes of Health20 (NIH). The laboratory animal protocol for this work was approved by the School of Medicine Animal Studies Committee of Washington University at Saint Louis.

5.

Results and Discussion

The lower bound of the SNR is estimated by imaging an edge of a highly optically absorptive black polyethylene film in water using the cw PA imaging system. The image contrast is found to be 43dB , which can also be interpreted as the lower bound of the SNR. A similar experiment is conducted using the pulsed system, the image contrast is found to be 77dB . Therefore, the pulsed system yields a 34dB greater SNR than the cw system. This difference agrees with the aforementioned theoretical expectation.

The lateral resolution of the cw system is tested by imaging a 60-μm-diam human hair immersed in a 0.2% Intralipid® solution. The hair is oriented along the Y axis, which is normal to the transducer axis (Z) . The PA signal amplitude is plotted as a function of the Z and X axes in Fig. 2 , which is referred to as the line spread function. The lateral resolution is estimated to be 450μm , which is in good agreement with the theoretical prediction [ 0.61λ /numerical aperture (NA)=430μm , where λ=600μm is the acoustic wavelength and NA=0.85 is the NA of the spherical transducer]. The axial resolution at the 6-dB level is estimated to be 1mm , which matches approximately the 2mm full width at half maximum of the pressure sensitivity distribution along the acoustic axis in the focal zone of the transducer. The pressure sensitivity distribution can be computed21 by P(u)=P0[sin(u4)(u4)] , where P0 is the peak pressure sensitivity and u=(2πZλ)(NA) .

Fig. 2

Line spread function of the cw PA imaging system.

024006_1_019802jbo2.jpg

To demonstrate the feasibility of cw PA imaging for biomedical applications, we imaged vascular networks in rabbit ears.22 An amplitude image—a plot of the PA signal amplitude versus the scanned positions of the ultrasonic transducer—of a rabbit ear is shown in Fig. 3a , which shows the vasculature clearly. The scanning step size was 150μm in both the X and Y directions. Raster scanning was performed with a scanning speed of 10 steps (pixels)/s, whereas the integration time of the lock-in amplifier was 30ms . The rabbit ear was then transilluminated for photography of the vasculature [Figs. 3b and 3c]. The major arteries and veins were located between the central layer of cartilage and the dorsal skin. The skin overlying the imaged vasculature measured on average 300μm in thickness. Figure 3b shows a close-up of the vascular anatomy within the PA scanning area, which displays small branches from the central auricular branch and tributaries from the rostral auricular vein. The smallest vessel visualized by the cw PA imaging system in this specific region of the ear has a diameter of approximately 200μm , estimated from the photograph. There is good correspondence between the PA image and the photograph, although some artifacts are apparent. The artifacts are probably due to the interference between PA signals originating from different locations. In addition, ultrasonic reflections from the sample boundaries and the transducer surface as well as leakage of the laser-modulation electrical signal into the lock-in amplifier can also cause artifacts.

Fig. 3

(A) Amplitude image of a rabbit ear vasculature acquired with the cw PA imaging system, (B) photograph of the imaged region of the ear, and (C) photograph of the entire ear. The imaged region in (B) is marked with a frame.

024006_1_019802jbo3.jpg

On the basis of the Fourier transform, the cw mode with the modulation frequency swept over a bandwidth is equivalent to the pulsed mode with the same bandwidth. Therefore, PA sources can be localized by sweeping the modulation frequency within the frequency band of the ultrasonic transducer.15 In the frequency domain, the transducer output voltage V(ω) as a function of frequency ω is proportional to the PA pressure p(ω) multiplied by both the frequency response of the ultrasonic receiver and the phase shift due to the PA wave propagation distance z0 from the PA source to the ultrasonic transducer. The frequency-dependent ultrasonic attenuation in the medium can be included as well. Here, we consider only a single small-sized PA source although the same principle applies to a distribution of PA sources. If the frequency response of the resonant transducer is assumed to have a Gaussian shape, we have

Eq. 1

V(ω)p(ω)exp[(ωω0)2(Δω)2]exp(iωz0c).
By virtue of the Fourier transformation and the relation z=ct , Eq. 1 can be converted into the space domain:

Eq. 2

V(z)p(z)exp{[π(zz0)λΔωω0]2},
where λ denotes the wavelength of the PA wave at frequency ω0 and denotes convolution. We can see that the PA source is localized with a resolution related to the ultrasonic wavelength multiplied by a factor of ω0πΔω . The resolution can be related to the pulse width in the time domain multiplied by the speed of sound. In practice, the system frequency response can be measured by recording the amplitude and phase of the transducer output in response to a small test target placed at the focal spot of the transducer, and the continuous Fourier transformation can be implemented in its discrete form. Since Δωω0 for a resonant transducer, frequency sweeping will not be broad enough to achieve high spatial resolution. If a broadband transducer is used instead, the sensitivity will be reduced.

In some cases, a few frequencies are sufficient to remove some most significant imaging artifacts. Some interference artifacts are indicated by arrows in Fig. 4a . Here, we assume that the interference is between the PA signals from blood vessels and the front surface of the transducer. We use two frequencies shifted by Δf=c2z0 and simply subtract the two transducer outputs. The difference image is shown in Fig. 4b. As can be seen, this simple procedure significantly improves the image quality. For example, some blood vessels that show lengthwise contrast inversion in Fig. 4a, as indicated by arrows, become uniformly bright in Fig. 4b.

Fig. 4

Amplitude images of a rabbit ear vasculature acquired with the cw PA imaging system (A) before and (B) after removal of interference artifacts. Arrows indicate some much improved structures.

024006_1_019802jbo4.jpg

To demonstrate the capability of the cw system to image relatively deep structures, we further imaged a rabbit leg using the cw PA imaging system [Fig. 5a ]. Two anatomical cross sections were photographed for comparison [Fig. 5b]. The bright structures shown in Fig. 5a are the TA and TV. The diameters of the TA and TV are both 1mm as estimated by optical microscopy. The scanning area covers 30×20mm through 240 and 160 scanning steps in the X and Y directions, respectively. As revealed by Fig. 5b, the main branches of both the tibial artery and tibial vein seat below a 2.4 to 3.4-mm-thick layer of skin and muscle tissue.

Fig. 5

(A) Amplitude image of a rabbit tibial artery (TA) and a tibial vein (TV) acquired with the cw PA imaging system and (B) photographs of the anatomical cross sections indicated by S1 and S2 in (A) through the leg muscle.

024006_1_019802jbo5.jpg

The current cw system has relatively long data acquisition time. If the data acquisition time per pixel is 90ms , the scan time for a 10,000-pixel 2=D en face image using the current cw system can be as long as 15min . If only a 100-pixel 1=D image is needed, the acquisition time is shortened to 9s . Use of an ultrasound array can potentially speed up the data acquisition.

If depth resolution is not important in a particular application, cw imaging can serve as a viable alternative to pulsed mode imaging. Like pulsed PA imaging,2, 10 cw imaging can provide functional information of biological tissue as well if multiple laser diodes of different wavelengths are used for spectral measurements. Possible applications of the cw technique may include for example sentinel lymph node mapping23 and cerebral blood oxygenation monitoring24.

6.

Conclusions

In this feasibility study, we demonstrated that the cw PA imaging technique can image blood vessels 3mm deep in tissue. The cw system was built using an intensity-modulated cw laser diode. As a result, the imaging system is inexpensive, compact, and durable. Although a single-frequency cw measurement cannot provide the time-domain information about the PA time of flight, a spherically focused acoustic lens can provide spatial resolution. Using a bowl-shaped piezoelectric transducer with an NA of 0.85 and a resonance frequency of 2.45MHz , we imaged biological tissues with a lateral resolution of 450μm , an axial resolution of 1mm , and an SNR as high as 43dB .

Acknowledgments

We thank Z. Huang for experimental assistance. This project is sponsored by National Institutes of Health Grants R01 EB000712 and R01 NS46214.

References

1. 

M. Xu, and L. V. Wang, “Photoacoustic imaging in biomedicine,” Rev. Sci. Instrum., 77 041101 (2006). https://doi.org/10.1063/1.2195024 0034-6748 Google Scholar

2. 

X. Wang, Y. Pang, G. Ku, X. Xie, G. Stoica, and L. V. Wang, “Noninvasive laser-induced photoacoustic tomography for structural and functional imaging of the brain in vivo,” Nat. Biotechnol., 21 803 –806 (2003). https://doi.org/10.1038/nbt839 1087-0156 Google Scholar

3. 

C. G. A. Hoelen, F. F. M. de Mul, R. Pongers, and A. Dekker, “Three-dimensional photoacoustic imaging of blood vessels in tissue,” Opt. Lett., 23 648 –650 (1998). 0146-9592 Google Scholar

4. 

A. Oraevsky and A. Karabutov, “Optoacoustic tomography,” Biomedical Photonics Handbook, PM125 CRC Press, Boca Raton, FL (2003). Google Scholar

5. 

R. A. Kruger, P. Liu, Y. R. Fang, and C. R. Appledorn, “Photoacoustic ultrasound (PAUS) reconstruction tomography,” Med. Phys., 22 1605 –1609 (1995). https://doi.org/10.1118/1.597429 0094-2405 Google Scholar

6. 

A. G. Bell, “On the production and reproduction of sound by light,” Am. J. Sci., 20 305 –324 (1880). 0002-9599 Google Scholar

7. 

G. J. Diebold, M. I. Khan, and S. M. Park, “Photoacoustic ‘signature’ of particulate matter: optical production of acoustic monopole radiation,” Science, 250 101 –104 (1990). https://doi.org/10.1126/science.250.4977.101 0036-8075 Google Scholar

8. 

A. C. Tam, “Applications of photoacoustic sensing techniques,” Rev. Mod. Phys., 58 381 –431 (1986). https://doi.org/10.1103/RevModPhys.58.381 0034-6861 Google Scholar

9. 

F. A. Duck, Physical Properties of Tissue, Academic Press, London (1990). Google Scholar

10. 

H. F. Zhang, K. Maslov, G. Stoica, and L. V. Wang, “Functional photoacoustic microscopy for high-resolution and noninvasive in vivo imaging,” Nat. Biotechnol., 24 848 –851 (2006). https://doi.org/10.1038/nbt1220 1087-0156 Google Scholar

11. 

, (2000) Google Scholar

12. 

A. Oraevsky and A. Karabutov, “Ultimate sensitivity of time-resolved optoacoustic detection, biomedical optoacoustics,” Proc. SPIE, 3916 228 –239 (2000). https://doi.org/10.1117/12.386326 0277-786X Google Scholar

13. 

K. Maslov, G. Stoica, and L. V. Wang, “In vivo dark-field reflection-mode photoacoustic microscopy,” Opt. Lett., 30 625 –627 (2005). https://doi.org/10.1364/OL.30.000625 0146-9592 Google Scholar

14. 

T. J. Allen and P. C. Beard, “Pulsed near-infrared laser diode excitation system for biomedical photoacoustic imaging,” Opt. Lett., 31 3462 –3464 (2006). https://doi.org/10.1364/OL.31.003462 0146-9592 Google Scholar

15. 

T. W. Murray and O. Balogun, “High-sensitivity laser-based acoustic microscopy using a modulated excitation source,” Appl. Phys. Lett., 85 2974 –2476 (2004). https://doi.org/10.1063/1.1802387 0003-6951 Google Scholar

16. 

Y. Fan, A. Mandelis, G. Spirou, and I. A. Vitkin, “Development of a laser photothermoacoustic frequency-swept system for subsurface imaging: theory and experiment,” J. Acoust. Soc. Am., 116 3523 –3533 (2004). https://doi.org/10.1121/1.1819393 0001-4966 Google Scholar

17. 

I. G. Calasso, W. Craig, and G. J. Diebold, “Photoacoustic point source,” Phys. Rev. Lett., 86 3550 –3553 (2001). https://doi.org/10.1103/PhysRevLett.86.3550 0031-9007 Google Scholar

18. 

G. J. Diebold, T. Sun, and M. I. Khan, “Photoacoustic monopole radiation in one, two, and three dimensions,” Phys. Rev. Lett., 69 3384 –3387 (1991). https://doi.org/10.1103/PhysRevLett.67.3384 0031-9007 Google Scholar

19. 

M. Robert, G. Molingou, and K. Snook, “Fabrication of focused polyvinylidene fluoride-trifluoroethylene PVDF-TrFE copolymer 4050MHz ultrasound transducers on curved surfaces,” J. Appl. Phys., 96 252 –257 (2004). https://doi.org/10.1063/1.1760233 0021-8979 Google Scholar

20. 

National Institutes of Health, Guide for the Care and Use of Laboratory Animals, U.S. Government Printing Office, Washington, DC (1985). Google Scholar

21. 

M. Born and E. Wolf, Principles of Optics: Electromagnetic Theory of Propagation, Interference and Diffraction of Light, 436 5th ed.Pergamon Press, Oxford,1975). Google Scholar

22. 

R. I. Siphanto, R. G. Kolkman, A. Huisjes, M. C. Pilatou, F. F. de Mul, W. Steenbergen, and L. N. van Adrichem, “Imaging of small vessels using photoacoustics: an in vivo study,” Lasers Surg. Med., 35 354 –362 (2004). https://doi.org/10.1002/lsm.20100 0196-8092 Google Scholar

23. 

A. Darzi and S. Mackay, “Recent advances in minimal access surgery,” Br. Med. J., 324 31 –34 (2002). https://doi.org/10.1136/bmj.324.7328.31 0007-1447 Google Scholar

24. 

Y. Y. Petrov, D. S. Prough, D. J. Deyo, M. Klasing, M. Motamedi, and R. O. Esenaliev, “Optoacoustic, noninvasive, real-time, continuous monitoring of cerebral blood oxygenation: an in vivo study in sheep,” Anesthesiology, 102 69 –75 (2005). https://doi.org/10.1097/00000542-200501000-00014 0003-3022 Google Scholar
©(2008) Society of Photo-Optical Instrumentation Engineers (SPIE)
Konstantin I. Maslov and Lihong V. Wang "Photoacoustic imaging of biological tissue with intensity-modulated continuous-wave laser," Journal of Biomedical Optics 13(2), 024006 (1 March 2008). https://doi.org/10.1117/1.2904965
Published: 1 March 2008
Lens.org Logo
CITATIONS
Cited by 141 scholarly publications and 25 patents.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Transducers

Imaging systems

Ultrasonics

Signal to noise ratio

Continuous wave operation

Photoacoustic imaging

Semiconductor lasers

Back to Top