1 January 2009 Compact portable ocular microtremor sensor: design, development and calibration
James Patrick Ryle, Mohammed Ali Al-Kalbani, Niamh Collins, Unnikrishnan Gopinathan, Gerard Boyle, Davis Coakley, John T. Sheridan
Author Affiliations +
Abstract
Ocular microtremor (OMT) is a physiological high-frequency (up to 150 Hz) low-amplitude (25–2500 nm peak-to-peak) involuntary motion of the human eye. Recent studies suggest a number of clinical applications for OMT that include monitoring the depth of anesthesia of a patient in surgery, prediction of outcome in coma, and diagnosis of brain stem death. Clinical OMT investigations to date have used mechanical piezoelectric probes or piezoelectric strain gauges that have many drawbacks which arise from the fact that the probe is in contact with the eye. We describe the design of a compact noncontact sensing device to measure OMT that addresses some of the above drawbacks. We evaluate the system performance using a calibrated piezoelectric vibrator that simulates OMT signals under conditions that can occur in practice, i.e., wet eye conditions. We also test the device at low light levels well within the eye safety range.
©(2009) Society of Photo-Optical Instrumentation Engineers (SPIE)
James Patrick Ryle, Mohammed Ali Al-Kalbani, Niamh Collins, Unnikrishnan Gopinathan, Gerard Boyle, Davis Coakley, and John T. Sheridan "Compact portable ocular microtremor sensor: design, development and calibration," Journal of Biomedical Optics 14(1), 014021 (1 January 2009). https://doi.org/10.1117/1.3083435
Published: 1 January 2009
Lens.org Logo
CITATIONS
Cited by 12 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Eye

Speckle

Signal to noise ratio

Calibration

Safety

Sensors

Signal processing

Back to Top