1.IntroductionThe development and use of intraocular lens (IOL) implants for refractive cataract surgery have changed the life of cataract patients and have become the most commonly performed surgical procedure since the first IOL implantation demonstrated by Ridley Harold in 1949,1 with an estimated 3 million surgeries per year in the United States2 and 20 million worldwide.3 Among many physical parameters that determine key performance characteristics, safety, and quality of an IOL, the focal length (or dioptric power), modulation transfer function (MTF), scattering, astigmatism, thickness, and refractive index play a critical role. MTF, astigmatism, and dioptric power can be measured using commercially available IOL-testing equipment. Recently, our group has developed and implemented test methodologies to measure dioptric power of various IOL designs with high precision.4–7 However, despite their significance in IOL characterization,8 conventional techniques employed for IOL thickness and refractive-index measurements show some specific drawbacks and limitations in precise IOL testing. For instance, the use of simple mechanical devices such as vernier calipers or micrometers for IOL thickness measurement can affect the IOL optical quality caused by the contact measurement, and the lens curvature introduces additional inaccuracies for precise measurement. Commonly used refractometers are also not suitable for the IOL refractive-index testing due to the same type of contact-based measurement. Therefore, the development and implementation of alternative noncontact sensing principles and measurement methods using remote, contact-free optical, or ultrasound irradiation sources have been considered to provide safer, more effective, and higher measurement accuracy approaches.3,7 The use of optical measurement tools is especially attractive due to the well-developed optical imaging, microscopic, and sensing techniques providing high measurement accuracy. There have been a number of published reports on the application of optical tools for measurement of thickness,9,10 refractive index,11,12 or both13–16 of transparent media or biological samples. These tools, however, impose some specific limitations for testing thickness and refractive index of IOL in terms of spatial sample alignment, testing accuracy, variations of IOL designs, and calibration requirements. Some of these approaches are based on optical interferometry9,10,12 that will have a limited applicability for testing of both IOL thickness and refractive index due to the complex measurements, the critical spatial alignment requirement of the IOL center to the beam axis, and the availability of IOL designs with different shapes (convex, concave, etc.). Optical coherence tomography (OCT)-based methods11,14 do not provide superior testing accuracy compared to confocal microscope method and have the OCT limitation in measuring thickness of nontransparent objects. The conventional confocal microscopy-based platforms13,16 have also some limitations for testing IOL in terms of confocal design and calibration requirements. In this work, we present a noncontact method for precise IOL thickness and refractive-index measurement employing an advanced fiber-optic self-calibrating dual-confocal laser caliper approach. The suggested sensing method can be applied for self-calibrating thickness measurement of any IOL thickness, shape, and transparency, and for the refractive-index measurement of a lens with surface curvature providing high accurate measurements for IOL. 2.Experimental Sensing ApproachWe developed a fiber-optic-based dual-confocal laser sensing system shown in Fig. 1.17 As for other similar sensor methods, the measurement accuracy is the key characteristic that is highly dependent on the alignment of laser beams and the calibration of distances between the optical components. Below, we present a step-by-step procedure for setting up and the alignment of the proposed dual-confocal laser caliper system.18 As a laser source, we used a solid-state diode laser with a center wavelength at 658 nm, an output power of 25 mW, and a beam diameter of 8.0 mm. The laser characteristics affect the overall performance of the confocal microscope sensor; the wavelength is directly related to the resolution, the larger beam diameter compared to a typical 1-mm diameter of He–Ne laser enables a high laser-to-fiber coupling efficiency due to the reduced size of the focused beam spot that is moved toward matching to the relatively small input area of the single-mode fibers used, and the laser low coherence reduces instability of the output power due to the coherent interference between the optical elements. The laser beam is launched into a 50:50 single-mode fiber coupler, and it is divided into two arms (O1 and O2, Fig. 1), which are configured as two identical fiber-optic confocal microscopes. The fiber coupler consists of single-mode fibers with a core diameter. Instead of using a single lens for the confocal microscope, we employed a collimating and focusing lens pair configuration that provides broader capabilities for precise collimating, focusing, and alignment of the confocal laser beams. Without using the focusing lenses (LF1 and LF2, Fig. 1), laser outputs from both arms of the fiber coupler are collimated using identical () microscope objective lenses, LC1 and LC2. By mounting the fiber-lens assemblies on tilting and translational mechanical stages, two collimated laser outputs were aligned to face each other on the same beam axis. This was performed by maximizing the measured laser intensity at the detector. The measured intensity was 7.4 mW, which provides 59% of overall laser-to-fiber coupling efficiency, when assumed only a half of the intensity will be delivered to the detector through 50:50 fiber coupler. Following the alignment of beam axes, a total-reflectance flat mirror was placed at the target position (T, Fig. 1), the reflective surface facing the collimating lens LC1. The mirror was adjusted to be normal to the beam axis, which was again performed by maximizing the intensity at the detector. This is to ensure the next step: the alignment of the focusing objective lens LF1 to the beam axis. Now, placing the LF1 at a position that is a focal length away from the mirror, we can detect the confocal reflection signal from the mirror, and then by maximizing the signal intensity the optimal alignment can be achieved. This procedure was repeated for the other confocal microscope, by placing the mirror facing the opposite side and placing the focusing lens LF2. The distance between the two confocal microscopes, eventually the distance between LF1 and LF2, depends on the working distance (WD) of LF1 (identical lens is used for LF2) and the thickness of the object to be measured. For example, when lenses (, ) were used, the distance between LF1 and LF2 must be () to measure a thickness of about 1 mm. The most convenient method is to place two lenses as far apart as possible within the travel limit of the mechanical stage on which a measurement sample will be mounted; however, the farther they are apart the more time it takes for measurements. The operating principle of the dual-confocal laser caliper is straightforward (Fig. 2). Let us assume the distance between LF1 and LF2 is . We placed a thin plate (OBJECT, Fig. 2) with parallel surface, which also has a certain degree of reflectivity on both surfaces, between LF1 and LF2. The thickness of the plate, , is known with high accuracy. The surface of the plate can be adjusted to be normal to the beam axis using the same procedure as described in the previous paragraph. The plate is mounted on a mechanical translational stage that moves along the beam axis (-axis, Fig. 2). Limiting the travel distance of the plate from close proximity of LF1 and to that of LF2, confocal reflection signals were recorded while the plate was moved from left to right. A silicon photodiode combined with an analog-to-digital converter was used as a detector. If the plate is not transparent, there will be one peak when the left surface of the plate is at (, Fig. 2), where the left surface of the plate is a focal length of LF1 () away from the lens. Another peak will be observed at (, Fig. 2), when the right surface of the plate is a focal length of LF2 () away from the lens. Now, the distance between LF1 and LF2 can be expressed as a function of other parameters described above If we repeat the same procedure using an object with unknown thickness (), the two peaks will be observed at and . Equation (1) is now expressed as By comparing Eqs. (1) and (2), we obtain an expression for the unknown thickness in terms of known and measured parameters Equation (3) suggests that we can measure the thickness of any unknown object, which has well-defined reflecting surfaces on both sides, after calibrating the distance between optical components using a plate of known thickness. This method does not require precise determination of focal lengths of lenses and the distance between them, and therefore, the possible errors are reduced. Two additional peaks will be observed when a transparent object with a thickness smaller than the WD of focusing lenses is used. As can be seen from Fig. 2, there is a backreflection from the right surface of the object at . This is due to the focusing through the transparent object. However, is not away from , since the refractive index of the object () increases the effective focal length of LF1. If we designate the distance between and as , the refractive index of the object can be calculated using other parameters13 Another peak observed at can also be used for the calculation by letting represent the distance between and . This is valid only when the measured object has inversion symmetry. The IOL thickness measurement requires an additional alignment. Unlike a parallel plate, the IOL not only needs to be placed normal to the beam axis but also the center of the IOL must coincide with the beam axis since the center is the thickest portion regardless if it is planoconvex or double-convex. Thickness measurement of negative-power IOL is also possible; however, it will not be discussed in this study. After the alignment of two collimating lenses, LC1 and LC2, a flat total-reflectance mirror (M) was placed to be normal to the beam axis (see Fig. 1). The position of the mirror M is approximately a focal length of the IOL away from T. Using one portion of the dual-confocal setup (O1), the dioptric power of the IOL is measured by placing the IOL at T. This method not only provides a high-precision measurement of the IOL dioptric power but also ensures the IOL alignment to the beam axis.5 After the power measurement, the mirror is removed and the two focusing lenses are placed back at LF1 and LF2. The rest of the measurement procedure is the same as that for a plate. Refractive index can also be measured when the thickness of the IOL is smaller than the WD of the lenses; however, the measured values using () and () are different when a planoconvex IOL is used. 3.Results and DiscussionsFigure 3 shows a confocal reflection signal measured for a 1-mm standard metal plate using objective lenses. The values and , as defined in the previous section, were 0.1927 and 0.9565 mm, respectively. The full-width at half-maximum (FWHM) of a peak was and the precision of the mechanical translational stage was . FWHM of our setup was larger than the theoretically predicted value of , which is calculated using an expression for the axial response of a confocal microscope 19 where is the laser wavelength and is the numerical aperture of the focusing lens. The setup showed a lower axial resolution compared to the theoretically predicted value since the setup was operated under the fiber-dominated mode for increased coupling efficiency by using a larger NA (0.25) of the collimating lens compared to that of the optical fiber NA (0.12).20Figure 4 shows data measured for a thin glass plate using objective lenses. The thickness of the glass plate was 1.00 mm as measured with a vernier caliper as one of the standard techniques listed in Sec. 1 using a mechanical tool to measure the IOL thickness providing an accuracy of . Four peaks were observed as the thickness of the plate is smaller than the WD of the lens. The intensity peaks at , , , and are measured to be 0.1415, 0.7967, 1.5406, and 2.1976 mm, respectively. The thickness and refractive index of the glass plate can be calculated using Eqs. (3) and (4): is 1.0199 mm and is 1.482. Both results match well to the known values. Calculated values of using () and () were 0.6552 and 0.6570 mm, respectively. The difference is , and it is well within the experimental error. The intensity peaks at and were smaller than those at and , because the direct reflection from the front surface is stronger than that from the back surface. Figure 5 shows data measured for a planoconvex IOL using objective lenses, planar surface facing LF1. The thickness of the IOL was 0.76 mm when measured with a vernier caliper. Four peaks at , , , and are measured to be 0.1266, 0.6406, 1.6297, and 2.1597 mm, respectively. The thickness and the refractive index calculated using Eqs. (3) and (4) were 0.7747 and 1.438, respectively. Both results match well to the known values. The calculated values of using () and () were 0.5140 and 0.5300 mm, respectively. The difference is , which is larger than the difference measured from the glass plate. It is because the reflections measured from LF1 (peaks at and ) are through the planar surface; however, the reflections measured from LF2 (peaks at and ) are through the convex surface. Additional focusing effect from the convex surface has decreased the effective focus of LF2, thus, the peak was observed when the object was closer to LF2, giving the larger value of () than (). We used () as for the calculation of refractive index. The intensity at is also smaller than that at . We assumed the convex surface is locally flat when it is at the laser focus (focal spot diameter is ). The IOL dioptric power measured during the alignment was . For comparison, the measurement was performed using objective lenses. Figure 6 shows the measured data for a standard 1-mm metal plate. and were 0.1365 and 1.2136 mm, respectively. Because of high NA of the lenses, FWHM was reduced to . Figure 7 shows data measured for the same IOL. The intensity peaks at and are measured to be 0.0906 and 1.3976 mm, respectively. The calculated IOL thickness was 0.7700 mm, which is closer to the value measured with a vernier caliper than when it was measured with lenses. Since the WD of objective lens is only 0.28 mm, the refractive index could not be measured. Objective lenses with high NA and long WD such as infinity-corrected lenses could be used when the measurement of refractive index while maintaining high accuracy is necessary. A few general factors that could limit the measurement accuracy of the presented method are discussed as follows.
4.ConclusionsIn conclusion, we presented a fiber-optic dual-confocal sensing method for noncontact high-precision measurement of thickness of an object with reflective surfaces. The method is proved to be specifically useful for the thickness measurement of IOL implants, where contact-free measurement is highly desirable. We have successfully measured the IOL thickness with accuracy as high as , when single-mode fibers, objective lenses, and 658-nm laser source were used. Since the alignment of optical components is critical in the proposed method, we developed a step-by-step procedure for the optical alignment to reduce any possible errors due to the misalignment. The measured IOL thickness using our method was 0.7700 mm, which agrees well with the value measured by the conventional mechanical methods. In addition, refractive index of an IOL is also measured using objective lenses, and the measured value was 1.438. The method is not only limited to the thickness and refractive-index measurement of thin transparent object but it can also be effectively employed to objects with any thickness and transparency if the thickness is within the travel limit of translational stage and the object has measurable backreflection at the surfaces. Furthermore, due to the high precision of the noncontact method, measurements can be performed without any physical contact on the sample, which ensures that the sample can be preserved in its original conditions. In addition, fragile or soft material, and tissue can also be used for the thickness measurement. DisclosuresThe authors have no relevant financial interests in this article and no potential conflicts of interest to disclose. AcknowledgmentsA basic part of this work was undertaken when D.-H.K. was a visiting scientist at the US Food and Drug Administration, Silver Spring, Maryland. The mention of commercial products, their sources, or their use in connection with material reported herein is not to be construed as either an actual or implied endorsement of such products by the Department of Health and Human Services. ReferencesD. Apple and J. Sims,
“Harold Ridley and the invention of the intraocular lens,”
J. Surv. Ophthalmol., 40 279
–292
(1996). https://doi.org/10.1016/S0039-6257(96)82003-0 Google Scholar
N. Congdon et al.,
“Prevalence of cataract and pseudophakia/aphakia among adults in the United States,”
Arch. Ophthalmol., 122 487
–494
(2004). https://doi.org/10.1001/archopht.122.4.487 AROPAW 0003-9950 Google Scholar
Global Intraocular Lens (IOL), “Market 2013-18: industry nanotechnology analysis, size, share, strategies, growth, trends and forecast research report,”
PRWeb,
(2018) http://www.prweb.com/pdfdownload/10886563.pdf January ). 2018). Google Scholar
I. K. Ilev,
“A simple confocal fibre-optic laser method for intraocular lens power measurement,”
Eye, 21 819
–823
(2007). https://doi.org/10.1038/sj.eye.6702406 12ZYAS 0950-222X Google Scholar
B. Walker et al.,
“Assessing the effect of laser beam width on quantitative evaluation of optical properties of intraocular lens implants,”
J. Biomed. Opt., 19 055004
(2014). https://doi.org/10.1117/1.JBO.19.5.055004 JBOPFO 1083-3668 Google Scholar
B. Walker et al.,
“Confocal laser method for quantitative evaluation of critical optical properties of toric intraocular lenses,”
J. Cataract Refractive Surg., 42 455
–461
(2016). https://doi.org/10.1016/j.jcrs.2015.09.031 JCSUEV 0886-3350 Google Scholar
J. Colliac,
“Matrix formula for intraocular lens power calculation,”
Invest. Ophthalmol. Visual Sci., 31 374
–381
(1990). IOVSDA 0146-0404 Google Scholar
P. Groot et al.,
“Optical interferometry for measurement of the geometric dimensions of industrial parts,”
Appl. Opt., 41 3853
–3860
(2002). https://doi.org/10.1364/AO.41.003853 APOPAI 0003-6935 Google Scholar
C. Poilane and P. Sandoz,
“Thickness measurement of nontransparent free films by double-side white-light interferometry: calibration and experiments,”
Rev. Sci. Instrum., 77 055102
(2006). https://doi.org/10.1063/1.2198707 RSINAK 0034-6748 Google Scholar
G. J. Teamey et al.,
“Determination of the refractive index of highly scattering human tissue by optical coherence tomography,”
Opt. Lett., 20 2258
–2260
(1995). https://doi.org/10.1364/OL.20.002258 OPLEDP 0146-9592 Google Scholar
K. Joo and S. Kim,
“Refractive index measurement by spectrally resolved interferometry using a femtosecond pulse laser,”
Opt. Lett., 32 647
–649
(2007). https://doi.org/10.1364/OL.32.000647 OPLEDP 0146-9592 Google Scholar
I. K. Ilev et al.,
“Dual-confocal fiber-optic method for absolute measurement of refractive index and thickness of optically transparent media,”
Opt. Lett., 27 1693
–1695
(2002). https://doi.org/10.1364/OL.27.001693 OPLEDP 0146-9592 Google Scholar
X. Wang et al.,
“Simultaneous refractive index and thickness measurements of bio tissue by optical coherence tomography,”
J. Biomed. Opt., 7 628
–632
(2002). https://doi.org/10.1117/1.1501887 JBOPFO 1083-3668 Google Scholar
Z. Jian, C. Hsu and D. Su,
“Improved technique for measuring refractive index and thickness of a transparent plate,”
Opt. Commun., 226
(6), 135
–140
(2003). https://doi.org/10.1016/j.optcom.2003.09.010 OPCOB8 0030-4018 Google Scholar
M. Ohmi et al.,
“High-speed simultaneous measurement of refractive index and thickness of transparent plates by low-coherence interferometry and confocal optics,”
Meas. Sci. Technol., 15 1531
–1535
(2004). https://doi.org/10.1088/0957-0233/15/8/017 MSTCEP 0957-0233 Google Scholar
D. Kim, I. Ilev and Y. Han,
“A thickness measurement method for biological samples using lensed-fiber sensors,”
Proc. SPIE, 7894 789416
(2011). https://doi.org/10.1117/12.881640 PSISDG 0277-786X Google Scholar
T. Gorle, G. Chou and G. Kino,
“Depth response of confocal optical microscopes,”
Opt. Lett., 11 770
–772
(1986). https://doi.org/10.1364/OL.11.000770 OPLEDP 0146-9592 Google Scholar
T. Dabbs and M. Glass,
“Fiber-optic confocal microscope: FOCON,”
Appl. Opt., 31 3030
–3035
(1992). https://doi.org/10.1364/AO.31.003030 APOPAI 0003-6935 Google Scholar
M. D. Shanna and C. J. R. Sheppard,
“Effects of system geometry on the axial response of the fibre-optical confocal microscope,”
J. Mod. Opt., 46 605
–621
(1999). https://doi.org/10.1080/09500349908231288 JMOPEW 0950-0340 Google Scholar
I. Ilev et al.,
“Simple fiber-optic confocal microscopy with nanoscale depth resolution beyond the diffraction barrier,”
Rev. Sci. Instrum., 78 093703
(2007). https://doi.org/10.1063/1.2777173 RSINAK 0034-6748 Google Scholar
BiographyDo-Hyun Kim finished his PhD in solid state physics in 2000 at Seoul National University in the Republic of Korea. He finished his second PhD in electrical engineering with biomedical optics as the topic of his research at Johns Hopkins University in 2006. He has performed research on biomedical optics with special interest in optical radiation safety and optical microscopy at the US Food and Drug Administration. He is a senior member of SPIE. Robert James has more than 40 years of experience in the US Food and Drug Administration, where he has developed laboratory evaluation procedures for optical radiation safety of light emitting products, medical lasers, and intraocular lenses. He has published in peer-reviewed journals, presented at conferences, and has been active in ANSI and ISO standards committees. In recent years he transitioned away from supervisory duties to a part-time research position concentrating on optical medical devices. Don Calogero has a master’s degree in biomedical engineering from the State University of New York at Stony Brook and has 38 years of experience evaluating intraocular devices such as intraocular lenses prior to their clinical investigation and FDA approval. He has published in peer-reviewed journals, presented at major national and international ophthalmic meetings, and has led the development of ANSI and ISO intraocular implant standards. Ilko K. Ilev is the leader of Optical Therapeutics and Medical Nanophotonics Laboratory, and a Senior Biomedical Research Service (SBRS) Scientist at FDA. He has a PhD in quantum and laser physics. He has over 25 years of experience in the United States, Europe, and Japan, with more than 415 publications in the field of laser technologies, biophotonics, nanobiophotonics, bioimaging, biosensing, and laser safety. He is an elected fellow of IEEE, OSA, SPIE, and ASLMS. |