11 September 2023 Multifocal acoustic radiation force-based reverberant optical coherence elastography for evaluation of ocular globe biomechanical properties
Author Affiliations +
Abstract

Significance

Quantifying the biomechanical properties of the whole eye globe can provide a comprehensive understanding of the interactions among interconnected ocular components during dynamic physiological processes. By doing so, clinicians and researchers can gain valuable insights into the mechanisms underlying ocular diseases, such as glaucoma, and design interventions tailored to each patient’s unique needs.

Aim

The aim of this study was to evaluate the feasibility and effectiveness of a multifocal acoustic radiation force (ARF) based reverberant optical coherence elastography (RevOCE) technique for quantifying shear wave speeds in different ocular components simultaneously.

Approach

We implemented a multifocal ARF technique to generate reverberant shear wave fields, which were then detected using phase-sensitive optical coherence tomography. A 3D-printed acoustic lens array was employed to manipulate a collimated ARF beam generated by an ultrasound transducer, producing multiple focused ARF beams on mouse eye globes ex vivo. RevOCE measurements were conducted using an excitation pulse train consisting of 10 cycles at 3 kHz, followed by data processing to produce a volumetric map of the shear wave speed.

Results

The results show that the system can successfully generate reverberant shear wave fields in the eye globe, allowing for simultaneous estimation of shear wave speeds in various ocular components, including cornea, iris, lens, sclera, and retina. A comparative analysis revealed notable differences in wave speeds between different parts of the eye, for example, between the apical region of the cornea and the pupillary zone of the iris (p = 0.003). Moreover, the study also revealed regional variations in the biomechanical properties of ocular components as evidenced by greater wave speeds near the apex of the cornea compared to its periphery.

Conclusions

The study demonstrated the effectiveness of RevOCE based on a non-invasive multifocal ARF for assessing the biomechanical properties of the whole eyeball. The findings indicate the potential to provide a comprehensive understanding of the mechanical behavior of the whole eye, which could lead to improved diagnosis and treatment of ocular diseases.

1.

Introduction

Knowledge of the biomechanical properties of the various interconnected ocular components is critical for understanding the physiology of the eye globe, detecting ocular diseases, and designing effective treatments and surgical procedures. For example, measuring the biomechanical properties of the cornea and sclera can help detect the early signs of glaucoma, a condition that can be characterized by increased intraocular pressure (IOP).1 By understanding how different ocular components work together to resist changes in IOP, clinicians can assess glaucoma risk and develop effective treatment strategies for managing glaucoma. Furthermore, biomechanical models of the eye can be used to design personalized therapeutic procedures for implanting intraocular lenses or performing corneal transplant surgeries, ensuring that the procedures are safe and effective.

Recent advances in innovative techniques for measuring the biomechanical properties of individual ocular tissues hold great promise for improving our understanding of ocular biomechanical properties. For example, the ocular response analyzer,2,3 CorVis,4 Brillouin microscopy,5,6 and optical coherence elastography (OCE)7 have shown potential for quantifying the biomechanical properties of individual ocular tissues with high precision and accuracy. However, there is still a gap in the methods used for detecting the biomechanical properties of various ocular components simultaneously. While it is important to understand the biomechanical properties of individual ocular tissues, the whole eye globe is a complex and interconnected system. The interactions between different ocular components play a crucial role in determining the overall ocular functions.1 For instance, previous research utilizing finite element analysis has demonstrated that the lens, iris, and other ocular muscles contribute significantly to the mechanical aspects of corneal deformation.8 Therefore, it is essential to determine the mechanical properties of the interrelated parts, such as the cornea, lens, sclera, iris, and retina, for an accurate understanding of the eye’s mechanical behavior.

Several studies have demonstrated promising outcomes in estimating the biomechanical properties of multiple ocular components simultaneously. One such study involved measuring displacement in response to eye globe inflation using multiple cameras and a laser displacement sensor to enable the assessment of regional variations in the stiffness of the cornea and sclera.9 However, this study was limited to the surface of the eye globe, and internal parts, such as the lens and iris, were not assessed. On the other hand, high-field magnetic resonance imaging has been used to assess the displacement of the whole eyeball with a reasonable resolution, but its clinical application is limited due to long scan times and motion artifacts.1 Ultrasound elastography (USE) is another promising method and has been used to determine the biomechanical properties of the sclera and optic nerve simultaneously.10 A recent study presented the first comprehensive elasticity assessment of the whole eyeball using USE, characterizing the biomechanical properties of the cornea, lens, iris, optic nerve hypoplasia, and peripapillary sclera.11 However, this method involved a mechanical shaker to induce deformation in the tissues, which may cause discomfort during in vivo application and hence may influence measurement results. Additionally, the resolution of USE is relatively low.

OCE12,13 has emerged as a high-resolution technique to quantify the biomechanical properties of multiple ocular components simultaneously. For example, OCE has been used to study the biomechanical properties of the anterior eye (cornea, limbus, and anterior sclera simultaneously),14 the lens and cornea,15 and the cornea and retina.16 Although these studies provided valuable insights into the biomechanical properties of different ocular components, measurements were limited to only a few ocular components at a time due to the limited tissue penetration depth. The recent introduction of reverberant OCE (RevOCE), which assumes the existence of complex three-dimensional reverberant shear wave fields in the tissue,17 aims to overcome the above limitations.18 However, conventional mechanical shakers employed in existing RevOCE techniques that directly contact tissues may not be suitable for assessing sensitive (delicate) ocular tissues.

In this context, the aim of the current study is to demonstrate the capability of a non-invasive RevOCE in assessing the biomechanical properties of the entire eye globe with high mechanical resolution and contrast. We present a multifocal acoustic radiation force (ARF) system, which was implemented using an array of acoustic lenses and a single-element ultrasound transducer as an excitation source to induce reverberant shear wave field in mouse eyeballs.19 The induced reverberant shear wave field was then imaged with phase-sensitive optical coherence tomography (PhS-OCT) to obtain a volumetric wave speed map of the entire eye globes. Shear wave speeds were quantified in different ocular components, including the cornea, iris, sclera, lens, and retina, yielding a comprehensive understanding of not only the relative stiffness variations among these interconnected parts but also the regional heterogeneity within each ocular component (e.g., cornea).

2.

Materials and Methods

2.1.

Mouse Eyeball Samples

Experiments were conducted on three freshly exercised mouse eyeballs (within 2 h after euthanasia). The eyeballs were cleaned of extraocular tissues, maintained in 1 X phosphate-buffered saline (PBS) solution, and frequently hydrated during imaging. The average IOP of the eyeballs, as measured by a rebound tonometer (Tonolab, iCare Finland Oy, Finland) before the start of OCE experiments, was 8.8  mmHg; thus, no IOP control was implemented.

2.2.

Systems and Data Acquisition

As shown in Fig. 1, the RevOCE system consisted of a lens-transducer system to generate multifocal ARF 19 and a PhS-OCT system to detect the resulting vibrations in the ocular tissues. A collimated ARF beam was produced by a 13 mm diameter single-element, 3.5 MHz ultrasound transducer (C382-SU, Olympus Co., Japan) and was coupled to an acoustic lens array composed of three plano-concave lenses. The individual lens geometries were specifically designed with a focal length of 9.7 mm, an aperture diameter of 6.6 mm, and a working distance of 6.4 mm. The lens array, which was 3D-printed using rubber-like material (TangoGray FLX950, Stratasys, Israel), was designed to produce three focused ARF beams with a spatial separation of 2.65  mm in the focal plane. Measurements with a hydrophone of 0.2 mm sensor diameter (NH0200, Precision Acoustics Ltd, United Kingdom) confirmed the presence of the three focal spots separated by 2.65  mm. The spot size, defined as the full width half maximum of the normalized pressure intensity at the focal plane, was determined to be 1.13 mm. The induced reverberant shear wave field was detected using a PhS-OCT system characterized by displacement stability, axial resolution in air, and transverse resolution of 0.28 nm, 9  μm, and 8  μm, respectively.20 The OCT system was operated at an A-line rate of 25 kHz. In prior research, we used seven foci ARF to generate reverberant shear wave fields.21 Despite utilizing only three foci excitation in the current study to accommodate the size of a mouse eyeball, our preliminary findings and prior studies22 confirm that using three or more sources can effectively generate reverberant shear wave fields in tissues.

Fig. 1

Schematic of the RevOCE system consisted of an ultrasound transducer coupled with array of acoustic lenses to produce multifocal ARF and a PhS-SDOCT system for imaging elastic wave propagation induced by the ARF. Amp, RF amplifier; C, collimator; FC, fiber coupler; FG, function generator; GS, 2D galvo scanner; M, reference mirror; P, pinhole; PC, polarization controller; S, spectrometer; SL, scan lens; and SLD, a superluminescent diode. The inset to the right is a 3D view of the acoustic lens and transducer system showing the multifocal ARF beams used to induce reverberant wave fields in mouse eyeballs that were mounted on a gelatin phantom (2% w/w concentration). The ARF beams were focused near the eyeball equator, aligned along the dashed lines surrounding its perimeter and positioned 120  deg apart from each other around the circumference. 3D OCE data were acquired by the 2D galvo scanner motion in the x-y plane. The sample setup was translated manually along the z-axis to repeat the 3D OCE measurement at the posterior of the eyeball.

JBO_28_9_095001_f001.png

During OCE imaging, the eyeball was mounted on a soft gelatin phantom (inset in Fig. 1, 2% w/w concentration), which had very low acoustic attenuation. To induce reverberant shear wave fields in the mouse eyeball, an arbitrary function generator (DG4162, RIGOL Tech, China) produced a 3.5 MHz sinusoidal signal modulated by ten cycles of a 3 kHz rectangular tone burst, which was then amplified by a power amplifier (1040L, Electronics & Innovation, Ltd., United States) feeding the transducer. Our selection of 3 kHz was a careful compromise between generating shorter excitation wavelengths necessary for inducing multiple internal reflections and avoiding excessive attenuation of waves. The excitation was synchronized with the OCT system frame trigger during OCT M-C mode scans, a technique utilized to obtain four-dimensional (4D) data in OCE, with “M” representing motion (temporal data) and “C” denoting volumetric (3D spatial data). This method involves capturing multiple A-scans at the same transverse location over time through M-mode scanning, which is repeated at every lateral location using a raster scanning procedure. In this study, the scan acquired 151×151 points over a lateral region of interest of 3.8 mm by 4.2 mm, with each M-mode scan consisting of 400 A-lines. As the OCT system axial imaging range was limited to 2.5  mm, we scanned the entire eyeball using a two-step imaging process. First, we adjusted the OCT imaging head position to enable RevOCE measurements on the anterior eye. Next, the imaging head was axially translated to focus on the posterior eye, and the RevOCE measurements were repeated.

Fig. 2

(a) A 3D OCT scan reconstruction of a mouse eyeball, illustrating its representative gross morphology. (b) Top: the volumetric image of the eyeball with the z-axis indicating the anterior-posterior (AP) direction which is along the optical axis of the OCT objective beam and eyeball optical axis. Bottom: a cross-sectional slice taken in plane C.

JBO_28_9_095001_f002.png

2.3.

Data Processing and Analysis

The 4D OCE data (3D spatial and 1D temporal) from the two-step measurement were processed to reconstruct a 3D map of the elastic wave speed in the whole eyeball using a custom MATLAB® R2021a (The MathWorks, Inc., United States) program. First, the phase shift, ΔØ, between successive A-scans was used to compute the axial particle velocity, vz=λoΔØ/(4πnΔτ), where λo=840  nm was the center wavelength of the OCT light source, n=1.46 was the refractive index of the ocular tissues, and Δτ=40  μs was the time interval between successive A-lines. The particle velocity volumetric frames were denoised using both temporal and spatial frequency filters. In the temporal domain, an impulse response filter centered around the excitation frequency of 3 kHz and full width half maximum bandwidth of 0.1 kHz, which was selected after some pre-analysis, was used to maintain signal integrity while minimizing noise. Similarly, a 2D spatial bandpass filter was applied to remove unwanted motion noise. The wavenumber filter was designed by setting the cutoff frequencies based on previously established ranges of shear wave speed in soft tissues, such as the mouse eye, which typically falls between 0.4 and 9.5  m/s.20,23 Then, the lower (kl) and upper (ku) limits of the wavenumber filter were determined using the relation [kl,ku]=[2πf/vu,2πf/vl], where f=3  kHz was the excitation frequency and vl=0.4  m/s and vu=9.5  m/s were the assumed speed limits. After denoising, the local shear wave speed, vs, was estimated using the relation vs=2πf/k, where k was the local wavenumber obtained by undertaking a 2D autocorrelation (window size = 0.4 mm by 0.4 mm) of the particle velocity volume followed by fitting the autocorrelation profiles to the analytical solutions of the reverberant shear wave field model.17,18,24,25 The window size was selected to be equal to at least half of the wavelength of the induced elastic wave to achieve a better estimate of the wave speed and not compromise the achievable elastic resolution. The described procedure was implemented on the particle velocity reverberant frames at all depths, encompassing all en-face frames of each 3D particle velocity volume. Wave speed results in different regions of the eye were compared using Student’s t-test.

3.

Results

3.1.

Structural OCT Images of the Eye Globe

The structural image of the whole eyeball was reconstructed from the M-C OCT scan. Figure 2 shows a typical structural image of mouse eyeball depicting both the anterior and posterior segments. The volume render in Fig. 2(a) clearly reveals the anterior segment with the cornea, iris, lens, and anterior region of the sclera. The major ocular components can be better revealed using a cross-sectional view indicated Fig. 2(b) (bottom), which was extracted from the yellow sectional plane indicated in Fig. 2(b) (top). Here, all the major ocular components from anterior (i.e., cornea, lens, iris, cornea, and anterior segment of sclera) to posterior (i.e., retina, choroid, and posterior scleral segment) are discernable. The ability of the OCT system to discern the ocular components is critical to correctly relate the elasticity differences measured by RevOCE.

3.2.

Wave Propagation and Speed Maps

The reverberant shear wave field induced in the eye globe can be visualized by taking the instantaneous particle velocity map in various en-face images positioned in parallel planes orthogonal to the anterior-posterior pole. Figure 3 depicts representative en-face OCT images (top row) taken from different en-face positions and instantaneous wave motion snapshots (bottom row) at t=2.12  ms after start of the excitation. Figure 3(a) shows the cornea region near the apex while Fig. 3(b) shows a partial view of the lens and iris in addition to the cornea. In Fig. 3(c), a partial view of the sclera as well as the cornea, lens and iris are visible. Finally, the last column, Fig. 3(d), shows a slice in the posterior segment near the retina. Evidently, the reverberant shear wave field was well-established throughout the whole eye globe (Fig. 3, bottom row).

Fig. 3

Elastic wave propagation snapshots at t=2.12  ms after excitation at different depths in en-face slices of the eyeball showing (a) cornea (near apex); (b) cornea, iris, and lens; (c) cornea, iris, lens, and sclera (partial view); and (d) posterior (across retina). Clearly, the reverberant shear wave field is generated in all regions of the eyeball from the apex of cornea to the posterior segment of the eye.

JBO_28_9_095001_f003.png

Figure 4 depicts a wave speed map in the anterior segment of the eyeball. This map allows us to identify the major components of the anterior eyeball, as seen in Figs. 4(a) and 4(b). The cross-sectional images shown in Figs. 4(c) and 4(d) reveal that there is a noticeable difference in the elastic wave speed among these components. Nevertheless, it is worth noting that the wave speeds at the air-tissue interfaces (e.g., cornea exterior surface) and the liquid-tissue interfaces (e.g., lens and iris anterior surface) seem to exhibit similar trends across all ocular components, with consistently lower values compared to the wave speed within their inner regions as shown in Fig. 4(d). This is mainly attributed to the correlation window in which tissue and non-tissue pixels are included in these regions. To avoid this issue, it is possible to segment the eyeball and only analyze the tissue regions using autocorrelation-based speed analysis. Nonetheless, one can still identify differences in the elastic wave speed by examining the cross-sectional view, i.e., x-z plane shown in Fig. 4(c), which is displayed in Fig. 4(d). The annotated regions display a significant variation in elastic wave propagation speed, with the lens and iris exhibiting a slower wave speed compared to the cornea and sclera.

Fig. 4

(a) A typical 3D speed map of anterior eye segment with slight cut plane in x-z plane to show the sclera. (b) An enface cut in the x-y plane revealing the lens, cornea and iris regions. (c) Partial planar cuts showing the depth and enface regions (sclera). (d) A cross-sectional slice taken from the x-z plane in panel (c) depicting elastic wave speed differences in the main components in the anterior segment.

JBO_28_9_095001_f004.png

3.3.

Comparison of Wave Speeds in Different Ocular Components

Figure 5 demonstrates the relative differences in wave speed measured in the anterior segment, including the cornea, lens, iris, and anterior sclera. Specific regions within each component were selected, and the means of their speed differences were plotted in a bar graph. As indicated in Fig. 5(b), the anterior sclera exhibits larger wave speed compared to the components in the anterior segment. However, it is essential to note that the corneal wave speed is not uniform, potentially indicating regional differences in corneal stiffness. The next section will detail a comparison of regional variations in corneal wave speed.

Fig. 5

Quantifying the elastic wave speeds in different ocular components. (a) Wave speed map in a representative anterior eye cross-section. Regions of interest for quantitative estimation of the wave speeds in each ocular component are indicated by the dashed rectangular regions; C, cornea; I, iris; L, lens; and S, anterior sclera. (b) Comparison of wave speeds of ocular components in the anterior eye. The mean wave speeds were quantified from the dashed rectangular regions in panel (a) excluding any fluid (N=3 eyes). (c) Partial view of a 3D wave speed map in the posterior segment of the eyeball. PS: posterior sclera. (d) Elastic wave speed profile along the anterior-posterior direction for the dashed rectangular region shown in panel (c) and annotated as ROI indicating increase in wave speed from the inner layer (i.e., retina) to the outer layer (i.e., posterior sclera).

JBO_28_9_095001_f005.png

Similarly, the wave speed map in the posterior segment shown in Fig. 5(c) indicates relative variations in the sclera, choroidal and retina regions. Figure 5(c) displays a partial view of a volumetric wave speed map of the posterior segment, showing variations across the thickness of the eyeball. By selecting differential rectangular regions indicated by the dashed rectangle annotated as ROI in Fig. 5(c), we quantified the speed variations along the posterior eyeball. Figure 5(d) presents the anterior-posterior profile of the wave speed in the dashed rectangular region, revealing a change in wave speed from approximately 2.4 to 3.2  m/s in the anterior (retina) to the posterior (sclera) regions, respectively. Moreover, the bumps in the profile between the extreme ends potentially indicate the presence of different layers (e.g., retina layers or the choroidal region). This information provides further insight into the differences in elasticity among the different layers of the posterior segment.

3.4.

Regional Variation of Wave Speed in the Cornea

The ability to quantify the biomechanical properties of the entire eye globe offers a significant advantage by allowing for the evaluation of the localized stiffness of specific ocular components such as the cornea. In Fig. 6, the variation in elastic wave speed of the cornea is illustrated. The depth-averaged (i.e., along the optical axis of the eye) structural image of the cornea is shown in Fig. 6(a), along with its corresponding wave speed map in Fig. 6(b). The wave speed map reveals that the highest elastic wave speed is observed at the corneal apex, with the speed decreasing progressively towards the pre-limbal region. A clearer representation of this regional variation can be obtained by referring to the two en-face speed maps displayed in Figs. 6(c) and 6(d), which were obtained from two parallel planes separated by 0.19 mm along the anterior-posterior pole. The speed profile from the apex to the periphery, an average of the profiles along the four paths represented by the arrows in Fig. 6(d), is shown in Fig. 6(e) and demonstrates a noticeable reduction from the apex to the pre-limbal region, decreasing from 3.50 to 2.35  m/s over a radial distance of 0.75 mm. Furthermore, the radial non-uniformity of the speed map in Fig. 6(a) is of interest since it potentially suggests that the stiffness of different corneal quadrants is heterogeneous. However, it is not possible to identify which regions are inferior-superior or temporal-nasal since the experiments did not note the anatomical alignment of the eyeball.

Fig. 6

Regional variations of elastic wave speed in the cornea. (a) En-face structural image of representative mouse cornea. (b) Depth averaged en-face elastic wave speed map in the cornea in panel (a). Representative elastic wave speed maps in the en-face of cornea taken at (c) 0.19 mm and (d) 0.36 mm posterior to corneal apex. (e) Radial wave speed profile along the four arrows shown in panel (b), i.e., from the apex to the periphery. SE: standard error of the four speed profiles taken along the directions shown by the arrows in panel (b).

JBO_28_9_095001_f006.png

4.

Discussion and Conclusion

The purpose of this study was to explore the potential of a novel RevOCE technique for quantifying the biomechanical properties of the whole eyeball. A multifocal ARF method was utilized to induce reverberant shear wave fields, which were then imaged using PhS-OCT. The results show that the technique can be used to quantify the biomechanical properties of the whole eyeball, providing insight into the relative differences in the stiffness of different ocular components of the intact whole eyeball. For the experimental conditions of the current study, the wave speed was found to be the greatest in the sclera, approaching 3.7  m/s in its anterior region. The mean elastic wave speed in the apical cornea region was found to be significantly higher when compared to the mean elastic wave speed in the apical region of the lens (p<0.005) and the pupillary zone of the iris (p=0.003) as indicated by a Student’s t-test. Likewise, the scleral elastic wave speed exhibited significantly higher value compared to that of the cornea, iris, and lens (p<0.001). Additionally, the iris showed a relatively higher wave speed compared to the lens (p=0.014). Although the posterior sclera had a relatively lower speed compared to the anterior sclera, this difference was not found to be statistically significant (p>0.05). If the IOP were maintained at a higher level (12  mmHg), it is possible that the contrast in wave speed between the cornea, lens, and iris would have been even more pronounced.11,14,26,27 The elastic wave speeds in the various components of the eye fall within a reasonable range as reported in previous studies.11,15,26,28 However, it is worth noting that these values may significantly differ depending on factors such as the IOP, boundary conditions, species, and age of the animal model.15,2830

The technique also allowed for the evaluation of localized elastic wave speed within specific ocular components. For instance, the cornea showed regional variation in elastic wave speed, with higher elastic wave speed observed in the central regions and lower speed towards the peripheral regions. This finding is consistent with prior studies,9,31 which has shown increased corneal compliance towards the limbal junction and minimal changes in central corneal curvature in response to IOP variations, suggesting that the central cornea is stiffer compared to the peripheral cornea. The observed variation in deformation across different regions is thought to be influenced by structural factors, including collagen distribution and orientation, as well as the distribution and concentration of elastin fibers in the cornea.9 These findings hold significant implications for understanding corneal biomechanics, the pathogenesis of corneal diseases, and the potential for tailored therapeutic approaches.32,33 Moreover, the apparent radial non-uniformity in the elastic wave speed map of the cornea, as shown in Fig. 6(b), particularly at the peripheral regions (i.e., the pre-limbal region), may suggest the existence of potential stiffness variation in the different corneal quadrants. For example, a previous study reported meridional variability in corneal deformation, showing greater deformation in the peripheral cornea at the superior and inferior poles compared to the nasal-temporal region.9 This information is essential to establish the correlation between regional biomechanical properties and the underlying collagen fibril structure in each region. Hence, this technique holds great promise in accurately diagnosing corneal diseases that exhibit regional biomechanical heterogeneity, such as keratoconus.34,35 However, it is crucial to note that further studies with a larger sample size and different species (e.g., human) are necessary to confirm this correlation and associate biomechanical heterogeneity with the quadrants of the cornea. Such studies can aid in the development of effective treatment strategies for corneal disorders.36,37

The elastic wave speed profile in the posterior segment in proximity to the optic nerve revealed a gradual increase from the inferior (2.4  m/s) to posterior (3.4  m/s) direction, with discernible bumps potentially indicating the presence of biomechanically distinct layers, such as the layers of the retina, the choroidal membrane, and the posterior sclera. Due to the relatively smaller thickness of the layers in the mouse ocular components (tens of micrometers) in comparison to the elastic resolution, it was challenging to visualize each layer of the retina and choroidal membrane accurately.38 Therefore, conducting further research on different species with eyeball sizes comparable to humans and with improved elastic resolution would offer a better understanding of the capabilities of RevOCE in identifying various layers of the posterior segment of the eyeball. However, previous work has shown that RevOCE has improved mechanical contrast as compared to traditional wave-based OCE,18 so further studies with higher excitation frequencies are underway.

While this study successfully determined the elastic wave speed of various ocular tissues using OCE, there are some limitations that could be addressed in future work. First, the current OCT system has a limited axial imaging range, which necessitates a two-step measurement process. To overcome this limitation, future work may utilize longer-range OCT systems like swept source OCT systems,39 enabling faster imaging without the need for adjustment of the axial position of the imaging probe. Second, the excitation orientation and the OCT objective lens were restricted to opposite sides of the eyeball, making it unsuitable for in vivo conditions. Future work will consider allowing excitation orientation on the same side of the objective lens to enable in vivo assessment, such as with an imaging hole as demonstrated with ring-shaped transducers.40,41 Third, the use of different excitation frequencies could provide valuable viscoelastic properties of the various ocular tissues.

Generally, the RevOCE technique employed in this study offers an unprecedented opportunity to evaluate the biomechanical properties of various ocular components simultaneously. This technique stands out for its superior elasticity resolution when compared to the USE previously reported11 and superior mechanical contrast compared to traditional wave-based OCE.18 Furthermore, unlike conventional RevOCE methods that utilize mechanical shakers that physically contact tissues directly, which can potentially compromise the integrity of delicate tissues while inducing a reverberant field,18,42,43 the presented technique relies on ARF, which does not necessitate direct contact between the actuator and the sample, making it more convenient for in vivo applications. By providing a more complete understanding of the interrelationship of the biomechanical properties of the different ocular components, this technique can have important implications for both research and clinical practice. For instance, it can help in analyzing the effect of a particular disease on different ocular components, which can lead to more effective diagnosis and treatment strategies that may have been unavailable due to focusing on a specific ocular component only. Furthermore, the ability to concurrently evaluate multiple ocular components is especially useful in monitoring disease progression, where the effect reflected on secondary ocular component due to a diseased ocular component would carry significant value, such as improved diagnostic efficacy. It can also be used to assess the efficacy and side effects of treatment strategies, such as corneal cross-linking for keratoconus,44,45 by monitoring changes in the elastic properties of the cornea as well as other adjacent ocular components, respectively.

Disclosures

MS and KVL have financial interests in ElastEye LLC., which is not directly related to this work.

Code, Data, and Materials Availability

All data in support of the findings of this paper are available from the authors upon reasonable request.

Acknowledgments

This work was funded in parts by the US National Institutes of Health (NIH, Grant Nos. R01EY022362, R01EY030063, and R01EY033978).

References

1. 

A. P. Voorhees et al., “Whole-globe biomechanics using high-field MRI,” Exp. Eye. Res., 160 85 –95 https://doi.org/10.1016/j.exer.2017.05.004 EXERA6 0014-4835 (2017). Google Scholar

2. 

A. B. Cankaya et al., “Ocular response analyzer to assess corneal biomechanical properties in exfoliation syndrome and exfoliative glaucoma,” Graefes Arch. Clin. Exp. Ophthalmol., 250 (2), 255 –260 https://doi.org/10.1007/s00417-011-1793-7 (2012). Google Scholar

3. 

L. Zimprich et al., “Corneal hysteresis as a biomarker of glaucoma: current insights,” Clin. Ophthalmol., 14 2255 –2264 https://doi.org/10.2147/OPTH.S236114 (2020). Google Scholar

4. 

R. Vinciguerra et al., “Detection of keratoconus with a new biomechanical index,” J. Refract. Surg., 32 (12), 803 –810 https://doi.org/10.3928/1081597X-20160629-01 JRSUEY 0883-0444 (2016). Google Scholar

5. 

S. H. Yun and D. Chernyak, “Brillouin microscopy: assessing ocular tissue biomechanics,” Curr. Opin. Ophthalmol., 29 (4), 299 –305 https://doi.org/10.1097/ICU.0000000000000489 COOTEF (2018). Google Scholar

6. 

G. Scarcelli et al., “Brillouin microscopy of collagen crosslinking: noncontact depth-dependent analysis of corneal elastic modulus,” Invest. Ophthalmol. Vis. Sci., 54 (2), 1418 –1425 https://doi.org/10.1167/iovs.12-11387 IOVSDA 0146-0404 (2013). Google Scholar

7. 

M. A. Kirby et al., “Optical coherence elastography in ophthalmology,” J. Biomed. Opt., 22 (12), 121720 https://doi.org/10.1117/1.JBO.22.12.121720 JBOPFO 1083-3668 (2017). Google Scholar

8. 

I. Issarti, C. Koppen and J. J. Rozema, “Influence of the eye globe design on biomechanical analysis,” Comput. Biol. Med., 135 104612 https://doi.org/10.1016/j.compbiomed.2021.104612 CBMDAW 0010-4825 (2021). Google Scholar

9. 

C. Whitford et al., “Ex vivo testing of intact eye globes under inflation conditions to determine regional variation of mechanical stiffness,” Eye Vision, 3 (1), 21 https://doi.org/10.1186/s40662-016-0052-8 (2016). Google Scholar

10. 

X. Qian et al., “Ultrasonic elastography to assess biomechanical properties of the optic nerve head and peripapillary sclera of the eye,” Ultrasonics, 110 106263 https://doi.org/10.1016/j.ultras.2020.106263 ULTRA3 0041-624X (2021). Google Scholar

11. 

R. Li et al., “Simultaneous assessment of the whole eye biomechanics using ultrasonic elastography,” IEEE Trans. Biomed. Eng., 70 (4), 1310 –1317 https://doi.org/10.1109/TBME.2022.3215498 IEBEAX 0018-9294 (2023). Google Scholar

12. 

J. Schmitt, “Oct elastography: imaging microscopic deformation and strain of tissue,” Opt. Express, 3 (6), 199 –211 https://doi.org/10.1364/OE.3.000199 OPEXFF 1094-4087 (1998). Google Scholar

13. 

K. V. Larin and D. D. Sampson, “Optical coherence elastography - OCT at work in tissue biomechanics [invited],” Biomed. Opt. Express, 8 (2), 1172 –1202 https://doi.org/10.1364/BOE.8.001172 BOEICL 2156-7085 (2017). Google Scholar

14. 

F. Zvietcovich et al., “Dynamic optical coherence elastography of the anterior eye: understanding the biomechanics of the limbus,” Invest. Ophthalmol. Vis. Sci., 61 (13), 7 https://doi.org/10.1167/iovs.61.13.7 IOVSDA 0146-0404 (2020). Google Scholar

15. 

Y. Li et al., “Simultaneously imaging and quantifying in vivo mechanical properties of crystalline lens and cornea using optical coherence elastography with acoustic radiation force excitation,” APL Photonics, 4 (10), 106104 https://doi.org/10.1063/1.5118258 (2019). Google Scholar

16. 

J. Zhang et al., “Elasticity measurements of ocular anterior and posterior segments using optical coherence elastography,” Opt. Express, 30 (9), 14311 –14318 https://doi.org/10.1364/OE.456065 OPEXFF 1094-4087 (2022). Google Scholar

17. 

K. J. Parker et al., “Reverberant shear wave fields and estimation of tissue properties,” Phys. Med. Biol., 62 (3), 1046 –1061 https://doi.org/10.1088/1361-6560/aa5201 PHMBA7 0031-9155 (2017). Google Scholar

18. 

F. Zvietcovich et al., “Reverberant 3D optical coherence elastography maps the elasticity of individual corneal layers,” Nat. Commun., 10 (1), 4895 https://doi.org/10.1038/s41467-019-12803-4 NCAOBW 2041-1723 (2019). Google Scholar

19. 

T. Mekonnen et al., “A multifocal acoustic lens-transducer system for reverberant optical coherence elastography,” Proc. SPIE, 12381 123810B https://doi.org/10.1117/12.2650633 PSISDG 0277-786X (2023). Google Scholar

20. 

T. Mekonnen et al., “Longitudinal assessment of the effect of alkali burns on corneal biomechanical properties using optical coherence elastography,” J. Biophotonics, 15 (8), e202200022 https://doi.org/10.1002/jbio.202200022 (2022). Google Scholar

21. 

T. Mekonnen et al., “Reverberant optical coherence elastography using multifocal acoustic radiation force,” Opt. Lett., 48 (11), 2773 –2776 https://doi.org/10.1364/OL.482201 OPLEDP 0146-9592 (2023). Google Scholar

22. 

G. Flores et al., “Experimental study to evaluate the generation of reverberant shear wave fields (R-SWF) in homogenous media,” in IEEE Int. Ultrasonics Symp. (IUS), 1 –4 (2020). https://doi.org/10.1109/IUS46767.2020.9251338 Google Scholar

23. 

J. Li et al., “Dynamic optical coherence tomography measurements of elastic wave propagation in tissue-mimicking phantoms and mouse cornea in vivo,” J. Biomed. Opt., 18 (12), 121503 https://doi.org/10.1117/1.JBO.18.12.121503 JBOPFO 1083-3668 (2013). Google Scholar

24. 

J. Ormachea, B. Castaneda and K. J. Parker, “Shear wave speed estimation using reverberant shear wave fields: implementation and feasibility studies,” Ultrasound Med. Biol., 44 (5), 963 –977 https://doi.org/10.1016/j.ultrasmedbio.2018.01.011 USMBA3 0301-5629 (2018). Google Scholar

25. 

J. Ormachea, K. J. Parker and R. G. Barr, “An initial study of complete 2D shear wave dispersion images using a reverberant shear wave field,” Phys. Med. Biol., 64 (14), 145009 https://doi.org/10.1088/1361-6560/ab2778 PHMBA7 0031-9155 (2019). Google Scholar

26. 

C. Zevallos-Delgado et al., “Assessing porcine iris elasticity and mechanical anisotropy with optical coherence elastography,” J. Biomed. Photonics Eng., 7 (4), 40304 https://doi.org/10.18287/JBPE21.07.040304 (2021). Google Scholar

27. 

S. Park et al., “The impact of intraocular pressure on elastic wave velocity estimates in the crystalline lens,” Phys. Med. Biol., 62 (3), N45 –N57 https://doi.org/10.1088/1361-6560/aa54ef PHMBA7 0031-9155 (2017). Google Scholar

28. 

C. Wu et al., “Measuring the elastic wave velocity in the lens of the eye as a function of intraocular pressure using optical coherent elastography,” Quantum Electron., 49 (1), 20 –24 https://doi.org/10.1070/QEL16900 QUELEZ 1063-7818 (2019). Google Scholar

29. 

S. Ye et al., “In vivo non-contact measurement of human iris elasticity by optical coherence elastography,” J. Biophotonics, 14 (9), e202100116 https://doi.org/10.1002/jbio.202100116 (2021). Google Scholar

30. 

Z. Han et al., “Optical coherence elastography assessment of corneal viscoelasticity with a modified Rayleigh-Lamb wave model,” J. Mech. Behav. Biomed. Mater., 66 87 –94 https://doi.org/10.1016/j.jmbbm.2016.11.004 (2017). Google Scholar

31. 

A. Wilson et al., “An interferometric ex vivo study of corneal biomechanics under physiologically representative loading, highlighting the role of the limbus in pressure compensation,” Eye Vision, 7 (1), 43 https://doi.org/10.1186/s40662-020-00207-1 (2020). Google Scholar

32. 

Y. Zhao et al., “Biomechanical analysis of ocular diseases and its in vitro study methods,” Biomed. Eng. Online, 21 (1), 49 https://doi.org/10.1186/s12938-022-01019-1 (2022). Google Scholar

33. 

C. J. Roberts and Jr. W. J. Dupps, “Biomechanics of corneal ectasia and biomechanical treatments,” J. Cataract. Refract. Surg., 40 (6), 991 –998 https://doi.org/10.1016/j.jcrs.2014.04.013 JCSUEV 0886-3350 (2014). Google Scholar

34. 

V. Mas Tur et al., “A review of keratoconus: diagnosis, pathophysiology, and genetics,” Surv. Ophthalmol., 62 (6), 770 –783 https://doi.org/10.1016/j.survophthal.2017.06.009 SUOPAD 0039-6257 (2017). Google Scholar

35. 

P. Padmanabhan and A. Elsheikh, “Keratoconus: a biomechanical perspective,” Curr. Eye Res., 48 1 –9 CEYRDM 0271-3683 (2022). Google Scholar

36. 

A. Elsheikh et al., “Experimental assessment of corneal anisotropy,” J. Refract. Surg., 24 (2), 178 –187 https://doi.org/10.3928/1081597X-20080201-09 JRSUEY 0883-0444 (2008). Google Scholar

37. 

M. Singh et al., “Assessing the effects of riboflavin/UV-A crosslinking on porcine corneal mechanical anisotropy with optical coherence elastography,” Biomed. Opt. Express, 8 (1), 349 –366 https://doi.org/10.1364/BOE.8.000349 BOEICL 2156-7085 (2017). Google Scholar

38. 

L. R. Ferguson et al., “Retinal thickness normative data in wild-type mice using customized miniature SD-OCT,” PLoS One, 8 (6), e67265 https://doi.org/10.1371/journal.pone.0067265 POLNCL 1932-6203 (2013). Google Scholar

39. 

J. Fujimoto and E. Swanson, “The development, commercialization, and impact of optical coherence tomography,” Invest. Ophthalmol. Vis. Sci., 57 (9), OCT1 –OCT13 https://doi.org/10.1167/iovs.16-19963 IOVSDA 0146-0404 (2016). Google Scholar

40. 

F. Zvietcovich et al., “Confocal air-coupled ultrasonic optical coherence elastography probe for quantitative biomechanics,” Opt. Lett., 45 (23), 6567 –6570 https://doi.org/10.1364/OL.410593 OPLEDP 0146-9592 (2020). Google Scholar

41. 

W. Qi et al., “Confocal acoustic radiation force optical coherence elastography using a ring ultrasonic transducer,” Appl. Phys. Lett., 104 (12), 123702 https://doi.org/10.1063/1.4869562 APPLAB 0003-6951 (2014). Google Scholar

42. 

L. A. Aleman-Castaneda, F. Zvietcovich and K. J. Parker, “Reverberant elastography for the elastic characterization of anisotropic tissues,” IEEE J. Sel. Top. Quantum Electron., 27 (4), 1 –12 https://doi.org/10.1109/JSTQE.2021.3069098 IJSQEN 1077-260X (2021). Google Scholar

43. 

G. R. Ge et al., “Assessing corneal cross-linking with reverberant 3D optical coherence elastography,” J. Biomed. Opt., 27 (2), 026003 https://doi.org/10.1117/1.JBO.27.2.026003 JBOPFO 1083-3668 (2022). Google Scholar

44. 

B. J. Blackburn, A. M. Rollins and Jr. W. J. Dupps, “Biomechanics of ophthalmic crosslinking,” Transl. Vision Sci. Technol., 10 (5), 8 https://doi.org/10.1167/tvst.10.5.8 (2021). Google Scholar

45. 

G. Wollensak, E. Spoerl and T. Seiler, “Riboflavin/ultraviolet-a-induced collagen crosslinking for the treatment of keratoconus,” Am. J. Ophthalmol., 135 (5), 620 –627 https://doi.org/10.1016/S0002-9394(02)02220-1 AJOPAA 0002-9394 (2003). Google Scholar

Biography

Taye Mekonnen is a postdoctoral fellow at the University of Houston. He received his PhD in biomedical engineering from Macquarie University in 2020. Furthermore, he completed his MSc degree in computer engineering at Addis Ababa University and a BSc degree in electrical engineering at Bahir Dar University, both in Ethiopia. His current research interest focuses on advancing and implementing high-resolution optical coherence tomography and optical coherence elastography imaging techniques in the fields of biomedical and pharmaceutical applications.

Christian Zevallos-Delgado is a biomedical engineering PhD student at the University of Houston. He received his Master of Science in bioinformatics and molecular biochemistry from George Washington University in 2020 and Bachelor of Engineering in biotechnology from Catholic University of Santa María, Arequipa, Peru in 2016. His research interest involves the assessment of biomechanical properties in developmental biology using mouse and zebrafish utilizing optical coherence elastography.

Manmohan Singh is currently a research assistant professor in Dr. Kirill Larin’s Biomedical Optics Laboratory at the University of Houston. He received his PhD in biomedical engineering in 2018 from the University of Houston and completed a fellowship from the National Library of Medicine in Data Science and Biomedical Informatics in 2020. His research interests are focused on developing optical imaging modalities and image processing techniques to detect diseases noninvasively.

Salavat R. Aglyamov is a research assistant professor of mechanical engineering at the University of Houston. He received the BS and MS degrees in applied mathematics in 1991 and 1993, respectively, from Moscow State University, Russia. He received a PhD in biophysics (1999) from the Institute of Theoretical and Experimental Biophysics, Russia. His research interests are in the areas of tissue biomechanics, elasticity imaging, medical imaging, ultrasound, and applied mathematics.

Kirill V. Larin is a Cullen College of Engineering Distinguished Professor of Biomedical Engineering at the University of Houston. He received his MS degree in laser physics and mathematics from Saratov State University (1995) and PhD in biomedical engineering from the University of Texas Medical Branch in Galveston (2002). He has published more than 200 papers in the field of biomedical optics and biophotonics. He is a fellow of SPIE, OPTICA, and AIMBE.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 International License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Taye T. Mekonnen, Christian Zevallos-Delgado, Manmohan Singh, Salavat R. Aglyamov, and Kirill V. Larin "Multifocal acoustic radiation force-based reverberant optical coherence elastography for evaluation of ocular globe biomechanical properties," Journal of Biomedical Optics 28(9), 095001 (11 September 2023). https://doi.org/10.1117/1.JBO.28.9.095001
Received: 24 April 2023; Accepted: 28 August 2023; Published: 11 September 2023
Advertisement
Advertisement
KEYWORDS
Cornea

Biomechanics

Eye

Elasticity

Acoustic waves

Sclera

Acoustics

Back to Top