1.IntroductionHistorically, the limit to the achievable spatial resolution of an optical microscope was defined by Abbe’s criteria based on principles of diffraction of light. With the development of high numerical aperture and low aberration optics, diffraction limited imaging modalities, including optical sectioning confocal fluorescence microscopy, have been used to investigate many interesting research questions in the biomedical field. However, many open questions in biology cannot be addressed without resolving finer details closer to the size of proteins and other biomolecules, typically tens of nanometers. Super-resolution (SR) microscopy techniques, such as stimulated emission depletion (STED),1,2 MINFLUX3,4 structured illumination microscopy (SIM),5,6 and single molecule localization techniques, such as PALM/STORM7,8 have moved fluorescence microscopy beyond the Abbe diffraction limit, allowing spatial resolutions of 10’s of nanometers. These SR microscopy techniques have opened up new possibilities for research using optical microscopy that were closed previously, examples of which include: observing the dynamics of dendritic spines in the brains of live zerbrafish larvae and mice,9 imaging the periodicity of the actin cytoskeleton using STORM in fixed neurons10 and STED in live neurons,11 and imaging the nanoscale distribution of Tom20 in mitochondria as a function of cell type, cell location, and colony location.12 STED microscopy, first demonstrated by the lab of Stefan Hell,2 has been utilized for live cell and in vivo SR imaging due to its fast times and ability to directly retrieve a super-resolved image without any required post-processing, as compared to PALM/STORM. STED microscopy can provide enhanced spatial resolution in comparison to SR-SIM. STED microscopy builds on conventional fluorescence microscopy by adding a second “STED” laser, tuned to the red tail of the fluorophore’s emission spectrum, co-aligned with the traditional excitation laser. The phase of the input STED beam is modified using a vortex waveplate to generate a donut spatial pattern at the focus of the objective, with a central null overlapped with the excitation focal spot. The co-aligned lasers are raster-scanned on the sample and the fluorescence is collected through a confocal pinhole to a photodetector. When a fluorophore is overlapped with the intense STED laser, outside of the central null, the stimulated emission rate is much higher than spontaneous emission, preventing fluorescence. The result is that only those molecules at the null in the STED donut beam can fluoresce, and thus be detected. With high STED laser intensities, resolution down to tens of nanometers is achievable in labeled biological samples.13,14 The resolution of a STED microscope is given as15 where NA is the numerical aperture of the objective, is the wavelength, is the maximum intensity of the STED beam, and is the saturation intensity of the fluorophore. To achieve as much as 10-fold improvement in resolution, the STED laser needs to be approximately three orders of magnitude higher power than the excitation laser ( of excitation, mW of STED power). Unfortunately, high STED intensity can cause photobleaching, and this imposes a limit to the intensity and therefore the attainable resolution.16 For dynamic imaging, photobleaching also limits the amount of time that live biological processes can be monitored with STED. In conventional STED, the high-power STED laser is scanned continuously over the sample regardless of whether there are fluorophores at a given pixel in the image, leading to excess sample photobleaching.In recent years, several methods have been developed to reduce photobleaching during STED imaging, including RescueSTED17 and dynamic intensity minimum (DyMIN).18 DyMIN, first demonstrated in 2017 by S. Hell, is an improvement on RescueSTED. DyMIN dynamically controls the STED laser intensity pixel by pixel to minimize the power on the sample. If there is no fluorescence detected from a pixel, the STED laser power remains off and the beam is moved to the next pixel. If fluorescence is detected above a threshold set for each DyMIN step, the DyMIN algorithm will incrementally increase the STED power. Ideally, the maximum STED power is only used when a fluorophore is present at the STED null, lower power is used when the fluorophore is located off of the null, and no STED power is used when no fluorophore is present. The DyMIN process can greatly reduce the overall exposure of the sample to the STED laser, especially for sparsely labeled samples, and results in decreased photobleaching. DyMIN can therefore allow acquisition of a time series over a longer period before the sample is photobleached, above a threshold set for each DyMIN step. The implementation of DyMIN on a STED microscope can involve substantial effort, including programming an field-programmable gate array (FPGA) to implement the DyMIN algorithm and combining it with custom electronics to allow fast switching of the STED optical power by modulating the radio frequency (RF) power controlling an acousto-optic modulator (AOM). Here, we discuss our implementation using inexpensive components and provide schematics and source code to make DyMIN simpler to use for labs employing STED microscopy. Previous biological applications of the DyMIN method include imaging synaptic vesicles19 on fixed samples, in addition to a related adaptive illumination method that used an FPGA to rapidly turn off the STED laser when scanning over highly reflective parts of malaria cells, which are sensitive to light.20 However, the DyMIN code and detailed methods used in these papers were not open-source. An open-source DyMIN implementation will bring this powerful method to more researchers using STED and assist with other adaptive illumination schemes. In the following sections, we present our custom DyMIN system and demonstrate its application to timelapse STED imaging, quantifying a reduction in photobleaching of beads. We also apply DyMIN to imaging of live cells and neural tissue samples. 2.Background: three-Step DyMIN AlgorithmThe DyMIN algorithm performs up to three steps of sequentially increasing STED laser power based on the number of collected fluorescent photons at each step, for each pixel in the image. Figure 1 illustrates the DyMIN algorithm process. In practice, the user selects a STED power value (), fluorescence collection time (), and a counts threshold (), for each DyMIN step (). The excitation power () is held constant during imaging for our implementation of DyMIN. At the th DyMIN step, the STED power and excitation power are incident on the sample, centered on a particular pixel, whereas the fluorescence photons are collected for a time . The total number of photons collected is denoted . Typically, the first DyMIN step is performed without any STED enhancement, i.e. , and successive DyMIN steps have increasing STED power, i.e., . If at DyMIN step , the photon counts fail to reach the predefined threshold (), the algorithm is aborted, and the number of photon counts is set to zero for that pixel. Subsequently, the STED laser returns to zero power and the algorithm waits for the next pixel clock and then starts over at step . If then the DyMIN algorithm stays on the pixel but moves on to the next DyMIN step, , by increasing the STED power to , collecting and counting the fluorescence photons and once again comparing the number of collected photons to the threshold value . If the final DyMIN step () is reached, the highest STED laser power () illuminates the sample and the number of photons detected during the final step is recorded at that pixel. Figure 1 outlines the three-step DyMIN algorithm for different scan positions approaching the location of a fluorophore. At the first location in Fig. 1(a) the fluorophore is outside the PSF of the excitation beam. The DyMIN algorithm starts at step 1, illuminates the sample with zero STED power, i.e., with only the confocal excitation, collects and thus stops after step 1, recording zero counts at that pixel location in the image. The DyMIN algorithm scans to next pixel, starting the DyMIN steps over with step 1 at location . In Fig. 1(b), the scan is now centered at position , and the fluorophore is within the confocal PSF, thus the number of photons collected exceeds the threshold and the DyMIN algorithm moves on to Step 2, changing the STED power to . The algorithm collects photon counts for a duration , but the effective PSF is smaller and there is no longer any overlap with the fluorophore position, thus the threshold is not reached (), the DyMIN steps are aborted, and the number of photons counted is recorded as zero at that pixel. The STED laser is reset to zero power and the DyMIN algorithm proceeds to the next pixel to start over at step 1. In Fig. 1(c) the center of the beam fully overlaps with the fluorophore at position , thus the counts are above threshold for steps 1 and 2, and step 3 is reached, the final DyMIN step. The number of photons collected at step 3, at the maximum STED power , will be counted over a collection time of and recorded. 3.Methods3.1.DyMIN STED ImplementationFigure 2(a) shows a schematic for our implementation of the DyMIN on a STED microscope. The custom STED microscope is described in further detail in Sec. 3.1. An AOM controls the STED laser power (Newport-EOS part # N24080). An RF driver with an analog amplitude modulation input (Newport-EOS part # N21080-0.9AMVCO) controls the AOM, which allows fast switching of the STED laser power in . The fluorescence signal is detected with an avalanche photodiode (APD, Excelitas SPCM-AQRH-15-FC) and the individual detected photons are converted to TTL pulses that are sent to the digital input on a field-programmable gate array (FPGA, NI myRIO-1900) programmed with the DyMIN algorithm using Labview FPGA (NI LabVIEW 2016, 32 bit). The FPGA counts the photons for each DyMIN step and then determines whether to go to the next step and raise the STED power or abort the DyMIN steps and move to the next pixel position. The analog output on the FPGA has a maximum slew rate of , too slow to directly control the AOM to change the output STED laser power on a time frame less than the dwell time used during imaging. Instead, we rely on the digital outputs from the FPGA to control a digital multiplexer circuit to rapidly modulate the control voltage to the AOM. A basic schematic of the multiplexer is shown in Fig. 2(b). Two digital outputs from the FPGA [labeled “Digital” in Fig. 2(a), DI0, DI1 in Fig. 2(b)] control the multiplexer (Analog Devices Inc, ADV3221 on ADV321-EVALZ) to switch between three different voltage values supplied by the analog outputs from the FPGA [labeled “Analog” in Fig. 2(a), AI0-AI2 in Fig. 2(b)]. Depending on the truth table in Fig. 2(b), the selected analog output [OUT in Figs. 2(a) and 2(b)] is sent from the multiplexer to the AOM RF driver, which allows faster control of the STED laser power than would be achievable using the analog output of the FPGA directly. The analog output values from the FPGA are chosen in software to select the STED power for the different DyMIN steps, , determined by calibration. In our implementation, we use Slidebook (Intelligent Imaging Innovations, Inc.) software for image acquisition and to control the scanning mirrors, however our DyMIN circuit and FPGA can be implemented on any image acquisition software that has a pixel clock and photon counting input. On our system, the pixel clock is supplied by a data acquisition (DAQ) board (NI PCIe-6259) to the FPGA digital input channel to synchronize the DyMIN steps with the image acquisition. The modified/unmodified (DyMIN/conventional STED) photon counts at each pixel are sent by a digital TTL output from the FPGA to the counter input on the DAQ to record photon counts in the software. For DyMIN, only the photon counts for the final DyMIN step are sent to the DAQ. Finally, we use a 10 MHz output clock signal from the FPGA to synchronize the STED laser clock so that the number of laser pulses per step stays constant over the measurement. The FPGA code and further details on the electronics are publicly available for download.21 3.2.Details of the DyMIN STED MicroscopeOur custom-built beam-scanning STED microscope for imaging yellow fluorescent protein (YFP) and green fluorescent protein (GFP) in live cells is shown schematically in Fig. 3. The excitation laser is a 485 nm pulsed diode laser (Picoquant, LDH-P-C-485-B controlled by PDL 800-B) and a 585 nm or 600 nm selectable pulsed output of a Rainbow laser (Mobius Photonics) serves as the STED laser. Both lasers are triggered by the same external pulse generator. The pulse widths of the STED and excitation laser are 800 ps and less than 140 ps, respectively. Both excitation and STED lasers pass through bandpass filters to clean up the spectra (Chroma, ET470/30X and ET600/40M) and a half-waveplate and a polarizer are used for power control. Each laser is then coupled into a PM fiber to clean up the spatial profiles. A vortex phase mask (RPC Photonics, VPP-1a) applies a spiral phase pattern to the STED laser to generate a donut shaped intensity pattern at the focal plane of the microscope. One dichroic beamsplitter (Chroma, ZT488RDC) combines the excitation and STED lasers while another (Chroma, ZT568RDC-XR) separates the fluorescence emission. A quarter waveplate (B. Halle, RAC 3.4.15), carefully aligned to obtain nearly perfect circular polarization of the STED beam (see Ref. 21 for details), is placed in the combined beam path of the excitation and STED beam before the galvo-scanner (Till Photonics, Yanus IV). The galvo-scanner is designed to image the first mirror onto the second mirror so that the STED beam remains stationary at the back focal plane of the objective during scanning. A scan lens (Thorlabs, LSM03-VIS) and tube lens (internal to Olympus IX-71) image the second galvo mirror onto the back focal plane of the 100X/1.4NA oil immersion objective lens (Olympus, uPlanSApo) that focuses the beams on the sample. The objective is mounted on the turret of an Olympus IX-71 microscope platform, which has been modified to incorporate a 3 mm thickness protected silver mirror that sends light that comes into the left side port straight up through the center of the tube lens and objective lens. A Nano-ZL XY microstage (Mad City Labs) controls the position of the sample and widefield epifluorescence, brightfield, or DIC imaging through the oculars is used for alignment. A Nano-Z100 piezo nanopositioning system (Mad City Labs) provides axial translation to perform z-stacks or x-z and y-z scans required for the STED alignment process, as detailed in Ref. 22, with insert adapters for different sample holders. The fluorescence collected by the objective is de-scanned and reflected by the dichroic beamsplitter to a lens that focuses the fluorescence into a multimode fiber that serves as a confocal pinhole (Thorlabs, AC254-200A and M43L01). The fiber relays the fluorescence to the APD (Excelitas, SPCM-AQRH-15-FC), which sends output TTL pulses to the FPGA running the DyMIN code and subsequently to the DAQ on the computer for photon counting, as described above. For DyMIN operation, the FPGA and multiplexer send an analog voltage signal to an AOM in the STED laser path to modulate the intensity. After the AOM, the STED laser is focused into a PM fiber. In characterizing the power control of the DyMIN and AOM, we found that the power level of the STED beam measured after the PM fiber has a time-dependent response after the AOM was switched on (see Fig. S1 in the Supplementary Material). The power stays constant over , then a gradual drop is seen, followed by an increase to eventually reach a constant average power by . We believe that this slower power modulation is due to heating that influences the coupling efficiency into the fiber, likely due to the laser mode not being perfectly . Therefore, less effective power is used for DyMIN compared with conventional STED if the same analog voltages are sent from the FPGA to the multiplexer and RF driver. To make comparisons between STED and DyMIN imaging, we carefully measured the steady state power (as would be used for STED when the laser is constantly on) and the power in the first 10 ms after the AOM is switched on. To compare photobleaching rates with our bead samples, we made sure to use a lower power for the STED beam when doing conventional STED so that we had the same effective power on the sample during the last DyMIN step. 3.3.Selecting the DyMIN ParametersDyMIN has clear benefits for dynamic time-series STED imaging due to decreased photobleaching. To maximize resolution and minimize exposure to STED light during scanning, the DyMIN three-step parameters , , , , and need to be optimized. For the remainder of this paper, we will write these parameters as threshold photon counts , fluorescence collection times where is the time used for the final DyMIN step and also the effective dwell time for the recorded image, and power levels where and is the final STED power. Below we outline how to set the intermediate parameters to achieve the desired final resolution while minimizing photobleaching during the DyMIN steps. The excitation power and STED power for the final DyMIN step are the same as what is used in conventional STED microscopy, with a pixel size of at least half the resolution required to achieve Nyquist frequency sampling. 3.3.1.Power parametersTo determine the DyMIN step power levels, first the resolution versus STED power was calibrated by imaging 23 nm fluorescent beads at a series of STED laser powers. Individual beads were identified from raster-scanned images, and a 2D Lorentzian function was then fit to this data using the method of least squares. The STED resolution was then determined by taking the mean FWHM values for all beads in the image. The values were then fit to a curve using Eq. (1) and plotted in Fig. 3(b). To determine the power levels for three-step DyMIN imaging of fluorescent beads, we first pick the resolution and subsequently the STED power to use for imaging, and this is the value for . To determine the value for , ideally one would like to minimize the STED light dose on the sample. In the original DyMIN manuscript (Ref. 18), the authors experimentally found the number of frames they could acquire before the signal dropped by 75% in a bead sample as a function of . The value of that maximized the number of frames was then used. Note that the value of also depends on the final STED power used. Based on our final STED power of 18.7 mW, we tested two values (3 mW, 8 mW); however, we did not see a significant difference in photobleaching rate [see Figs. 5(a) and 5(b)]. 3.3.2.Collection timesFor the step 1 and step 2 collection times ( and ), we used Eq. (S1) in Ref. 18. The dwell time for the final step () is a tradeoff between photobleaching and signal to noise ratio (SNR). If the goal is to take a long image sequence, you want to minimize the dwell time to avoid photobleaching, however this will also reduce SNR. For the bead samples, we used a final dwell time of and for the live cell and tissue we used 10 and , respectively. 3.3.3.Threshold parametersTo aid in the selection of the count thresholds for steps 1 and 2, we took a confocal image and a STED image at with the calculated collection times. We calculated of the peak fluorescence value of those images for use as and , respectively, as suggested in Ref. 17. However, we found that for time-series imaging we needed smaller threshold values to reach threshold on the later images in a series, after the detected signal has dropped with photobleaching. This step would not be required for taking a single DyMIN image. An improvement would be to allow the threshold to decrease dynamically over time during a time series based on the fluorescence in the most recent image, but we did not explore that option for this paper. 3.4.Sample PreparationFor the DyMIN bead images, 45 nm Yellow-Green FluoSpheres™ (ThermoFisher Scientific, F8795) were diluted by 1:5000 in 1% phosphate buffered saline (PBS) sonicated for 10 min, dispersed on a poly-l-lysine coated coverslip for 5 min, washed and then immersed in Prolong Glass (ThermoFisher Scientific, P36982) immersion media and attached to a microscope slide. A commercially prepared sample slide with 23 nm DNA origami beads labeled with Atto 488 was used for STED resolution calibration (Gattaquant GmbH, GATTA-Beads B – brightline). For the live T cell samples, lymph nodes and spleen were harvested from LifeAct-GFP donor mice. The lymph nodes and spleen were mechanically dissociated to a single cell suspension. T cells in the suspension were then activated by plating on anti-CD3 and anti-CD28 antibody ( each) coated tissue culture plates. Two days after activation, the T cells were transferred to culture flasks and further grown in complete Roswell Park Memorial Institute (RPMI) media containing of IL-2 for 3 more days. On day 5 post-activation, the T cells were used for microscopy studies. For the imaging, the activated T cells were plated on chambered coverslips (-Slide 8 well from IBIDI) coated with of recombinant ICAM-1 (R&D systems) in RPMI media without Phenol Red and with 2% (wt/vol) BSA and 10mM HEPES. The T cells were then imaged at 37°C after allowing them to adhere to the ICAM-1 coated surface for at least 1 h. The live tissue that was imaged was an organotypic cerebellar explant culture. To prepare the samples ex vivo cerebellar explants were prepared as previously described.23 Cerebella were dissected from P10 PLP-EGFP mice24 (Jax#33357), sliced at 300 mm on a McIlwain Tissue Chopper, and cultured on MilliCell membrane inserts (Millipore) in media containing 15% Hank’s balanced salt solution, 15% heat-inactivated horse serum, 68% minimum essential media, 125 mM HEPES, 28 mM D-Glucose, 2 mM L-Glutamine, and penicillin/streptomycin (Life Technologies) at 37°C. Media were exchanged every 2-3 days. Slices were incubated for 10 days prior to imaging. For imaging, slices were cut out of their inserts and placed tissue side down into a 35 mm glass-bottom dish. For the live cell and tissue measurements, we used a Tokai HIT heated stage to keep the cells and tissue at 37°C during the measurements. All experiments were performed according to protocols approved by the University of Colorado Anschutz Medical Campus Institutional Animal Care and Use Committee. 4.Results: DyMIN Imaging of Fluorescent Beads and Live Cells and TissueTo demonstrate our DyMIN implementation, we imaged 45 nm fluorescent bead samples. Five different 50-image sequence were taken with conventional STED and DyMIN STED on different areas of the sample with a field of view (FOV) of at , with 15 s between the start of each image. We took DyMIN sequences at four different combinations of the parameters, labeled DyMIN1-DyMIN4; see Table 1 for the parameters of each sequence. Note that all DyMIN sequences had the same step times and final step effective power, and the dwell time and STED power for the conventional STED sequence was the same as the final DyMIN step time and effective power. A larger FOV confocal image centered on the same field as the sequence (, ) was acquired before and after each sequence so that the bleaching before and after the image sequence could be measured. Table 1Parameters used for DyMIN and STED sequences.
Figure 4(a) shows the confocal image taken after each sequence with a larger FOV so the relative intensity compared to the unimaged regions can be seen. The fields that were imaged are dimmer than the surrounding field, but the photobleaching for all DyMIN sequences is lower than the STED comparison image. Figure 5(a) shows a bar graph of the ratio of the total confocal image counts in the regions of interest (ROIs) shown in Fig. 4(a) after to before the image sequence. The error bars indicate the standard deviation for the ratio, calculated by splitting the ROI into nine equal regions. The ratio for STED is lower than the ratio for all DyMIN sequences, with DyMIN maintaining 14% to 46% more of the initial fluorescence intensity, depending on the parameters selected. Note that the effective STED powers used in all these sequences are much lower than typically used in STED imaging, due to the lower effective STED power achieved during DyMIN (see Fig. S1 in the Supplementary Material), so these results are a lower bound on the bleaching prevented with DyMIN at the 2-3x higher STED powers typically used for imaging. To characterize the DyMIN and conventional STED image sequence resolutions we calculated the fixed 1/7 threshold Fourier ring correlation (FRC)25,26 between subsequent images () in the sequence and plotted the results for each sequence on the same axes. The FRC was calculated using the ImageJ plugin available from the BioImaging and Optics Platform (PTBIOP) update site. Figure 5(c) shows the resulting FRC curves, with the curves starting near the same value, indicating we had equivalent STED powers during imaging for the sequences. Over the course of the sequences the fluorescence signal drops due to photobleaching and thresholding during previous steps, which causes larger variation in the FRC calculated at higher image numbers. We performed Lorentzian fits on a 15-pixel ROI around each bead in the DyMIN/STED images and calculated the mean values of the amplitude of the beads over the course of the imaging sequence, scaled to the mean of the first three points. Figure 5(b) shows these plots, with the STED data showing a much faster decrease than the DyMIN traces. However, as shown in Fig. 5(d), the number of beads found by the fitting routine for the DyMIN traces decreases faster for all DyMIN sequences than for conventional STED. The DyMIN algorithm defines a pixel value at zero when the threshold is not reached. As a result, with changing sample intensities due to photobleaching, beads that are not bright enough to achieve threshold in steps 1 or 2 in DyMIN will disappear over time. The DyMIN algorithm defines a pixel value at zero when the threshold is not reached. As a result, with changing sample intensities due to photobleaching, the effective size of the bead decreases since the threshold levels remain the same over time. Figure S7 in the supplement from Ref. 18 showed a similar effect: that DyMIN can lead to scan-dependent bleaching effects in the sample. We next wanted to demonstrate that our DyMIN system could image live cells and will extend the number of images that can be acquired with good SNR. We imaged primary mouse T cells expressing lifeactGFP (a reporter for F-actin). In this sample, the T cells are very mobile and were relatively dim, making it a challenging sample for STED microscopy. We imaged a FOV with , with 30 s between images at DyMIN effective powers . The DyMIN step time parameters used were . The final dwell time was selected based on previous conventional STED time-lapse imaging on the samples, where we found it to be a reasonable balance between imaging speed and SNR. The thresholds selected were counts. Although these low thresholds make it more likely for a pixel without significant fluorescence to be exposed to STED light in the first images of the series, it did allow a series of 30 images to be taken for DyMIN while maintaining good visibility of the sample (Fig. 6 right), whereas the conventional STED image series quickly bleaches so that the sample is no longer visible on the same intensity scale (Fig. 6 left). The conventional STED imaging was done using the same dwell time, at a STED power of 24.0 mW. The final sample imaged was a cerebellar slice from a PLP-eGFP mouse prepared as described above. We imaged the tissue on a heated stage to maintain sample at a 37°C culturing environment. The parameters for DyMIN were counts, and effective STED powers of , using a FOV with . Figure 7 shows a timelapse imaging series of myelin along the length of an axon with 30 s total delay between the start of each image. The maximum fluorescence collected during the first image was 59 photons and that is the scale used to show the entire image series so that the effects of photobleaching will be evident. Although some bleaching is seen between the first and final images, the resulting series allows the sample to be imaged throughout the time series. 5.Discussion and ConclusionsWe have demonstrated an open-source implementation of the DyMIN method first reported by the Hell group in 2017. The DyMIN add-on is an inexpensive FPGA programmed with LabVIEW FPGA and custom electronics. Open source DyMIN should allow implementation on any STED microscope with the only requirements being photon counting detection, which is generally used for STED imaging, a pixel clock signal from the image acquisition software to provide a start signal for the DyMIN algorithm at each pixel, and the space to add an AOM into the STED beam path. DyMIN can be used to achieve increased resolution or signal-to-noise in 2D STED but is especially powerful for 3D STED imaging,18 or the focus here, time series imaging of live samples. We are interested in using DyMIN to extend the number of images we can take over the same field while imaging live cells and tissue, to observe dynamics as live samples interact with their environment or experience a photo-activation or uncaging laser that initiates a dynamical cellular process. One disadvantage of using DyMIN in image sequences is that, to extend the image sequence, one must start with a smaller threshold than ideal so that later images still reach the threshold. Unfortunately, this choice causes more photobleaching during the first images of a sequence. One modification that would improve DyMIN for longer term image sequences would be to use the photon counts in an image to modify the DyMIN step thresholds dynamically during image sequences so that the threshold decreases over time. Dynamic STED imaging of living samples interacting with the environment will be difficult to achieve without the ability to prevent significant photobleaching that DyMIN provides. The described DyMIN system will allow more research groups to implement this powerful technique in their labs and, by providing this in an open-source format, can help foster a community of active participation in microscopy technology development. Code and Data AvailabilityThe code, custom electronics design, and the data in this article are available in a GitHub repository, with a link provided in Ref. 21. AcknowledgmentsWe acknowledge assistance from Brendan Heffernan, who contributed to modifications to the MATLAB code for calculating the bead size and measured the laser pulse duration. Donald David from Cooperative Institute for Research in Environmental Sciences at the University of Colorado Boulder built the multiplexer circuit. Karl Kilborn provided custom modifications of the image acquisition software, Slidebook, to control the galvo mirror. Dr. Roland Wedlich-Söldner kindly gave the LifeAct-GFP donor mice that the T cell samples were prepared from. This work was supported in part by the following grants from the National Science Foundation (NSF) and the National Institute of Allergy and Infectious Diseases (NIAID), National Institutes of Health (NIH): NSF (Grant No. DBI-1919361) (E.A.G and D.R.), NSF (Grant No. DBI-1919541) (J.T.G and M.E.S.), and NIAID/NIH (Grant No. R01AI167943) (J.J.). The content of this work is solely the responsibility of the authors and does not necessarily represent the official views of the NIH or other funding agencies. ReferencesW. H. Stefan and W. Jan,
“Breaking the diffraction resolution limit by stimulated emission: stimulated-emission-depletion fluorescence microscopy,”
Opt. Lett., 19 780 https://doi.org/10.1364/OL.19.000780 OPLEDP 0146-9592
(1994).
Google Scholar
A. K. Thomas et al.,
“Fluorescence microscopy with diffraction resolution barrier broken by stimulated emission,”
Proc. Natl. Acad. Sci. U. S. A., 97 8206
–8210 https://doi.org/10.1073/pnas.97.15.8206
(2000).
Google Scholar
F. Balzarotti et al.,
“Nanometer resolution imaging and tracking of fluorescent molecules with minimal photon fluxes,”
Science, 355
(6325), 606
–612 https://doi.org/10.1126/science.aak9913 SCIEAS 0036-8075
(2017).
Google Scholar
K. C. Gwosch et al.,
“MINFLUX nanoscopy delivers 3D multicolor nanometer resolution in cells,”
Nat. Methods, 17
(2), 217
–224 https://doi.org/10.1038/s41592-019-0688-0 1548-7091
(2020).
Google Scholar
M. G. L. Gustafsson,
“Surpassing the lateral resolution limit by a factor of two using structured illumination microscopy. Short communication,”
J. Microsc., 198
(2), 82
–87 https://doi.org/10.1046/j.1365-2818.2000.00710.x JMICAR 0022-2720
(2000).
Google Scholar
M. G. Gustafsson,
“Nonlinear structured-illumination microscopy: wide-field fluorescence imaging with theoretically unlimited resolution,”
Proc. Natl. Acad. Sci. U. S. A., 102
(37), 13081
–13086 https://doi.org/10.1073/pnas.0406877102
(2005).
Google Scholar
E. Betzig et al.,
“Imaging intracellular fluorescent proteins at nanometer resolution,”
Science, 313
(5793), 1642
–1645 https://doi.org/10.1126/science.1127344 SCIEAS 0036-8075
(2006).
Google Scholar
J. R. Michael, B. Mark and Z. Xiaowei,
“Sub-diffraction-limit imaging by stochastic optical reconstruction microscopy (storm),”
Nat. Methods, 3 793
–796 https://doi.org/10.1038/nmeth929 1548-7091
(2006).
Google Scholar
R. Turcotte et al.,
“Dynamic super-resolution structured illumination imaging in the living brain,”
Proc. Natl. Acad. Sci. U. S. A., 116
(19), 9586
–9591 https://doi.org/10.1073/pnas.1819965116
(2019).
Google Scholar
C. Leterrier et al.,
“Nanoscale architecture of the axon initial segment reveals an organized and robust scaffold,”
Cell Rep., 13
(12), 2781
–2793 https://doi.org/10.1016/j.celrep.2015.11.051
(2015).
Google Scholar
E. D’Este et al.,
“Sted nanoscopy reveals the ubiquity of subcortical cytoskeleton periodicity in living neurons,”
Cell Rep., 10
(8), 1246
–1251 https://doi.org/10.1016/j.celrep.2015.02.007
(2015).
Google Scholar
C. A. Wurm et al.,
“Nanoscale distribution of mitochondrial import receptor tom20 is adjusted to cellular conditions and exhibits an inner-cellular gradient,”
Proc. Natl. Acad. Sci. U. S. A., 108
(33), 13546
–13551 https://doi.org/10.1073/pnas.1107553108
(2011).
Google Scholar
F. Gottfert et al.,
“Coaligned dual-channel STED nanoscopy and molecular diffusion analysis at 20 nm resolution,”
Biophys. J., 105
(1), L01
–L03 https://doi.org/10.1016/j.bpj.2013.05.029 BIOJAU 0006-3495
(2013).
Google Scholar
D. Wildanger et al.,
“A compact STED microscope providing 3D nanoscale resolution,”
J. Microsc., 236
(1), 35
–43 https://doi.org/10.1111/j.1365-2818.2009.03188.x JMICAR 0022-2720
(2009).
Google Scholar
V. Westphal and S. W. Hell,
“Nanoscale resolution in the focal plane of an optical microscope,”
Phys. Rev. Lett., 94
(14), 143903 https://doi.org/10.1103/PhysRevLett.94.143903 PRLTAO 0031-9007
(2005).
Google Scholar
J. Oracz et al.,
“Photobleaching in STED nanoscopy and its dependence on the photon flux applied for reversible silencing of the fluorophore,”
Sci. Rep., 7
(1), 11354 https://doi.org/10.1038/s41598-017-09902-x SRCEC3 2045-2322
(2017).
Google Scholar
S. Thorsten et al.,
“Far-field optical nanoscopy with reduced number of state transition cycles,”
Opt. Express, 19 5644 https://doi.org/10.1364/OE.19.005644 OPEXFF 1094-4087
(2011).
Google Scholar
J. Heine et al.,
“Adaptive-illumination STED nanoscopy,”
Proc. Natl. Acad. Sci. U. S. A., 114 9797
–9802 https://doi.org/10.1073/pnas.1708304114
(2017).
Google Scholar
N. Upmanyu et al.,
“Colocalization of different neurotransmitter transporters on synaptic vesicles is sparse except for VGLUT1 and ,”
Neuron, 110
(9), 1483
–1497.e7 https://doi.org/10.1016/j.neuron.2022.02.008 NERNET 0896-6273
(2022).
Google Scholar
J. G. Schloetel et al.,
“Guided STED nanoscopy enables super-resolution imaging of blood stage malaria parasites,”
Sci. Rep., 9
(1), 4674 https://doi.org/10.1038/s41598-019-40718-z SRCEC3 2045-2322
(2019).
Google Scholar
S. A. Pierce et al.,
“DyMINSTED4all,”
https://github.Com/cuneurophotonics/dyminsted4all
(2024).
Google Scholar
S. A. Meyer et al.,
“Super-resolution imaging of ciliary microdomains in isolated olfactory sensory neurons using a custom two-color stimulated emission depletion microscope,”
J. Biomed. Opt., 21
(6), 066017 https://doi.org/10.1117/1.JBO.21.6.066017 JBOPFO 1083-3668
(2016).
Google Scholar
Y. Liu et al.,
“Myelin-specific multiple sclerosis antibodies cause complement-dependent oligodendrocyte loss and demyelination,”
Acta Neuropathol. Commun., 5
(1), 25 https://doi.org/10.1186/s40478-017-0428-6
(2017).
Google Scholar
B. S. Mallon et al.,
“Proteolipid promoter activity distinguishes two populations of nG2-positive cells throughout neonatal cortical development,”
J. Neurosci., 22
(3), 876
–885 https://doi.org/10.1523/JNEUROSCI.22-03-00876.2002 JNRSDS 0270-6474
(2002).
Google Scholar
R. P. Nieuwenhuizen et al.,
“Measuring image resolution in optical nanoscopy,”
Nat. Methods, 10
(6), 557
–562 https://doi.org/10.1038/nmeth.2448 1548-7091
(2013).
Google Scholar
M. van Heel and M. Schatz,
“Fourier shell correlation threshold criteria,”
J. Struct. Biol., 151
(3), 250
–262 https://doi.org/10.1016/j.jsb.2005.05.009 JSBIEM 1047-8477
(2005).
Google Scholar
BiographyStephanie A. Pierce (formerly Meyer) is a senior science writer for New Product Information at Thorlabs. She was previously a senior research associate in the Department of Bioengineering, University of Colorado Anschutz Medical Campus. She earned her PhD in physics from the University of Colorado at Boulder with a certificate from the Optical Science and Engineering Program. Her PhD research was performed at NIST in the Optical Frequency Metrology group. Jordan Jacobelli is a cellular immunologist and associate professor in the Department of Immunology and Microbiology, University of Colorado Anschutz Medical Campus. He earned his PhD in immunology at the University of Rome ‘La Sapienza’ and did his post-doctoral training at the University of California San Francisco. His research is focused on understanding how cytoskeletal remodeling regulates lymphocyte migration, activation, and effector functions in response to environmental stimuli during homeostasis and in disease settings. Wendy B. Macklin is Distinguished Professor and Chair of the Department of Cell and Developmental Biology at the University of Colorado School of Medicine. She received her PhD from Stanford University and was a postdoctoral fellow at Harvard Medical School/Massachusetts General Hospital. She was elected a fellow of AAAS in 2020. She studies oligodendrocyte development and myelination, investigating animal models of altered brain development and those related to multiple sclerosis. Juliet T. Gopinath is the Alfred T. and Betty E. Look Professor of Electrical, Computer and Energy Engineering and Physics at the University of Colorado Boulder. She received her BS degree in electrical engineering from the University of Minnesota and her MS and PhD degrees at MIT. She has published over 200 peer-reviewed journal and conference publications. She is the recipient of an R&D 100 Award (2012) and is an Optica Fellow (2021). Mark E. Siemens is a professor in the Department of Physics and Astronomy at the University of Denver. He has a BS degree in engineering physics from Colorado School of Mines and a PhD in physics from the University of Colorado at Boulder. He was an NRC postdoctoral fellow, received an NSF CAREER award, and a Fulbright fellowship. His research focuses on generation, propagation, measurement, and interaction of optical vortices. Diego Restrepo is a systems neuroscientist with a background in biophysics. The goal of his research is to understand how brain circuits mediate learning and decision making to complex sensory input. His lab uses an interdisciplinary approach employing awake behaving multielectrode and advanced optical recording, closed loop optogenetics, and computational neuroscience. He is involved in developing novel approaches to study circuit function in robust collaborations with physicists and engineers. Emily A. Gibson is an associate professor in the Department of Bioengineering, University of Colorado Anschutz Medical Campus, with a joint appointment in the Department of Physiology and Biophysics and Neuroscience Program. She earned her PhD in physics from the University of Colorado at Boulder with a specialization in nonlinear optics. She is a DARPA early career award recipient. Her research focus is the development of microscopy and spectroscopy technologies for clinical and biomedical research. |
Stimulated emission depletion microscopy
Biological samples
Field programmable gate arrays
Photobleaching
Photons
Fluorescence
Microscopes