Paper
2 May 1997 Institutional safety certification of a research class IV laser for clinical photodynamic therapy
Jon A. Schwartz, Sharon L. Thomsen M.D., Erle Janssen, Sharon Erickson, Daniel E. Supkis, Jr.
Author Affiliations +
Abstract
In preparation for photodynamic therapy clinical trials, a research class IV argon ion-pumped dye laser was modified to allow delivery of laser radiation from the Laser Biology Research Laboratory (LBRL) to a surgical operating room (OR) located one floor below and over 50 meters away. Optical fibers and coaxial cable, protected by flame retardant conduit, were fed from the LBRL to the OR. A remote control box was constructed to allow physician control of the laser output from the OR. A safety-off and output-power control wee included in the control system. Safety issues involved calculations of maximum permissible exposure of OR staff to optical radiation, the classification of the laser output in the OR and dealing with the placement of relatively high-flux carrying optical fiber placed in utility chases. Operational considerations involved the calculation and measurement of optical transmission losses, the procedure for relaying operational status between the OR and laser facility and the necessity of conducting practice runs with the laser and OR staff.
© (1997) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Jon A. Schwartz, Sharon L. Thomsen M.D., Erle Janssen, Sharon Erickson, and Daniel E. Supkis, Jr. "Institutional safety certification of a research class IV laser for clinical photodynamic therapy", Proc. SPIE 2974, Laser and Noncoherent Ocular Effects: Epidemiology, Prevention, and Treatment, (2 May 1997); https://doi.org/10.1117/12.275242
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Laser safety

Photodynamic therapy

Safety

Optical fibers

Control systems

Fiber lasers

Eyewear

RELATED CONTENT

Laser safety aspects for medical applications
Proceedings of SPIE (December 12 2003)
Can broken fiber optics produce hazardous laser beams?
Proceedings of SPIE (July 07 1993)
Ophthalmological LH-58 set with the holmium laser
Proceedings of SPIE (March 01 1995)
Laser safety at high profile projects
Proceedings of SPIE (February 18 2011)
Angioscopic Monitoring of In Vivo Laser Angioplasty
Proceedings of SPIE (January 14 1986)

Back to Top