19 February 2009 3D in vivo swept source endoscopic OCT based on rigid GRIN lens rod probe
Author Affiliations +
Abstract
In this manuscript, we report a fast speed swept source endoscopic OCT system utilizing a 4.5mm diameter rigid GRIN lens rod probe. The rigid probe has a tunable working distance with tuning range 0~7mm. The system could take front scanning images with speed of 40frames per second (512 A line per frame). The probe can work in contact and non contact mode. Fast speed contact and non-contact imaging was demonstrated.
© (2009) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Gangjun Liu, Tuqiang Xie, Jun Zhang, George M. Peavy, Matthew Brenner, and Zhongping Chen "3D in vivo swept source endoscopic OCT based on rigid GRIN lens rod probe", Proc. SPIE 7168, Optical Coherence Tomography and Coherence Domain Optical Methods in Biomedicine XIII, 716812 (19 February 2009); https://doi.org/10.1117/12.809624